Year 2000 Computing Crisis: Key Actions Remain to Ensure Delivery of
Veterans Benefits and Health Services (Testimony, 04/20/99,
GAO/T-AIMD-99-152).
Pursuant to a congressional request, GAO discussed the Department of
Veterans' Affairs (VA) efforts to address the year 2000 computer
problem, focusing on the: (1) year 2000 readiness of automated systems
that support the delivery of benefits and health care services; (2)
compliance status of biomedical equipment used in patient care; and (3)
year 2000 readiness of the pharmaceutical and medical-surgical
manufacturers upon which VA relies. In discussing biomedical equipment
and pharmaceutical manufacturers, GAO provided information on the Food
and Drug Administration's (FDA) year 2000 efforts.
GAO noted that: (1) VA continues to make progress in its year 2000
readiness; (2) however, key actions remain to be performed; (3) for
example, the Veterans Benefits Administration and Veterans Health
Administration (VHA) have not yet completed testing of their
mission-critical systems to ensure that these systems can reliably
accept future dates--such as January 1, 2000; (4) also, VHA has not
completed year 2000 assessments of its facility systems, which can be
essential to ensuring continuing health care; (5) in addition, neither
VA nor FDA had implemented GAO's prior recommendation to review the test
results for biomedical equipment used in critical care/life support
environments; (6) further, VHA's pharmaceutical operations are at risk
because the automated systems supporting its consolidated mail
outpatient pharmacies are not year 2000 compliant; (7) VHA does not know
if its medical facilities will have a sufficient supply of
pharmaceutical and medical-surgical products on hand because it does not
have complete information on the year 2000 readiness of these
manufacturers; and (8) it is critical that these concerns be addressed
if VA is to continue reliably delivering benefits and health care.
--------------------------- Indexing Terms -----------------------------
REPORTNUM: T-AIMD-99-152
TITLE: Year 2000 Computing Crisis: Key Actions Remain to Ensure
Delivery of Veterans Benefits and Health Services
DATE: 04/20/99
SUBJECT: Y2K
Systems conversions
Information resources management
Veterans benefits
Computer software verification and validation
Strategic information systems planning
Medical equipment
Computer software
Systems compatibility
Pharmaceutical industry
IDENTIFIER: Y2K
Veterans Health Information Systems and Technology
Architecture
VA Veterans Integrated Service Network
VA Year 2000 Strategy
VBA Beneficiary Identification and Record Locator
Subsystem
******************************************************************
** This file contains an ASCII representation of the text of a **
** GAO report. This text was extracted from a PDF file. **
** Delineations within the text indicating chapter titles, **
** headings, and bullets have not been preserved, and in some **
** cases heading text has been incorrectly merged into **
** body text in the adjacent column. Graphic images have **
** not been reproduced, but figure captions are included. **
** Tables are included, but column deliniations have not been **
** preserved. **
** **
** Please see the PDF (Portable Document Format) file, when **
** available, for a complete electronic file of the printed **
** document's contents. **
** **
** A printed copy of this report may be obtained from the GAO **
** Document Distribution Center. For further details, please **
** send an e-mail message to: **
** **
** **
** **
** with the message 'info' in the body. **
******************************************************************
ai99152t GAO United States General Accounting Office
Testimony Before the Committee on Veterans' Affairs, U. S. Senate
For Release on Delivery Expected at 2: 30 p. m. Tuesday, April 20,
1999
YEAR 2000 COMPUTING CRISIS
Key Actions Remain to Ensure Delivery of Veterans Benefits and
Health Services
Statement of Joel C. Willemssen Director, Civil Agencies
Information Systems Accounting and Information Management Division
GAO/T-AIMD-99-152
GAO/T-AIMD-99-152
Page 1 GAO/T-AIMD-99-152
Mr. Chairman and Members of the Committee: We appreciate the
opportunity to participate in today's hearing on the Department of
Veterans' Affairs (VA) efforts to address the Year 2000 (Y2K)
computer problem. We will focus on the Y2K readiness of automated
systems that support the delivery of benefits and health care
services, the compliance status of biomedical equipment used in
patient care, and the Y2K readiness of the pharmaceutical and
medical- surgical manufacturers upon which VA relies. In
discussing biomedical equipment and
pharmaceutical manufacturers, we will also share with you
information on the Food and Drug Administration's (FDA) Y2K
efforts. 1 In brief, VA continues to make progress in its Y2K
readiness. However, key actions remain to be performed. For
example, the Veterans Benefits Administration (VBA) and Veterans
Health Administration (VHA) have not yet completed testing of
their mission- critical systems to ensure that these systems can
reliably accept future dates such as January 1, 2000. Also, VHA
has not completed Y2K assessments of its facility systems, which
can be essential to ensuring continuing health care. In addition,
neither VA nor FDA had implemented our prior recommendation to
review the test results for biomedical equipment used in critical
care/ life support environments. Further, VHA's pharmaceutical
operations are at risk because the automated systems supporting
its consolidated mail outpatient pharmacies are not Y2K compliant.
