[Lamp, Therapeutic, Mercury Arc, Air-Cooled, Mobile, Complete, 110-Volt, 60-Cycle, Item 7126005] [From the U.S. Government Publishing Office, www.gpo.gov] Document Reserve W Ml WAR DEPARTMENT TECHNICAL MANUAL non-circulating LAMP, THERAPEUTIC, MERCURY ARC, AIR-COOLED, MOBILE, COMPLETE, 110-VOLT, I 60-CYCLE (ITEM 7126005) WAR DEPARTMENT • 25 SEPTEMBER 1944 <«/. >/ WAR DEPARTMENT TECHNICAL MANUAL TM 8-621 LAMP, THERAPEUTIC, MERCURY ARC, AIR-COOLED, MOBILE, COMPLETE, 110-VOLT, 60-CYCLE (ITEM 7126005) WAR DEPARTMENT • 25 SEPTEMBER 1944 United Stales Government Printing Office Washington : 1944 WAR DEPARTAIENT, Washington 25, D. C., 25 September 1944. 1 Al 8-621, Lamp, Therapeutic, Alercury Arc, Air-Cooled, Alobile, Complete, 110-Volt, 60-Cycle (Item 7126005), is published for the information and guidance of all concerned. [A. G. 300.7 (16 Aug 44).] By order of the Secretary of War : G. C. MARSHALL, Chief of Staff. Official : J. A. ULIO, Major General, The Adjutant General. Distribution : Armies (2) ; Corps (2) ; Depts (10) ; Sv C (5) ; ID 8 (4) ; lb 8 (1) ; IR 8 (3) (2) (1) ; IC 8 (2) ; Base C (5) ; Numbered Air Forces (5) ; Aled Deps, Continental except St Louis Aled Dep (25), St Louis Med Dep (50) ; T of Opn (10) ; PE (2). ID 8: T/O & E 8-550. IB 8: T/O & E 8-611; 8-610. IR 8 (3) : T/O & E 8-581 ; 8-580, Evac. Hosps. IR 8 (2) : T/O & E 8-560; 8-510; 8-590. IR 8 (1): I/O & E 8-580, Convalescent Hosps. IC 8: T/O & E 8-187; 8-572S; 8-667. For explanation of symbols, see FAI 21-6. ii CONTENTS Paragraph Page CHAPTER 1. INTRODUCTION. Section I. General ........................... 1-2 1 II. Description and data.................. 3-4 1 CHAPTER 2. OPERATING INSTRUCTIONS. Section I. General ................................. 5 4 II. Service upon receipt of equipment.... 6-7 4 III. Controls and instruments............... 8-9 20 IV. Operating instructions ............... 10-14 20 CHAPTER 3. MAINTENANCE INSTRUCTIONS. Section I. General ................................. 15 23 II. Lubrication ............................. 16 23 III. Preventive maintenance services....... 17-18 23 IV. Trouble shooting ..................... 19-21 23 V. Maintenance operations ............... 22-24 24 VI. Packing and storage................... 25-26 25 APPENDIX. LIST OF PARTS....................................... 26 iii CHAPTER 1 INTRODUCTION Section I. GENERAL 1. SCOPE, a. These instructions are published for the information and guidance of the personnel to whom this equipment is assigned. They contain information on the operation and maintenance of the equipment, together with illustrations identifying the component parts of the equipment. They apply only to lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, Medical Department Item No. 7126005, and are arranged in three chapters: Chapter 1. Introduction; Chapter 2, Operating Instructions; Chapter 3, Maintenance Instructions. b. All service parts are listed in the appendix. 2. RECORDS. There are no forms or records which are required to be kept in conjunction with equipment. Section II. DESCRIPTION AND DATA 3. DESCRIPTION. Lamp, therapeutic, mercury arc, air-cooled, mobile, is an ultraviolet ray generator for irradiating areas of the body. The ultraviolet rays are produced by a self-starting, alternating-current, hot cathode lamp made of quartz containing a minute quantity of mercury in vacuum. The lamp is so constructed that the distance and direction from the ultraviolet ray source to the surface to be irradiated is adjustable. 4. DATA. a. Electrical characteristics. (1) Voltage—100 to 125 volts. (2) Frequency—60 cycles. (3) Line current, (a) Burdick Model No. QA-45O-NH and General Electric No. P2013 Model F—approximately 6 amperes. (b) Hanovia Model No. S-2307-A—approximately 18 amperes. (4) Line wiring specifications—No. 14 conductors for power service up to 100 feet. (5) Fuse size required on electric current supply line—30 amperes, b. Radiation intensity. The lamp produces ample ultraviolet radiation, 1 Figure 1 2 Lamp, therapeutic, mercury arc, air-cooled, mobile, complete, UO-volt, 60-cycle. at a distance of 30 inches from the burner to the patient, to cause a first-degree erythema (mild reddening) on the average person after an exposure of 30 seconds. c. Manufacturers. (1) Burdick Model No. QA-450-NH—the Burdick Corporation, Milton, Wisconsin. (2) General Electric No. P2013 Model F the General Electric X-Ray Corporation, Chicago, Illinois. (3) Hanovia Model No. S-2307-A—the Hanovia Chemical and Manufacturing Company, Newark, New Jersey. 3 CHAPTER 2 OPERATING INSTRUCTIONS Section I. GENERAL 5. SCOPE. I his chapter contains information for the guidance of the personnel responsible for the operation of this equipment. Section II. SERVICE UPON RECEIPT OF EQUIPMENT 6. NEW EQUIPMENT, a. Unpacking. (1) Open shipping boxes carefully. (2) Remove packing materials. (o) Lift assemblies and parts from the boxes and unwrap. b. Assemblies. 1 he lamp is issued broken down into the following assemblies: (1) Base upright assembly. (2) Burner housing assembly. (3) Burner assembly. c. Assembling the Burdick model. (1) Screw the casters into the sockets on the bottom of the cast iron base. 7R03608 (fig. 2), and set on the floor. (2) Remove the two wing nuts from the screws on the bottom of the base housing, 7R03612 (fig. 2). (3) I lace the base housing on the cast iron base in such a manner as to allow the two screws from which the wing nuts were removed to go through the two holes provided for this purpose in the cast iron base. (4) Replace the two wing nuts on the screws from which they were removed on the bottom side of the cast iron base. Draw the wing nuts up tight. (5) Lock the cast iron carriage, 7R03656 (fig. 2), of the upright column assembly on the upright column. 7R03640 (fig. 2). by tightening the vertical track locking knob. 7R03668 (fig. 2). (6) Raise the upright column assembly so that the bottom end may be placed in the hole provided in the cast iron cover plate, 7R03616 4 1. 7R03608 2. 7R03618 3. 7R03720 4. 7R03638 5. SR00398 6. 7R03668 7. 7R03672 8. 7R03670 9. 7R03656 10. 7R03658 11. 7R03644 12. 7R03650 13. 7R03646 Base, cast iron Plate, cover, front panel, sheet metal Handlebar Switch, toggle, line, 7/|6-inch shank Cord, 3-conductor, No. 14, asbestos-packed Knob, locking, vertical track Arm, supporting, complete Knob, supporting arm, plastic Carriage, cast iron Bearing, ball, nice 402-2 Track, upright, complete Fork, pulley Pulley, wheel 14. 7R03652 15. 7R03710 16. 7R03678 17. 7R03704 18. 7R03702 19. SR00396 20. 7R03712 21. 7R03642 22. 7R03640 23. 7R03616 24. 7R03612 25. 7R03610 Cable, steel, stranded Ventilator, burner housing top, complete Fork Plate, side Housing, burner, sheet metal Nut, cap, 5/t6 x 18, hex Shutter Collar, upright column Column, upright Plate, cover, cast iron, complete Housing, base, sheet metal, complete Caster, swivel, 3-inch wheel ® © (Th* —(iT T/ 41 I aB—(20) ® / a ® I .—@ I-—@ \ ®—\ \ —<24) Figure 2. Item 7126005 lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle; Burdick model. 5 (fig. 2). Slide the upright column down until it rests on the cast iron base. (7) Tighten the setscrew on under side of the cast iron base against the lower end of the upright column. (8) Remove the supporting arm plastic knob, 7R03672 (fig. 2), of the burner housing assembly by loosening the setscrew in the knob and unscrewing the knob. Insert the supporting arm in the ball-bearing roller track of the cast iron carriage. 