Finally, VHA does not know if its medical facilities will have a
sufficient supply of pharmaceutical and medical- surgical products
on hand because it does not have complete information on the
Y2K readiness of these manufacturers. It is critical that these
concerns be addressed if VA is to continue reliably delivering
benefits and health care.
1 Biomedical equipment refers to both medical devices regulated by
FDA, within the Department of Health and Human Services, and
scientific and research instruments, which are not subject to FDA
regulation.
Page 2 GAO/T-AIMD-99-152
Key Actions Remain to Ensure That VA Can Deliver Benefits and
Health Care Into the Next Century
Like many organizations, VA faces the possibility of computer
system failures at the turn of the century due to incorrect
information processing relating to dates. The reason for this is
that in many systems, the year 2000 is indistinguishable from
1900, since the year is represented only by 00. This could make
veterans who are eligible for benefits and medical care
appear ineligible. If this happens, the issuance of benefits and
the provision of medical care that veterans rely on could be
delayed or interrupted.
As we reported last August, 2 VBA had made progress in addressing
the recommendations in our May 1997 report 3 and in making its
information systems Y2K compliant. For example, VBA changed its
Y2K strategy from developing new systems to converting existing
ones. It also reported it had
renovated 75 percent of its mission- critical applications as of
June 1998. At the same time, VHA reported it had assessed all and
renovated the vast majority of its mission- critical information
systems. Despite this progress, VBA was making limited progress in
renovating two key mission- critical applications the compensation
and pension online application and the Beneficiary Identification
and Record Locator Sub- System. And, except for its Insurance
Service, VBA had not developed business continuity and contingency
plans for its program services Compensation and Pension (the
largest), Education, Loan Guaranty, and Vocational Rehabilitation
and Counseling to ensure that they would continue to operate if
Y2K failures occurred.
VHA's Y2K program likewise had areas of concern. For example,
although VHA's medical facilities had hospital contingency plans,
as required by the Joint Commission on Accreditation of Healthcare
Organizations, they had not yet completed Y2K business continuity
and contingency plans. To address these areas and to reduce the
likelihood of delayed or interrupted benefits and health care
services, we recommended that VA
reassess its Y2K mission- critical efforts for the compensation
and pension online application and the Beneficiary Identification
and Record Locator Sub- System, as well as other information
technology initiatives, such as special projects, to ensure that
the Y2K efforts have 2 Year 2000 Computing Crisis: Progress Made
in Compliance of VA Systems, But Concerns Remain (GAO/AIMD-98-237,
August 21, 1998).
3 Veterans Benefits Computer Systems: Risks of VBA's Year 2000
Efforts (GAO/AIMD-97-79, May 30, 1997).
Lett er
Page 3 GAO/T-AIMD-99-152
adequate resources, including contract support, to achieve
compliance in time; establish critical deadlines for the
preparation of business continuity and contingency plans for each
core business process or program
service so that mission- critical functions affecting benefits
delivery can be carried out even if software applications and
commercial- off- theshelf (COTS) products fail, including a
description of resources, staff roles, procedures, and timetables
needed for implementation; and ensure rapid development of
business continuity and contingency plans
for each medical facility so that mission- critical functions
affecting patient care can be carried out if software
applications, COTS products, and/ or facility- related systems and
equipment do not function properly, including a description of
resources, staff roles, procedures, and timetables needed for
implementation. 4
VA Continues to Make Progress
VA has been responsive to our recommendations. For example, VBA
reassessed its mission- critical efforts for the compensation and
pension online application and the Beneficiary Identification and
Record Locator Sub- System, as well as other information
technology initiatives. It also reallocated resources to ensure
that the Y2K efforts had adequate resources, including contract
support, to achieve compliance.
In addition, VBA completed a draft business continuity and
contingency plan in January 1999 for its core business processes,
as well as a related planning template for its regional offices.
The plan provides a high- level overview of the resources, staff
roles, procedures, and timetables for its implementation. It
addresses risks, including mitigation actions to reduce
the impact of Y2K- induced business failures, and analyzes the
effect on each business line of a number of potential Y2K
disasters such as loss of electrical power, loss of
communications, loss of data processing
capabilities, and failure of internal infrastructure. According to
VBA, the plan, which it expects to test this August, is an
evolving document, to be revised and updated periodically until
January 1, 2000. VBA's plan makes no reference to contingencies
for the failure of three of VBA's benefits payment systems
Compensation and Pension, Education, and Vocational Rehabilitation
and Counseling. However, it is currently developing a payment
contingency plan for these systems and expects this 4 GAO/AIMD-98-
237, August 21, 1998.
Page 4 GAO/T-AIMD-99-152
to be completed in May 1999. A VBA official told us that the
payment contingency plan should have been referenced in VBA's
business continuity and contingency plan and will be in future
versions. The current plan also does not contain the designation
of an information technology security coordinator and a physical
security coordinator individuals that VBA
acknowledges are essential to the agency's Y2K efforts with
responsibility for ensuring overall security for VBA's network and
web site and for backing up data storage before, during, and
following January 1, 2000. This type of information will be
necessary if security- related failures occur. According to VBA,
it expects to designate these individuals by August 1999. VHA has
also made progress in developing business continuity and
contingency plans for its medical facilities. Last month, VHA
issued its Patient- Focused Year 2000 Contingency Planning
Guidebook to its medical facilities describing actions they can
take to minimize Y2K- related disruptions to patient care. The
guidebook discusses how a medical facility should develop
contingency plans for each major hospital function such as
radiology, pharmacy, and laboratory as well as for each major
support function such as telecommunications, facility systems,
medical devices, and automated information systems. The guidebook
also contains examples of plans, policies, and solutions for
problems that a medical facility may face and provides Y2K
templates describing the areas a facility should address by
specific hospital function. VA provided this guidebook to the
medical facilities early last month and expects the facilities to
use it to
prepare their individual business continuity and contingency
plans, set to be completed by April 30. The guidebook stresses
that these plans should be tested and suggests that the medical
facilities begin testing in June. The guidebook addresses external
emergency preparedness as well as internal operations.
Specifically, it discusses three functions that a medical facility
should perform in order to ensure that potential external hazards
are considered and planned for. These are (1) performing an
assessment of hazard vulnerabilities that is, the types and kinds
of Y2K problems that are anticipated within the community, (2)
conducting an inventory of
community resources people, money, clinical space, supplies, and
equipment available to address these hazards, and (3) closing the
gap between vulnerabilities and capabilities by putting into place
measures that will mitigate potential disruptions in critical
services by developing new working relationships with various
government agencies, non- VA health care organizations, and
vendors of critical supplies.
Page 5 GAO/T-AIMD-99-152
In addition to implementing our recommendations, VA continues to
make progress renovating, validating, and implementing its
systems. On March 31, 1999, VA reported to the Office of
Management and Budget (OMB) that the department had renovated and
implemented all of the mission- critical applications supporting
its 11 systems areas. As shown in table 1, VBA has
six of these areas, and VHA has two.