7R03656 (fig. 2). The “Burdick” 1. 7R03715 2. 7R03686 3. 7R03680 4. 7R03700 5. 7R03706 6. 7R03714 Clamp, shutter rod Reflector, complete Lever, friction locking, plastic Measure, tape, steel Plate, side, with spacer nut, complete Rod, shutter, complete 7. 7R03712 8. 7R03704 9. 7R03690 10. 7R03602 11. SR00401 12. 7R03692 Shutter Plate, side Clip, burner Burner, Burdick HB—455 Nut, thumb, battery terminal, 6 x 32 brass Screw, thumb, 6-32 x % inch |T~ . X. V- - ~~..... c' ■A 9 V \ I .. ..s ' ...............\ \ . © ® ® Figure 5. Burner housing assembly, Burdick model—bottom view. 6 1. 7R03948 2. 7R03942 3. 7R03944 4. 7R03946 5. 7R03910 3. SR00025 7. 7R03938 3. SR00075 9. 7R03940 10. SR00318 11. 7R04004 12. SR00332 Spring, adjustable upright Upright, stationary Upright, adjustable Collar, stationary upright Caster, swivel Plug, female, 2-prong, G.E. GC 363 Hook, burner cord Cord, Neoprene, No. 14, 2-conductor Bushing, plastic Nut, lock, conduit, ^-inch Clamp, cable Terminal, wire clamp, O.D. '/z-inch, 3/i6-inch hole, 1 inch long 13. 7R04006 14. 7R03936 15. 7R03934 16. SR00076 17. SR00026 18. SR00317 19. SR00316 20. 7R03908 21. 7R03922 22. 7R03954 23. 7R03952 Terminal, wire clamp Spring, strain release Bushing, plastic Cord, Neoprene, No. 14, 3-conductor Plug, male, small, 2-prong Plug, ground, female Plug, ground, male Base, cast iron Autotransformer, 400-watt Plate, friction Arm, supporting i I tr® g I IA JVJ 11 I ■■ ®2) •sk-**isw / y x "x.X"'~'\ .\ ) _____ i \ I’'?® Figure 4. Parts for lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, General Electric model F. 1. 7R03914 Pedal, line toggle switch 7. 2. SROO317 Plug, ground, female 8. 3. SR00075 Cord, Neoprene, No. 14, 2-conduotor 9. 4. 7R03938 Hook, burner cord 10. 5. 7R03978 Shutter 11. 6. 7R03988 Measure, tape, steel 7R03956 Lever, locking stud 7R03952 Arm, supporting 7R03944 Upright, adjustable 7R03950 Knob, locking, adjustable upright 7R03942 Upright, stationary /©) III * x—z / f (6) 1 If/® (5/ 0 0^ j y Figure 5. Lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, General Electric model F, 8 trademark on the supporting arm should be on the outside. Replace supporting arm plastic knob and tighten setscrew in knob. (9) Remove the hexagon nut, flat washer, and spacer bushing from the other end of the supporting arm, 7R03672 (fig. 2). Attach burner housing assembly to supporting arm by placing the end of the supporting arm through the center of the fork, 7R03678 (fig. 2) ; replace spacer bushing, washer, and hexagon nut. Draw up hexagon nut tightly. (10) To place the burner in the burner housing, proceed as follows: (a) Loosen the battery terminal thumb nuts, SR00401 (fig. 3), on the burner clips, 7R03690 (fig. 3). (b) Slide the end of the burner with the flexible wire lead into the right-hand burner clip, and push it through far enough so that the opposite end of the burner clears the other burner clip. (c) Slide the burner back so that it is evenly spaced in the two burner clips, and tighten the battery terminal thumb nuts. (d) Attach the flexibl e wire lead from the burner to the binding post on the reflector near the right-hand burner clip. (e) Remove the protective paper covering from the burner. Do not touch the quartz tube with the fingers. d. Assembling the General Electric model. (1) Set the base assembly on its four casters. (2) Make up the upright assembly by placing the adjustable upright spring, 7R03948 (fig. 4). inside the stationary upright, 7R03942 (figs. 4 and 5). Press the flanged end of the adjustable upright, 7R03944 (figs. 4 and 5), against the spring and force it about halfway into the stationary upright. Slip the stationary upright collar, 7R03946 (fig. 4), over the adjustable upright, and fasten it to the stationary upright by tightening the setscrews provided. Lock the adjustable upright by tightening the adjustable upright locking knob, 7R03950 (fig. 5). (3) Place the bottom of the upright assembly in the socket of the cast iron base and tighten the Allen head setscrews with the Allen setscrew wrench, SR00333 (fig. 6). provided. (4) Attach the supporting arm, 7R03952 (fig. 5), with the burner housing attached, by slipping it over the end of the adjustable upright. Line up the holes in the supporting arm with those at the end of the adjustable upright. Fasten the supporting arm in place by tightening the two fillister head screws provided. (5) Insert the cor;l burner hook. 7R03938 (fig. 5), into the hole on the under side of the supporting arm, and secure it in place by tightening the setscrews provided. (6) To place the burner in the burner housing, proceed as follows: (a) Place the burner. 7R03902 (fig. 7), in position by carefullv push- 9 1. 7R03902 2. 7R03996 3. SR00079 4. SR00325 5. 7R03956 6. 7R03958 7. SR00324 8. SR00115 9. SR00326 10. SR00103 11. SR00321 12. SR00319 13. SR00323 14. SR00322 15. SR00331 16. SR00236 17. 7R03994 18. SR00185 19. 7R03990 Burner, G.E. UA 2682 Plug, male, 2-prong, Hubbell P-646 Wire, Flamenol, No. 14, stranded, white Terminal, wire clamp, i.D. %-inch, 3/i6-inch hole, 1 inch long Lever, locking stud Stud, locking Screw, 8-32 x % inch, fl. H.M. Screw, 8-32 x % inch, R.H.M. Screw, 6-32 x % inch, R.H.M. Screw, 6-32 x 8 inch, R.H.M. Screw, 10-32 x % inch, till. H.M. Screw, 10-32 x 3/t6 inch, fill. H.M. Screw, 8-32 x 1 inch, fill. H.M. Screw, 8-32 x i/8 inch, fill. H.M. Screw, 8-32 x i/2 inch, fill. H.M. Screw, 6-32 x i/4 inch, B.H.M. Screw, 4-36 x '/4 inch, B.H.M. Nut, 8-32, hex, brass Washer, O.D., l9/32-inch, I.D. 3/8-inch, */|6-inch thick, fiber 20. 7R03992 21. SR00327 22. SR00328 23. SR00329 24. SR00152 25. 7R03998 26. 7R03984 27. 7R04000 28. 7R03976 29. 7R04002 30. 7R03986 31. SR00333 32. 7R03982 33. 7R03988 34. 7R03950 35. 7R03972 36. 7R03974 37. 71390 Washer, O.D. 19/32-inch, I.D. inch, '/32-inch thick, brass Washer, No. 6, brass Washer, No. 8, brass Washer, No. 9, brass Washer, lock, No. 8 Spacer, O.D. >/2 inch, I.D. scinch thick, fiber Spring, shutter Thumbscrew, 4-36 x 1 '/16-inch Clip, burner, quartz tube Nut, round, 4 x 36 Spring, quartz tube burner clip Wrench, setscrew, Allen, 5/l6 inch-18 Knob, shutter Measure, tape, steel Knob, locking, adjustable upright Spring, clip Clip, burner end Spectacles, ultraviolet i(7) it (?) (?) ’ t 1 lV • z® CFt V " , I । (V tit ■ / I X —*Q14y I! ® @ । X® .(4) /Jb O'-® M.. /fU w «« _ ■ o-@ o-@ VSl @z w ^VH) V ___jf J Figure 6. Parts for lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, General Electric model F. 10 1. 7R03978 Shutter 2. 7R03902 Burner, G.E. UA 2682 3. 7R03982 Knob, shutter 4. 7R03988 Measure, tape, steel 5. 7R03976 Clip, burner, quartz tube 6. 7R03974 Clip, burner end 7. 7R03968 Reflector, burner / —. .. —..........— M’—(3J / r a — / I / I J®X I J® I • |llh iiin^ [_ ’ O________(T) ”Tj) Figure 7. Burner housing, General Electric model—bottom view. ing it into the burner end clips, 7R03974 (fig. 7), and the quartz tube burner clip, 7R03976 (fig. 7). The seal-off tip of the burner (fig. 7 (2) ) is placed at the end away from the quartz tube burner clip. (Note that the quartz tube burner clip does not make contact with any metallic part.) (b) Place the thumbscrew, 7R04000, in the slots of the burner end clips, attach the round nuts, 7R04002, and tighten so that proper contact will be made. T1 1. 7R04252 Base, cast iron 19. 7R04300 Cable, stranded, steel 2. SR00075 Cord, Neoprene, No. 14, two-conductor 20. 7R04310 Housing, pulley wheel, cast iron 3. 4. 7R04214 7R04218 Plate, base cover Washer, 0.0. S/g-inch, I.D. I'/a-inch i/8 inch thick 21. 22. 7R04302 7R04260 Rod, locking carriage, steel, %-inch dia., 39-inch long Plate, right-hand side carriage 5. 7R04306 Collar, locking carriage rod 23. 7R04298 Knob, burner housing rail locking 6. SR00195 Plug, female, twist lock, Hubbell 24. 7R04270 Axle 7. SR00191 Plug, male, twist lock, Hubbell XT 7465 25. 26. 7R04318 7R04280 Spacer, burner housing rail Knob, vertical carriage locking 8. 9. 7R04286 7R04354 Hook, burner cable Trunnion, clamp 27. SR00075 Cord, Neoprene, No. 14, two-conductor 10. 7R04362 Shutter 28. 7R04304 Upright, tube, 6 feet 11. 7R04372 Knob, shutter 29. 7R04210 Switch, line toggle, Hubbell P 1650 12. 13. 7R04360 7R04334 Knob, trunnion locking Hood, sheet metal 30. 7R04208 Switch, canopy rotary starter, 3/8-inch shank, Leviton 14. 7R04358 Plate, friction, trunnion clamp 31. 7R04216 Plate, switch 15. 7R04324 Fork 32. 7R04220 Cylinder, base housing 16. 17. 7R04314 7R04322 Block, rail support Knob, rail-supporting block locking 33. SR00075 Cord, Neoprene, No. 14, two-conductor 18. 7R04316 Rail, burner housing 34. 7R04212 Caster, swivel, 3-inch wheel ®®@®®®-Gl „ \\W \ ffl \\\\ \ 'f / @1 /Xa\\ ' III //® \\\V\ EMM® |- If; ®—I I •—® ©/ Ad M® IWR @- ._L®.---------- I Rfrr rrWL @ ( 1 )—M I la-._®V_LullhilJj / O MTx-------------W- nfc (34) Figure 8. Lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, 11 anovia model. 12 (c) Remove the protective paper cover from the burner. Do not touch quartz tube with the fingers. (7) Set the steel tape measure, 7R03988 (fig. 6), in operating position by pulling out the steel tape and placing it in the slot of the burner housing directly below the tape reel (fig. 7 (4) ). e. Assembling the Hanovia model. (1) Slip the casters into the sockets on the bottom of the cast iron base, 7R04252 (fig. 8). Set the base assembly on its four casters. (2) Lock the carriage (fig. 8 (22) ) on the upright tube, 7R04304 (fig. 8), of the upright assembly by tightening the vertical carriage locking knob, 7R04280 (fig. 8). (3) Raise the upright assembly so that the bottom end may be placed in the hole provided in the base cover plate, 7R04214 (fig. 8). Lower the upright assembly until it seats. (4) Remove the two screws located at the extreme end of the burner housing rail (fig. 8 (25) ) of the burner housing assembly. Insert the burner housing rails, 7R04316 (fig. 8), horizontally into the wheel track of the carriage (fig. 8 (22) ) until the screw holes project on the opposite side of the carriage. Replace the two screws and tighten. If the burner housing rail does not move freely, loosen the rail-locking knob, 7R04322 (fig. 8). (5) To install the burner, proceed as follows: (a) Place the burner, 7R04336 (fig. 9), in the burner supports (fig. 9 (3) ) by grasping the metal end of the burner with the right hand in a position such that the fitting seal and the slotted metal fittings (fig. 9 (5) ) are toward the lamp. W ith the left hand push and hold to the left the left-hand burner support (fig. 9). Insert the right end of the burner into the right-hand burner support, -taking care that the right slotted metal burner fitting engages the Y-shaped guide. Fit the left-hand burner support over the left end of burner. Release the left-end support and then the burner. (b) Insert the burner pin leads (fig. 9 (7) ) into their respective porcelain terminal blocks, 7R04346 (fig. 9). (c) Grasp the burner end shield, 7R04338 (fig. 9), with the hand and place the fingers near the flanged edges. Slip the burner end shields over the ends of the burner and the beaded leads. Squeeze the flanged edges of the shield together with the fingers and insert the end burner shield flanges into the opening in the lamp reflector. Release the fingers so that the flanges engage the reflector. f. Prestarting instructions. (1) Clean the burner and reflector as outlined in paragraphs 22 and 23. (2) Before operating the lamp for the first time from any electric 13 1. 7R04338 Shield, burner end 2. 7R04340 Reflector, burner 3. 7R04342 Bracket, burner, complete 4. 7R04336 Burner, Hanovia S-2051 5. Slotted fitting issued as part of 7R04336 Burner, Hanovia S—20251 6. 7R04346 Block, terminal, porcelain 7. Burner pin leads issued as part of 7R04336 Burner, Hanovia S—20251 y® ®® ® ® & ® ® Figure 9. Burner housing, Hanovia model—bottom view. current supply source, determine the actual voltage at the receptacle through the use of a voltmeter, and make certain that both voltage and frequency correspond to that stamped on the nameplate. The lamp is adjusted at the factory for operation on 115-volt supply. If the supply voltage differs from this value by more than plus-or-minus 4 volts, the lamp will give better service if adjustment is made for the supply voltage. A terminal board with voltage adjustment taps is provided on the transformer located in the base assembly. W hen making voltage adjustment, it is only necessary to change one wire—the line connections to the transformer terminal board. After making the lamp voltage adjust- 14 1. 7R03926 Board, terminal 2. 7R03920 Bracket, starting relay 3. SR00313 Screw, 5/i6—18 x i/2 inch, R.H.M. 4. SROO144 Screw, 10-32 x 1'/4 inch, R.H.M. 5. SR00045 Screw, 10-32 x 3/8 inch, R.H.M. 6. SR00300 Screw, 10-32 x i/2 inch, R.H.M. 7. SR00010 Screw, 8-32 x % inch, R.H.M. 8. SR00040 Screw, 6-32 x '/4 inch, R.H.M. 9. SR00264 Screw, 10-32 x '/4 inch, B.H.M. 10. 7R03930 Screw, line toggle switch pedal 11. 7R03932 Spacer, 0.0. i/2 inch, I.D. 3/16 inch, Vi6 inch thick 12. SR00185 Nut, 8-32, hex, brass 13. SR00330 14. SR00152 15. SR00151 16. SR00156 17. SR00294 18. SR00328 19. SR00303 20. 7R03928 21. 7R03914 22. 7R03916 23. 7R03918 Nut, 10 x 32, hex Washer, lock, No. 8 Washer, lock, No. 10 Washer, lock, 5/|6-inch Washer, shakeproof, No. 8, internal teeth Washer, No. 8, brass Washer, No. 10, brass Condenser, with terminals Pedal, line toggle switch Switch, line toggle Relay, starting 15 ______(J) ' © ----© shmI —Cl) a a ~ *-----© ■HF &) ■hH ’^Mi O'—® WlatMl -J^f ©*—® ® JAL &-----® & ® o------® fi . O-------------® 1 aZ" o—® ■ B* ® Q.---------------------@ iL ____________o—® (SSh-® o-—® Figure 11. Farts for lamp, therapeutic, mercury-arc, air-cooled, mobile complete, 110-volt, 60-cycle, General Electric model F. incut, tighten all terminal nuts. Make the lamp voltage adjustment as follows: (a) Burdick model. Remove the front panel cover plate, 7R03618 (fig- 2), by unfastening the screws holding it in place so as to make the 1. 7R03968 Reflector, burner 2. 7R03980 Rod, shutter 3. 7R03970 Board, burner clip, asbestos 4. 7R03978 Shutter 5. 7R03962 Frame, right end 6. 7R03964 Frame, left end 7. 7R03960 Bracket, swivel 8. 7R03912 Plate, cover 9. 7R03966 Cover a. (5) I /z (T) .. . - ®r L pF^ Figure 11. Parts for lamp, therapeutic, mercury-arc, air-cooled, mobile complete, 110-volt, 60-cyclc, General Electric model F. 16 plastic terminal board, 7R03634 (fig. 14), accessible. Connect the line lead (fig. 14 (6) ) to the proper terminal tap as indicated in the table below: Actual line ■voltage Place lead on terminal tap 100-104 ................................................ 