Table 1: Reported Status of VA's Mission- Critical Computer
Systems Areas and Their Applications a Of this total, 316
applications were renovated and two were replaced.
Source: VA. We have not independently verified this information.
Component/ office (number of systems) Systems areas Number of
applications
renovated or replaced
Veterans Benefits Administration (6) Compensation and Pension 30
Education 24 Insurance 3 Loan Guaranty 19 Vocational
Rehabilitation 4 Administrative 27
Total 107
Veterans Health Administration (2) Veterans Health Information
Systems and Technology Architecture 105 Veterans Health
Administration Corporate Systems 95
Total 200
National Cemetery System (1) Burial Operations Support System/
Automated Monument Application System 1 Reengineer 1
Total 2
Office of Financial Management (2) Personnel and Accounting
Integrated Data 8 Financial Management System 1
Total 9 VA total 11 318 a
Page 6 GAO/T-AIMD-99-152
Testing of Mission- Critical Systems Not Yet Complete Complete and
thorough Y2K testing is essential to providing reasonable
assurance that new or modified systems will process dates
correctly and
will not jeopardize an organization's ability to perform core
business operations. Because the Y2K problem is so pervasive,
potentially affecting an organization's systems software,
applications software, databases, hardware, firmware, embedded
processors, telecommunications, and interfaces, the requisite
testing can be extensive and expensive. Experience is showing that
Y2K testing is consuming between 50 and 70 percent of a
Y2K project's time and resources. According to our Y2K guide, 5 to
be done effectively, testing should be planned and conducted in a
structured and disciplined fashion. Our guide describes a step-
by- step framework for managing Y2K testing, which includes the
following key processes:
Software unit testing to verify that the smallest defined module
of software (individual subprograms or procedures) continues to
work as intended. Software integration testing to verify that
units of software, when combined, continue to work together as
intended. Typically, integration testing focuses on ensuring that
the interfaces work correctly and that the integrated software
meets requirements.
System acceptance testing to verify that the complete system that
is, the full complement of application software running on the
target hardware and systems software infrastructure satisfies
specific requirements and is acceptable to users. This testing can
be run
separately or in some combination in an operational environment
(actual or simulated) and collectively verifies that the entire
system performs as expected.
According to VBA and VHA officials, their testing criteria were
based on their software development life cycle guidance documents.
They said that upon successful completion of software unit and
integration testing, a
system is considered Y2K compliant. They said this type of testing
had been completed for all of their mission- critical systems.
As of March 31, 1999, neither VBA nor VHA had completed systems
acceptance testing which requires that each system be tested,
including 5 Year 2000 Computing Crisis: A Testing Guide (GAO/
AIMD- 10. 1.21, November 1998).
Page 7 GAO/T-AIMD-99-152
full future- date testing, on a compliant platform for all their
missioncritical systems. Specifically, according to VBA officials,
the agency had completed systems acceptance testing for half of
its mission- critical systems Insurance, Loan Guaranty, and
Vocational Rehabilitation and
Counseling. According to VBA's Hines, Illinois, data center Y2K
coordinator, systems acceptance testing of the Compensation and
Pension systems just started on April 14, 1999. According to a VBA
official, one of the reasons for the late systems testing was that
the new IBM compiler 6 at its Hines data center was not available
for use until February 1999. According to VBA, the Compensation
and Pension and most of the Education systems will be future- date
tested throughout April.
VHA also plans to begin system acceptance testing of its mission-
critical systems this month and complete it this June. According
to VHA officials, they could not perform this type of testing
before March of this year because VHA did not have a separate Y2K-
compliant test environment to
isolate the testing from the hospital systems in use. In addition
to testing of individual systems, end- to- end testing of multiple
systems is also critical. End- to- end testing, as defined in our
test guide, verifies that a defined set of interrelated systems,
which collectively
support an organizational core business area or function,
continues to work as intended in an operational environment,
either actual or simulated. For example, in order to successfully
process a compensation benefit payment to a veteran, VBA's
Compensation and Pension System must work correctly with its
Beneficiary Identification and Records Locator SubSystem,
Treasury's Financial Management System, the Federal Reserve
System, and financial institution systems.