2 104-111 ................................................ 3 111-118 ................................................ 4 118-125 ................................................ 5 Replace the front panel cover by fastening with the screws that were removed. (b) General Electric model. Turn the lamp on its side. Remove the toggle switch pedal, 7R03914 (fig. 10), and the cover plate, 7R03912 (fig. 11), by unfastening the screws holding them so as to make the terminal board, 7R03926 (fig. 10), accessible. Connect the lead stamped 1. 7R04208 Switch, canopy rotary starter, 3/8-inch shank, Leviton 2. 7R04232 Board, terminal, transformer, complete 3. 7R04234 Spacer, fiber, 0.0. %-inch, 12-inch long 4. 7R04210 Switch, line toggle, Hubbell P 1650 5. 7R04230 Transformer 'X __________ Ct IBXX 5 * ® & ■&. *« w mm \ \i u a -- 9 \ A \l * * Elfe- d ® X x XX X X 1^,5 ? Sv*- " \ \ ’ % . v \ \ \ fira&W* Jef .x ■ x; » ■ -- «s»iv sioagiga IS “•■• ....... f - u Figure 12. Parts for lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, Hanovia model. 17 1. 7R04238 Board, fiber, 1 x 1% x 3/32 inch, %-inch hole 2. 7R04240 Board, fiber, 1 x 1% x 3/32 inch, !/4-inch hole 3. 7R04242 Washer, liber, O.D. %-inch, I.D. 5/i6-inch, 7/32 inch thick 4. 7R04256 Relief, strain, line cord 5. 7R04254 Relief, strain, burner cord 6. 7R04258 Spacer, O.D., %-inch, I.D. i/4 inch, 1 */2 inches long 7. SROO366 Bolt, 5/|6-18 x 2 inch, square H.M. 8. SR00156 Washer, lock, 5/)6 inch 9. SR00367 Nut, 5/|6 x 18, square 10. SR00355 Setscrew, i/4-20 x 3/8 inch, headless M., cup point 11. SR00372 Screw, 6-32 x 2 inch, fl. H.M. 12. 7R04250 Spacer, O.D. %-inch, I.D., i/4-inch, 3/4 inch long 13. SR00356 Bolt, 3/8-16 x 6 inch, square H.M. 14. SR00359 Washer, lock, 3/8-inch 15. SR00357 Nut, 3/8 x 16, square 16. SR00361 Bolt, |JZ4—20 x 1 >/4 inches, square H M 17. SR00293 Washer,' '/4 inch 18. SR00145 Screw, 10-32 x 1i/2 inches, R.H.M. 19. SR00277 Washer, shakeproof, No. 10, internal teeth 20. SR00273 Nut, 10-32, hex, brass 21. SR00358 Washer, lock, i/2-inch 22. SR00151 Washer, lock, No. 10 23. 7R04244 Fiber, sheet 24. 7R04236 Strap, condenser, brass 25. 7R04218 Washer, O.D. 3'/8 inch, I.D. 1’/z inches, i/8 inch thick 26. 7R04248 Board, terminal, line 27. SR00360 Condenser, 0.1 Mfd., 600 volts (jj __ z~x fl® Hi (?) ® ® ® ®. (TT) (7) © if (?) ® (SY I | I ° \ ** @ I fl s ■ O ^24) I Z® Figure 13. Parts for lamp, therapeutic, mercury arc, air-cooled, mobile, complete, 110-volt, 60-cycle, Pianovia model. 18 “X” to the proper terminal tap, as indicated in the table given in (a) on page 16. Replace the cover plate and toggle switch pedal by fastening with the screws that were removed. (c) Hanovia model. Remove the switch plate, 7R04216 (fig. 8), by unfastening the canopy rotary starter switch lock nut (fig. 8 (30) ) and the two switch-plate screws. Unfasten the screws holding the base cover plate, 7R04214 (fig. 8), and raise it to make the terminal board accessible. Connect the line lead to the proper terminal tap as indicated by the voltage markings on the terminal board, 7R04232 (fig. 12). Replace the 1. 7R03626 2. 7R03628 3. 7R03634 Receptacle, 3-prong, large, twist lock, polarized, Hubbell Autotransformer, 360-watt, G.E. cat. NO. 58 G 251 Board, terminal, plastic 4.. 7R03630 Bracket, terminal board 5. 7R03636 Socket, thermal switch 6. SR00390 Wire, No. 14, fixture, stranded 7. 7R03716 Switch, thermal 8. 7R03638 Switch, toggle, line, 7/|&-inch shank ® ------—j Z--X 1 Figure IF Farts for lamp, therapeutic, mercury arc, air-cooled, Burdick model. 19 base cover plate and switch plate by fastening with the screws and the lock nut that were removed. (3) With the line toggle switch in the “off” position, check to see that the burner clips are tight and burner leads are properly connected. (4) Both patient's and operator’s eyes should always be protected by wearing protective ultraviolet spectacles, Item No. 7139000, or X-ray operator’s goggles, Item No. 6031000. Failure to do so may result in an uncomfortable or dangerous case of conjunctivitis. 7. USED EQUIPMENT. To prepare used or reconditioned therapeutic lamps for operation, proceed exactly as for new equipment. (See par. 6.) Section III. CONTROLS AND INSTRUMENTS 8. CONTROLS a. Burdick model. T he Burdick model has a line toggle switch in the center of the cover plate on the base housing. This switch is marked “on” and “off”. b. General Electric model. The “on” and “off” switch on this model is of the foot treadle type. The ‘‘off” position is away from the white dot marked on the base housing. c. Hanovia model. (1) Line toggle on this model is marked “on" and “off,” and is located in the center of the base housing. (2) Rotary starting switch. The rotary starting switch is located adjacent to the line toggle switch on the base assembly. 9. INSTRUMENTS. The therapeutic lamp employs no instruments. Section IV. OPERATING INSTRUCTIONS 10. STARTING INSTRUCTIONS, a. Snap the line toggle switch to the “off” position. (1 ) Line toggle switch on the Burdick model, 7R03638 (tig. 14), is marked “on” and “off”. (2 ) For line toggle switch on the General Electric model (fig. 6(1)) the “off” position is away from the white dot. (3 ) Line toggle switch on the Hanovia model, 7R04210 (fig. 8), is marked “on” and “off". b. Connect burner line plugs. Ground General Electric model lamp by removing one screw holding the outlet faceplate and screw in the ground plug (fig. 5 (2) ). If the conduit system is not grounded, then provide a separate ground. c. Close burner housing shutters. 20 d. Snap the line toggle switch to the "on" position. A while clot on the base of the General Electric model indicates the “on" position. e. When a bluish glow is observed in the quartz burner tube, the lamp has started. If the bluish glow is not observed in the Hanovia model, twirl the rotary starter switch, 7R04208 (fig. 8), slowly until it is started. f. After the lamp has started, allow approximately 4 minutes for it to build up to its full operating intensity. The burner tube will then appear as a very brilliant narrow cord of blue light. Open the burner housing shutters. The lamp is then ready for use. 11. EXTINGUISHING INSTRUCTIONS. Snap the line toggle switch to the “off’' position. 12. EXTINGUISHING AND RELIGHTING INSTRUCTIONS, a. Do not try to relight the lamp immediately after it has been extinguished. 1 he burner must cool to a temperature at which the discharge can be reestablished. Wait approximately 6 minutes before relighting. b. If the lamp is to be employed for a number of successive applications, it is recommended that it be continued in operation instead of being turned off after each treatment. 13. LAMP MANIPULATING INSTRUCTIONS, a. Burdick model. (1) To rotate housing about its horizontal axis, loosen the plastic friction locking lever, 7R03680 (fig. 3), adjust to desired position, and tighten the friction locking lever. (2) To rotate the burner housing about its vertical axis, turn the burner housing fork. 7R03678 (fig. 2), to the desired position. (3) To adjust the height of the burner housing, loosen the vertical track locking knob, 7R03656 (fig. 2), move the burner housing assembly up or down to the desired position, and tighten the vertical track locking-knob. (4) To adjust the horizontal position of the burner housing, loosen the burner housing supporting arm locking knob, slide the supporting arm in or out to the desired position, and tighten the burner housing supporting arm locking knob. b. General Electric model. (1) To angulatc the burner housing to any position from the vertical to horizontal, loosen the locking stud lever, 7R03956 (fig. 5), adjust to the desired position, and tighten the locking stud lever. (2) To adjust the height of the burner housing, loosen the adjustable upright locking knob. 7R03944 (fig. 5), move the burner housing assembly up or down to the desired position and tighten the upright locking knob. 21 c. Hanovia model. (1) To move the burner housing on its horizontal axis, loosen the trunnion locking knob, 7R04360 (fig. 8), adjust to the desired position and tighten the trunnion locking knob. (2) To rotate the burner housing on its vertical axis, loosen the rail supporting block locking knob, 7R04322 (fig. 8), turn to desired position, and tighten the rail supporting block locking knob. (3) To adjust the height of the burner housing, loosen the vertical carriage locking knob, 7R04280 (fig. 8), move the carriage (fig. 8 (22) ) up or down to the desired position, and tighten the vertical carriage locking knob. (4) To adjust the horizontal position of the burner housing, loosen the burner housing rail locking knob, 7R04298 (fig. 8), slide the rails in or out to the desired position, and tighten the burner housing rail locking knob. 14. MEASURING THE DISTANCE BETWEEN BURNER AND PATIENT. Use the steel tape measure located on the burner housing. The intensity of ultraviolet radiation varies inversely as the square of the distance between the burner and the patient. For example, moving the patient a distance of 30 to 15 inches from the lam]) makes the radiation four times as intense, rather than twice as intense. Moving the patient three times closer would make the radiation nine times as intense. If this principle is not constantly borne in mind when positioning the patient, severe burns may result from moving him a short distance closer to the lamp. 22 CHAPTER 3 MAINTENANCE INSTRUCTIONS Section I. GENERAL 15. SCOPE. Chapter 3 contains information for the guidance of the personnel of the using organization responsible for the maintenance (1st and 2nd echelon) of this equipment. It also contains information needed for the performance of the preventive maintenance service and replacement of burners. Section II. LUBRICATION 16. BALL BEARINGS. Wear on these bearings is negligible, ft is not necessary to lubricate them. Keeping them clean and free of dust and dirt is all that is necessary. Section III. PREVENTIVE MAINTENANCE SERVICES 17. OPERATOR MAINTENANCE (1ST ECHELON), a. Weekly. (1) Clean the burner. (2) Clean the reflector. (3) Dust the entire lamp thoroughly, particularly the ball bearings. b. When lamp is put in operation for the first time, returned from repair, or a new burner'is installed. (1) Clean the burner. (2) Clean the reflector. 18. ORGANIZATIONAL MAINTENANCE (2D ECHELON). Replace burner when necessary. Section IV. TROUBLE SHOOTING 19. IF LAMP FAILS TO LIGHT, a. Check current source to see that it agrees with the voltage and cycle data stamped on the nameplate of the lamp. 23 b. Check current source to see that current is available at the receptacle into which the lamp line cord is plugged. c. Check line fuse to see that it is intact and of the proper amperage. d. Burdick model only. (1) Snap the line toggle switch to the “on” position. (2) Test to determine if current is reaching the burner, by placing a 300-volt a-c voltmeter across the burner clips. A normal “no load” voltage reading will be approximately 250 to 260 volts. If there is no voltage at the burner clips, snap the line toggle switch to the “off” position. (3) Make the thermal switch. 7R03716 (fig. 14), accessible by unfastening the screws holding the front panel cover plate, and removing the front panel cover plate. 7R03618 (fig. 2). (4) Tighten the thermal switch in its socket, 7R03636 (fig. 14), by rotating it slightly clockwise. (5) Snap the line toggle switch to the “on” position and test for current at the burner clips as described above. If there is no voltage at the burner clips, remove thermal switch and replace it with a new one. 20. IF THE LAMP LIGHTS AND IMMEDIATELY EXTINGUISHES ITSELF. I f the lamp lights and immediately extinguishes itself, it is likely that a. fuse in the electric supply circuit has blown. In the event that this is the cause of operating failure, replace fuse. 21. IF BURNER FAILS TO REACH NORMAL BRILLIANCY OR GOES OUT. I he burner should be replaced if it fails to reach normal brilliancy or goes out and does not light after being in use over a period of time. Section V. MAINTENANCE OPERATIONS 22. CLEANING BURNER. To clean the burner, snap the line toggle switch to the “off position, moisten a small piece of clean cloth with either acetone (Item No. 1003000), carbon tetrachloride (Item No. 1127000), or ethyl alcohol (Item No. 1048000), gently wipe the quartz tube with the moistened cloth, and dry with a clean cloth. Never use rubbing alcohol. I he cleaning solvents are listed in the order of their preference. If these solvents are not available, distilled water may be used. The burner should not be removed for cleaning. DO NOT ALLOW the fingers to come in contact with the quartz tube at any time. In the event that they do, clean as directed above. 23. CLEANING REFLECTOR. I o clean the reflector, snap the line toggle switch to the “off” position, wipe lightly with a clean cloth slightly moistened with water, and dry with a soft clean cloth. 24 24. REPLACEMENT OF BURNER, a. Remove the spent burner by carrying out the assembling instructions in reverse. (See par. 6.) (1) Burdick model. See paragraph 6c (10). (2) General Electric model. See paragraph 6d(6). (3) Hanovia model. See paragraph 6c(5). b. Install the new burner by carrying out the assembling instructions as referred to in a above. Section VI. PACKING AND STORAGE 25. DISASSEMBLING. To disassemble the lamp, carry out the assembly instructions for the respective models (par. 6c, d, and e) in reverse. 26. PACKING AND STORING, a. Disassemble the lamp into the base assembly, upright assembly, and burner housing assembly, with the burner in the burner clips. b. Wrap each assembly in heavy protective paper. Place each assembly in a sturdily constructed wooden box of suitable size. Use sufficient bracing and packing material to protect the parts in shipment and prevent the movement of parts within the boxes. Nail the lids in place on the boxes and band with steel strapping. Fasten the steel strapping with staples. 25 APPENDIX LIST OF PARTS 1. ITEM NO. 7126005, LAMP, THERAPEUTIC, MERCURY ARC, AIRCOOLED, MOBILE, GENERAL ELECTRIC X-RAY CORPORATION. Medical 1 )ept. No. Nomenclature Figure No. 10030 ACETONE, ACS 71390 SPECTACLES, ULTRAVIOLET 6 SR00006 COMMON PARTS Base Assembly WIRE No. 16, FIXTURE, STRANDED, for line toggle switch SR00010 SCREW. 8-32 x INCH, R.H.M., for line toggle switch 10 SR00040 SCREW, 6-32 x % INCH, R.H.M., for starting relay 10 SR00142 SCREW, 10-32 x X INCH, R.H.M., for terminal board SR00144 SCREW, 10-32 x 1% INCH, R.H.M., for terminal board 10 SR00152 WASHER, LOCK, No. 8, for line toggle switch 6 SR00156 WASHER, LOCK, %6 INCH, for autotransformer mounting plate 10 SR00264 SCREW, 10-32 x 14 INCH, B.H.M., for line cord relief spring .... 10 SR00294 SR00300 WASHER. SHAKEPROOF, No. 8, INTERNAL TEETFT, for starting relav. . SCREW, 10-32 x y2 INCH, R.H.M., for relay starting bracket 10 10 SR00313 SR00314 SCREW. 5/6-18 x y2 INCH, R.H.M., for autotransformer mounting plate SCREW. 10-32 x y, INCH, R.H.M.. for terminal board . 10 SROO315 SETSCREW. 5/lfi_i8 x INCH. ALLEN H.M., FLAT END, for stationary upright. *SR00025 Burner cord assembly PLUG, FEMALE, TWO-PRONG, G.E. G C 3C3 4 -SR00075 CORD. NEOPRENE. No. 14, TWO-CONDUCTOR . 4,5 4 SR00318 NUT. LOCK. CONDUIT, %-INCH, for plastic bushing — *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 26 Medical Dept. No. Nomenclature Figure No. *SR00079 Burner housing assembly WIRE, FLAMENOL, No. 14, STRANDED, XN^EII^ri^ for burner clamp bne 6 SR00103 SCREW, 6-32 x % INCH, R.H.M., for 6 SR00115 SCREW, 8-32 x % INCH, R.H.M., for 6 SR00152 WASHER, LOCK, No. 8, for burner clip SR00236 SCREW, 6A32 x % INCH, B.H.M., for 6 SR00319 SCREW, 10-32 x 3/16 INCH, FILL. H.M., 6 SR00320 SETSCREW. 8-32 x 5/16 INCH, HEAD-T A/T fnr burner cable book SR00321 SR00322 1 j 1 j 1V± « J. I) 1 ID ll 1 1 IV 1 l f I V— i ivi jj\ • • • • • • SCREW, 10-32 x % INCH, FILL. H.M., for locking stud lever assembly SCREW, 8-32 x % INCH, FILL. H.M., 6 6 SR00323 SCREW, 8-32 x 1 INCH, FILL. H.M.. 6 SR00324 SCREW, 8-32 x INCH, FL. H.M., for 6 SR00325 TERMINAL. WIRE CLAMP. O.D.. •%-INCH, 3/16-INCH HOLE, 1 INCH 6 SR00326 SCREW. 6-32 x % INCH, R.H.M., for 6 SR00327 WASHER. No. 6, BRASS, for burner 6 SR00328 WASHER, No.' 8, BRASS, for burner 6 SR00329 WASHER, No. 9, BRASS, for burner 6 *SR00026 *SR00076 Line cord assembly PLUG. MALE, SMALL. TWO PRONG. . CORD. NEOPRENE, No. 14, 3-CON- DUCTOR 4 4 SR00316 SROO317 SROO185 SROOO45 SR00151 PT UG GROUND MALE 4 PT UG GROUND FEMALE 4. 5 Miscellaneous NUT 8-32. HEX. BRASS SCREW. 10-32 x INCH, R.H.M WASHER LOCK No 10 6. 10 10 10 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 27 Medical Dept. No. Nomenclature Figure No. SROO3O3 WASHER, No. 10. BRASS . 10 SR00330 NUT, 10 x 32, HEX 10 SROO331 SCREW, 8-32 x y2 INCH, FILL. H.M.. 6 SR00332 TERMINAL, CLAMP, WIRE O.D. y2-INCH, 3/16-INCH HOLE, 1 INCH LONG Accessories 4 SR00333 WRENCH. SETSCREW, ALLEN. 5/io-INCH UNCOMMON PARTS Base Assembly 6 7R03908 BASE. CAST IRON . 4 7RO391O CASTER, SWIVEL. H-13696 x 20, 3-inch wheel, Bassick . . , 4 7R03912 PLA 1 E, COVER, for under side of cast iron base 11 7R03914 PEDAL. LINE TOGGLE SWITCH. 5. 10 *7R03916 SWITCH. LINE TOGGLE. G.E. No 1689 1 0 *7R03918 RELAY, STARTING. G.E. No. 59-165.. 10 7R03920 BRACKET, STARTING RELAY . 10 7R03922 AUTOTRANSFORMER, 400-WATT G.E. No. 59-155 . 4 7R03924 PLATE, AUTOTRANSFORMER MOUNTING 7R03926 BOARD, TERMINAL . . . 10 *7R03928 CONDENSER. WITH TERMINALS.. 10 7R03930 SCREW. LINE TOGGLE SWITCH PEDAI 10 7R03932 SPACER. O.D. %-INCH, T.D. 3/16-INCH, %6 INCH THICK, for terminal board. . . Burner Cord Assembly 10 7R03938 HOOK, BURNER CORD, attached to 7R03940 supporting arm 4.5 BUSHING, PLASTIC, for burner line cord. Burner Housing Assembly 4 *7R03902 BURNER. G.E. UA 2682. . 6.7 7R03952 ARM. SUPPORTING, for burner housing. 4.5 7RO3954 , PLATE, FRICTION, for locking stud lever. 4 7R03956 LEVER, LOCKING STUD /. 5.6 7R03958 STUD. LOCKING . .. 6 7R03960 BRACKET, SWIVEL, for burner housing. 11 7R 03962 FRAME, RIGHT END, for burner housing. 11 7R03964 FRAME, LEF1 END, for burner housing. 11 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 28 Figure No. Medical Dept. No. Nomenclature 7R03966 7R03968 7R03970 7R03972 7R03974 7R03976 7R03978 7R03980 7R03982 7R03984 7R03986 7R03988 7R03990 7R03992 7R03994 *71403996 7R03998 7R04000 7R04002 COVER, for burner housing...... REFLECTOR, BURNER.............. BOARD, BURNER CLIP, ASBESTOS. .. SPRING, CLIP, for encl burner clips. CLIP, BURNER END .............. CLIP, BURNER, QUARTZ TUBE ..... SFIUTTER ........................... ROD, SHUTTER ....................... KNOB. SHUTTER ................. SPRING, SHUTTER ............... SPRING, OUARTZ TUBE BURNER CLIP .2...................... MEASURE, TAPE, STEEL........... WASHER. O.D. 1%2-INCH, I.D. ^-INCH, i/16 INCH THICK, FIBER, for shutter assembly....................... WASHER, O.D. 1%2-INCH, I.D. ^-INCH, i/32 INCH THICK. BRASS, for shutter assembly....................... SCREW, 4-36 x % INCH, B.H.M., for shutter assembly............... PLUG, MALE, TWO-PRONG. HUB- BELL P-646, for burner cable. SPACER. O.D. ^-TNCH. I.D. %2-INCH, % INCH THICK, FIBER, for burner clip assembly ................. THUMBSCREW, 4-36 x U/16 INCH, for burner and clip assembly...... NUT, ROUND, 4 x 36, for burner end clip thumbscrew ................... 11 7,11 11 6 6.7 6,7 7,H 11 6.7 6 6 5,6,7 6 6 7R03934 7R03936 7R03942 7R03944 7 R 03946 7R 03948 7R03950 7R04004 7R04006 Line Cord Assembly BUSHING, PLASTIC, for strain release spring........................... SPRING, STRAIN RELEASE........... Lamp upright assembly UPRIGHT. STATIONARY, Outside tube. . UPRIGHT. ADJUSTABLE, Inside tube. . . COLLAR. STATIONARY UPRIGHT. . . . SPRING. ADJUSTABLE UPRIGHT....... KNOB. LOCKING. ADJUSTABLE UP- RIGHT ........................ M i s c e 11 a n c o it s CLAMP, CABLE, for line and burner cord as sen iljlies TERMINAL, WIRE CLAMP, G.E. No. 2034 CORD TIP.................... 4 4 4.5 4.5 4 4 5.6 6 6 6 6 6 6 4 4 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 29 2. ITEM NO. 7126005, LAMP, THERAPEUTIC, MERCURY ARC, AIR-COOLED, MOBILE, BURDICK CORPORATION. Medical Dept. No. Nomenclature Figure No. 10030 ACETONE, ACS 71390 SR00015 SPECTACLES, ULTRAVIOLET COMMON PARTS Base assembly SCREW, 8-32 x 1 INCH, R.H.M., for terminal board bracket SR00185 NUT, 8 x 32, HEX, BRASS *SR00026 *SROOO75 PLUG, MALE, SMALL, TWO-PRONG. . CORD, NEOPRENE, No. 14, TWO-CON- DUCTOR SR00107 SCREW, 6-32 x y2 INCH, R.H.M., for base housing SR00130 SCREW, 6-32 x INCH, O.H.M., for base housing SR00140 SCREW, 10-32 x 14 INCH, R.H.M., for base housing SROO152 WASHER, LOCK, No. 8, for terminal board SR00229 WASHER, LOCK, %-INCH, for mounting transformer SR00293 WASHER, %-INCH, for fastening base housing SR00378 SETSCREW. 5/16-18 x y INCH, SQUARE H.M., CUP POINT, for lamp upright SR00379 BOUT/ 5/16-18 x 1 INCH, HEX H.M., for handle bar SR00380 SR00381 SCREW, %-20 x 2 INCH, R.H.M., For fastening base housing . . f NUT, THUMB, % x 20, WING, for fastening base housing SR00382 SCREW, >4-20 X y INCH, FILL, H.M., for mounting transformer SR00383 NUT, LOCK, CONDUIT, ^-INCH, for line porcelain bushing SR00384 SCREW, 4-40 x J4 INCH, R.H.M., for line toggle switch SR00385 SR00386 SR00387 * ■ NU4,4 x 40, HEX, for line toggle switch. . SETSCREW, 10-24 x y INCH. HEAD- LESS, CUP POINT, for cast iron cover plate SCREW, 10-24 x y INCH, R.H.M., for terminal board bracket * lo be requisitioned, when required, from the supply depot. (No asterisk indicates thi’t the item is not stocked as a spare part, but can be obtained by special requisition.) 30 Figure No. Medical Dept. No. Nomenclature SR00388 SR00389 *SR00390 SR00014 SR00125 SR00369 SR00393 SR00394 SR00395 SR00396 *SR00397 *SR00398 SR00003 SR00106 SR00391 SR00392 SR00040 SR 00043 SR00105 SR00106 SR00151 SROO153 SR00279 SR00293 SR00297 SCREW, 10-24 X 1 INCH, R.H.M., for terminal board bracket.............. SCREW, 6-32 x 1% INCH, R.H.M., for terminal board bracket.............. WIRE, No. 14, FIXTURE, STRANDED. . Burner housing assembly SETSCREW, %-20 x % INCH, HEADLESS, CUP POINT, for supporting arm plastic knob ....................... SCREW, 8-32 x H INCH, O.H.M., BRASS, for shutter clamp............ BOLT, 5/i6-18 x 1% INCH, HEX H.M., for side plate with spacer nut...... WASHER, %6 INCH, FIBER, for supporting arm............................. WASHER, 5/16 INCH, BRASS, for support 14 mg arm.................................. SETSCREW, 8-32 x %6 INCH, HEADLESS M. CUP POINT, for supporting arm nut ......................... NUT, CAP, 5/16 x 18, HEX, for burner housing assembly .................. PLUG MALE. LARGE. THREE PRONG. POLARIZED, TWIST LOCK, HUBBELL ......................... CORD, 3-CONDUCTOR, No. 14, ASBESTOS PACKED, for .burner line. SCREW, 8-32 x y2 INCH, R.H.M., for carriage assembly.................. SCREW, 6-32 x 5/16 INCH, R.H.M., for upright column..................... SETSCREW. J4-20 x yA INCH, ALLEN H., CUP POINT, for upright column collar ............................ SCREW, %-20 x % INCH, FILL. H.M.. for carriage ball bearings......... 2 2 Miscellaneous parts SCREW. 6-32 x y INCH, R.H.M............ NUT. 6 x 32. HEX....................... SCREW. 6-32 x y INCH, R.H.M............ SCREW. 6-32 x 5/16 INCH, R.H.M......... WASHER. LOCK No. 10.................... WASHER, LOCK No. 6..................... WASHER, ^-INCH, BRASS.................. WASHER, % INCH................,........ WASHER. No. 10................|.... • *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part., but can be obtained by special requisition.) 31 Medical Dept. No. Figure No. Nomenclature SR00303 SR00388 SR00400 SR00401 7R03608 7R03610 7R03612 7R03614 7R03616 7R03618 7R03620 7R03622 7R03624 *7R03626 7R03628 7R03630 7R03632 7R03634 7R03636 *7R03638 *7R03716 7R03720 7R03602 7R 03670 7R03672 7R03674 7R03676 WASHER, No. 10, BRASS....... SCREW, 10-24 x 1 INCH, R.H.M. NUT, 10 x 24, HEX BRASS...... NUT, THUMB, BATTERY TERMINAL, 6 x 32, BRASS.............. UNCOMMON PARTS Base assembly BASE, CAST IRON................. CASTER, SWIVEL, 3-INCH WHEEL. . . HOUSING. BASE, SHEET-METAL, COMPLETE, assembly............ PLATE, COVER, BOTTOM, SHEET- METAL, for under side of base housing. . PLATE, COVER, CAST IRON, COM- PLETE, Assembly for top of base housing PLATE. COVER, FRONT PANEL, SHEET-METAL, for base housing. CLAMP, LINE CORD ............... SHIELD. LINE CORD, SHEET-METAL. BUSHING, PORCELAIN, for line cord. . . RECEPTACLE, THREE PRONG, LARGE TWIST LOCK, POLARIZED. HUBBELL ...................... AUTOTRANSFORMER, 360-WATT, G.E. No. 58 G 251............. BRACKET. TERMINAL BOARD, On autotransformer................ SPACER, O.D. H-INCH, I.D. 5/16 INCH. Tib INCH LONG, for terminal board. BOARD, TERMINAL, PLASTIC........ SOCKET, THERMAL SWITCH, on terminal board bracket......... SWITCH, TOGGLE, LINE, %6-INCH SHANK .......................... 3 2 2 2 2 2 14 14 14 14 14 14 SWITCH, THERMAL, Burdick Uviarc cutout ........................ 14 HANDLEBAR......................I 2 BURNER. Burdick HB-455 .......... 3 KNOB, SUPPORTING ARM. PLASTIC. . 2 ARM, SUPPORTING. COMPLETE, assembly for burner housing.... 2 BUSHING. O.D. %-INCH, I.D. 5/1(;-TNCH. %6 INCH THICK, for burner housing fork ................................ NUT, %6 x 18 HEX, tapped 8 x 32 for set-screw, used on supporting arm........ *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 32 Figure No. Medical Dept. No. Nomenclature 7R03678 *7R03680 7R03682 7R03684 7R03686 7R03688 7R03690 7R03692 7R03694 7R03696 7R03700 7R03702 7R03704 7R03706 7R03708 7R03710 7R03712 7R03714 7R03715 7R03640 7R03642 7R03644 7R03646 7R03648 7RO365O 7R03652 7R03654 7R03656 7R03658 7R03660 FORK, for burner housing....... LEVER, FRICTION LOCKING, PLASTIC, for burner housing........ SPRING, STRAIN RELIEF, for burner cord .......................... BUSHING, STRAIN RELIEF, for burner cord .......................... REFLECTOR, COMPLETE, assembly. . . . BRACKET, BURNER CLIP............ CLIP, BURNER................... SCREW, THUMB, 6-32 x INCH, for burner clip ....................... BUSHING, INSULATING, for burner lead binding post and burner clips. WASHER, O.D. 9/16-INCH, I.D. .^-INCH, % INCH THICK, FIBER, for burner lead binding post and burner clips. MEASURE, TAPE, STEEL........... HOUSING. BURNER, SHEET-METAL. with outside plates and ventilator top. . . . PLATE, SIDE, for burner housing. PLATE, SIDE, WTTPI SPACER NUT. COMPLETE, assembly........... NUT, SPACER. y16 x 18, ROUND, O.D. 1% INCH, 9/1(. INCH THICK, for side plate ....................... VENTILATOR. BURNER HOUSING TOP. COMPLETE, assembly...... SHUTTER ........................ ROD, SHUTTER, COMPLETE, assembly. CLAMP, SHUTTER ROD............. Upright and carriage assembly COLUMN, UPRIGHT ............... COLLAR, UPRIGHT COLUMN......... TRACK. UPRIGHT, COMPLETE, assembly ...................... PULLEY. WHEEL .................. 3 3 3 3 3 2 2 3 2 2 3 3 2 2 2 2 AXLE. PULLEY ................. FORK. PULLEY ................ CABLE. STEEL. STRANDED ....... COUNTERWEIGHT, in upright column. . . CARRIAGE. CAST IRON. . ....... BEARING, BALL. NICE 402-2, for carriage assembly ............... SPACER, O.D. SAB-INCH, I.D. %-INCH. 1/16 INCH THICK, for carriage assembly, j 2 2 2 2 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 33 Medical Dept. No. Nomenclature Figure No. 7R03662 BRAKE, BURNER HOUSING SUPPORTING ARM *7R03664 KNOB, LOCKING, BURNER HOUSING SUPPORTING ARM 7R03666 BRAKE. VERTICAL TRACK 7 R03668 KNOB. LOCKING. VERTICAL TRACK. . 2 3. ITEM No. 7126005, LAMP, THERAPEUTIC, MERCURY ARC, AIRCOOLED, MOBILE, HANOVIA CHEMICAL AND MANUFACTURING COMPANY. Medical Dept. No. Nomenclature ... r igure No. 10030 71390 ACETONE, ACS SPECTACLES, ULTRAVIOLET COMMON PARTS Base assembly *SR00026 PLUG. MALE, SMALL, TWO-PRONG. . SR00040 SCREW, 6-32 x <4 INCH, R.H.M., for SR00105 switch plate SCREW, 6-32 x INCH, R.H.M., for SR00145 transformer terminal board SCREW, 10-32 x \y2 INCH, R.H.M., for SR00156 line terminal board WASHER, LOCK, 5/16 INCH, for upright 13 *SR00195 support tube PLUG, FEMALE, TWIST LOCK, 13 SR00229 HUBBELL, 7464 WASHER, LOCK, J4-INCH, for trans- 8 SR00230 former mounting NUT, % x 20, HEX, for transformer SR00273 mounting NUT, 10-32, HEX, BRASS, for line SR00293 terminal board \\ ASHER, )4-INTCH, for transformer 13 SR00303 mounting WASHER, NTo. 10, BRASS, for line 13 SROO355 terminal board SETSCREW, 14-20 x INCH, HEAD- LESS M., CUP POINT, for swivel SR00356 casters BOLT, J4-16 x 6 INCH, SQUARE H.M., 13 for counter weights in base I 13 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 34 Medical Dept. No. Nomenclature Figure No. SROO357 NUT, y x 16, SQUARE, for counter SROO358 weights in base WASHER, LOCK, %-INCH, for starter 13 SROO359 switch WASHER, LOCK, ^4-INCH, for counter 13 *SR00360 weights CONDENSER, 0.1 MFD., 600-VOLT, 13 SR00361 D.C • BOLT, X-20 x Uy INCH, SQUARE 13 SR00366 H.M., for transformer mounting BOLT. 5/16-18 x 2 INCH, SQUARE H.M.. 13 SR00367 for upright support tube NUT, %6 x 18, SQUARE, for upright 13 SR00368 support tube SCREW, 10-32 x 7/8 INCH, R.H.M., 13 BRASS, for line terminal board Burner housing assembly *SR00078 WIRE, FLAMENOL, No. 14, SR00109 STRANDED, BLACK, for burner lines. . SCREW, 6-32 x 1 INCH, R.H.M., for porcelain terminal blocks *SR00191 PLUG, MALE, TWIST LOCK. HUB- BELL XT 7465, for burner line cord.... 8 SR00369 BOLT, 5/16-18 x \y4 INCH, HEX H.M., for trunnion assembly SR00267 WASHER, SCREW TYPE, SIZE y, FIBER, for SR00404 WASHER, 5/16-INCH, for trunnion assemblies . . Upright and carriage assembly SR00147 PIN, COTTER, i/16-INCH DIA. x y INCH, for rail looking knob Miscellaneous parts SR00003 SR00010 SR00015 SR00017 SCREW, 8-32 x y2 INCH, R.H.M SCREW, 8-32 x H INCH. R.H.M SCREW, 8-32 x 1 INCH, R.H.M NUT, 8-32, HEX SR00043 NUT, 6-32, HEX SR00045 SCREW, 10-32 x y INCH, R.H.M *SROOO75 CORD, NEOPRENE, No. 14, TWO- SR00111 CONDUCTOR SCREW, 8-32 x y4 INCH, R.H.M 8 SR00114 SCREW, 8-32 x y INCH. R.H.M SR00142 SCREW, 10-32 x y INCH, R.H.M *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 35 Medical Dept. No. Figure No. Nomenclature SR00151 SR00158 SR00229 SR00277 SR00294 SR00297 SR00300 SR00314 SROO33O SROO372 SR00374 SROO375 SJ400376 *7R04208 *7R04210 7R04212 7R04214 7R04216 7R04218 7R04220 7R04222 7R04224 7R04226 7R04228 7R04230 7R04232 7R04234 7R04236 7R04238 WASHER, LOCK, No. 10......... 13 WASHER, SHAKEPROOF, No. 6, INTERNAL TEETH .................. WASHER, LOCK, % INCH........... WASHER, SHAKEPROOF, No. 10, INTERNAL TEETH............. 13 WASHER, SHAKEPROOF, No. 8, INTERNAL TEETH ......... 1 WASHER, No. 10................ SCREW, 10-32 x /2 INCH, R.H.M.. SCREW, 10-32 x 5/8 INCH, R.H.M. NUT, 10 x 32 HEX............ SCREW, 6-32 x 2 INCH, FL.H.M. 13 WIRE, No. 14, COPPER, SOLID, INSULATED .................. WIRE, No. 16, COPPER SOLID INSULATED .................. NUT CAP, 10-X 32, HEX, BRASS.- UNCOMMON PARTS Base Assembly SWITCH. CANOPY ROTARY STARTER ^-TNCH SHANK, LEVITON........... SWITCH, LINE TOGGLE, HUBBELL P 1650 ......................... CASTER, SWIVEL, 3-INCH WHEEL. . PLATE, BASE COVER, for top of base assembly................. PLATE, SWITCH ............... WASHER, O.D. 3%-INCH, I.D. \/2- INCH, %-INCH THICK, for upright opening on base cover plate... CYLINDER, BASE HOUSING......... BRACKET, STARTER SWITCH ....... BRACKET, LINE TOGGLE SWITCH.. BASE, LINE TOGGLE SWITCH...... COUNTERWEIGHT, BASE............ TRANSFORMER ........ BOARD, TERMINAL, TRANSFORMER, I COMPLETE, assembly.............. SPACER, FIBER. O.D. INCH, \/2 INC I T LONG, for transformer terminal board ................ STRAP. CONDENSER. BRASS BOARD. FIBER, 1 INCH x l/4 INCH x 3/32 INCH, 3/ INCH HOLE, for transformer mounting .......... 8. 12 8, 12 8 8 8 8 8 12 12 12 13 13 *4o be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 36 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 37 Medical , Figure Nomenclature Dept. No. 7R04240 BOARD, FIBER, 1 INCH x 1^4 INCH x %2 INCH, 44 INCH HOLE, for transformer mounting ............. 13 7R04242 WASHER, FIBER, O.D. y INCH, I.D. 5/16 INCH, %2 INCH THICK, for transformer mounting .............. 13 7R04244 FIBER, SHEET, for condenser........... 13 7R04246 TUBE, UPRIGHT SUPPORT, CAST IRON............. • • v.................. 7R04248 BOARD, TERMINAL, LINE................. 13 7R04250 SPACER, O. D. 44-INCH, I. D. %-INCH 44-INCH LONG, for line terminal board. . 13 71404252 BASE, CAST IRON........... .. ......... 8 7R04254 RELIEF, STRAIN, BURNER CORD. ... 13 7R04256 RELIEF, STRAIN, LINE CORD. ...... 13 7R04258 SPACER, O. D. J^-INCH, I. D. %-INCH. I44-INCH LONG, for starter and line toggle switch...................... 1° B u r n er Housing A s s e mb I y 7R04314 BLOCK, RAIL SUPPORT, attached to burner housing fork.............................. 8 7R04316 RAIL, BURNER HOUSING.................... 8 7R04318 SPACER, BURNER HOUSING RAIL. ... 8 7R04320 BUMPER, BURNER HOUSING RAIL, RUBBER .................................... *71404322 KNOB, RAIL-SUPPORTING BLOCK LOCKING............................. 8 7R04324 FORK, for burner housing.............. 8 7R04326 BUSHING, TRUNNION CLAMP, O. D. %6 INCH, I.D. INCH, y INCH LONG: ..................................... 7R04328 MEASURE, TAPE, STEEL........................... 71404330 BOLT, 5/16-18 x ^-INCH, HEX H. M., for fastening steel tape measure............... 7R04332 BUSHING, TRUNNION, O. D. y16-TNCH. I. D. 5/16-INCH, %2 LONG................... 7R04334 HOOD, SHEET-METAL..................... *71404336 BURNER : Hanovia S-2051............. 7R04338 SHIELD, BURNER END.................. 7R04340 REFLECTOR, BURNER..................... 9 7R04342 BRACKET, BURNER, COMPLETE, ( assembly with clamps.............. 7R04344 BRACKET, for complete burner bracket assembly.................... • ........... 7R04346 BLOCK, TERMINAL, PORCELAIN, for burner lugs. .............................. Medical Dept. No. Nomenclature Figure No'. 7R04348 7R04350 7R04352 7R04354 7R04356 7R04358 *7R04360 7R04362 7R04364 7R04366 7R04368 7R04370 *7R04372 7R04374 7R04260 7R04262 7R04264 7R04266 7R04268 7R04270 7R04272 7R04274 7R04276 7R04278 *7R04280 7R04282 7R04284 7R04286 7R04288 BUSHING, PLASTIC, for burner line cord. I. . , NUI, LOCK, CONDUIT, for burner line I cord plastic bushing. SPRING, BURNER CLIP....‘ ' TRUNNION, CLAMP, for burner hous- _ ing........................... 8 TRUNNION, for burner housing PLATE, FRICTION, TRUNNION CLAMP............................ 8 KNOB, TRUNNION LOCKING.’....... 8 SHUTTER............ ' 8 ROD, SHUTTER........... CLAMP, SHUTTER ROD....... .....i” REFLECTOR, SHUTTER............. STRIP, TWO HOLE TAPPED, for shutter rod clamp........ KNOB, SHUTTER..........8' " ’ BOARD, TERMINAL, BURNER HOUS- ING.......................... Upright and Carriage Assembly PLATE, RIGHT-HAND SIDE CARRIAGE 8 PLATE, LEFT-HAND SIDE CARRIAGE SHAFT, ROLLER, COMPLETE, assembly...... for carriage................. BEARING, BALL, for burner housing rail. . SPACER, MEI AL, COMPLETE, assembly for carriage..... AXLE, for burner housing rollers........... 8 GUIDE, ROLLER, for burner housing rail. . BRACKET, VERTICAL CARRIAGE LOCK............. GUIDE, VERTICAL CARRIAGE LOCK,’ PLASTIC........................ SHOE, LOCKING, PLASTIC, for vertical carriage lock...... KNOB, VERTICAL CARRIAGE LOCK-’..... ING............................ 8 COLLAR STOP, PLASTIC, for vertical carriage travel.......... SPACER. O. D. %-INCH, I. D. %-INCH,.... % INCH LONG, for vertical carriage locking knob........... HOOK, BURNER CABLE: / ......’ ' ’ s ’ ’ ’ ’ WASHER, O. D. 34-INCH, I. D. %-INCH, ¥16 INCH THICK, for burner housing rail locking knob............... *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) 38 Figure No. Medical Dept. No. 7R04290 7R04292 7R04294 7R04296 *7R042f>8 7R04300 7R04302 7R04304 7R04306 7R04308 7R04310 7RO4312 7R04376 Nomenclature WASHER. O. D. M-INCH. I. D. ^-INCH, i/16-INCH THICK, FIBER, for burner housing rail locking knob........... SPRING RETAINER, for burner housing-rail locking knob.................. BRAKE, SHOE, PLASTIC, for burner housing rail....................... SPACER. O. D. H-INCH, I. D. %-INCH, %6-JNCH LONG, for housing rail-locking knob........................... KNOB, BURNER HOUSING RAIL LOCKING ........................... CABLE, STRANDED, STEEL, for carriage....................... ROD, LOCKING CARRIAGE, STEEL H-INCH DIA., 39 INCHES LONG. UPRIGHT. TUBE. 6 FEET:....... COLLAR, LOCKING CARRIAGE ROD.. WHEEL, PULLEY: .............. HOUSING, PULLEY WHEEL, CAST IRON ...................... COUNTERWEIGHT, CARRIAGE, in upright tube.. ANCHOR, CABLE 8 8 8 8 8 8 *To be requisitioned, when required, from the supply depot. (No asterisk indicates that the item is not stocked as a spare part, but can be obtained by special requisition.) U.S. GOVERNMENT PRINTING OFFICE: 1944-607134 For Sale by Superintendent of Documents, Washington 25, D. C. 39 UNT LIBRARIES DENTON TX 76203 lillHIIIIIIIIIH 1001895492