VBA and VHA plan to conduct end- to- end testing between now and
this July. VBA is defining end- to- end testing as verification
that core missioncritical business functions, including benefits
payments and vendor and payroll payments, process correctly. The
interfaces between VBA's benefits systems and Treasury's Financial
Management System are to be tested in
May. VBA also plans to test transactions that interface with VHA
systems, such as information related to veteran eligibility. VHA
is defining end- toend testing as verification that core mission-
critical business functions, including patient- care transactions
and vendor and payroll payments,
6 A compiler is a computer program that converts human- readable
source code into a sequence of machine instructions that the
computer can run.
Page 8 GAO/T-AIMD-99-152
process correctly. Once these tests are completed, VBA and VHA
plan to conduct a business process simulation during the July 4,
1999, weekend. This simulation of day- to- day work at VA is to
include users at the VBA regional offices and VHA test
laboratories, who will simulate various transactions and process
them through a set of interrelated systems necessary to complete a
core business function. VBA expects to pretest the business
process simulation during May. Assessment of VHA's
Facility Systems Not Yet Complete
VA's facility systems are essential to the continued delivery of
health care services. For example, heating, ventilating, and air
conditioning equipment is used by hospitals to ensure that
contaminated air is confined to a specified area such as an
isolation room or patient ward. If computer systems used to
maintain these systems were to fail, any resulting climate
fluctuations could affect patient safety. Despite their
importance, VHA has not yet completed assessments of its facility
systems. As of February 28, 1999, VHA medical facilities reported
that they had assessed 55 percent of their facility systems.
According to VHA's Director of Safety and Technical Programs, the
remaining 45 percent have not been fully assessed primarily
because (1) facility systems tend to
be a combination of unique elements that have to be separately
assessed for compliance a time- consuming process and (2) VHA is
still awaiting compliance status information from facility- system
manufacturers. VHA
has not established milestones for completing assessments and
implementation of compliant facility systems. To help ensure that
sufficient time remains to complete these activities, we recommend
that VHA consider setting such deadlines.
In the event that facility- related systems and equipment do not
function properly due to Y2K problems, VHA medical facilities will
need to ensure that they have business continuity and contingency
plans addressing how
mission- critical functions affecting patient care will be carried
out. According to VHA's Director of Safety and Technical Programs,
most of VHA's facility systems have some kind of manual override
or reset that will allow them to continue functioning after a Y2K
problem. The director agreed, however, with the importance of
developing contingency plans that
fully document continued delivery of essential services in the
event of a facility system failure. VHA medical facilities expect
to have individual business continuity and contingency plans
completed by April 30.
Page 9 GAO/T-AIMD-99-152
On April 14, 1999, VA informed us that its February 28, 1999,
report contained an error. The corrected numbers for facility
systems (after fixing the reporting error) at the end of February
were 91 percent assessed and 9 percent not assessed.
Biomedical Equipment: Additional Status Information Available, But
Test Results Not Reviewed The question of whether biomedical
equipment such as magnetic resonance imaging (MRI) systems, x- ray
machines, pacemakers, and cardiac monitoring equipment can be
counted on to work reliably on and
after January 1, 2000, is also critical to VHA. To the extent that
this equipment uses embedded computer chips, it is vulnerable to
the Y2K problem. Such vulnerability carries with it possible
safety risks. This could range from the more benign such as
incorrect formatting of a printout
to the most serious such as incorrect operation of equipment with
the potential to adversely affect the patient. The degree of risk
depends in large part on the role the equipment plays in a
patient's care.
Additional Biomedical Equipment Status Information Available
Last September we testified that VHA was making progress in
assessing its biomedical equipment, but that it did not know the
full extent of the Y2K problem with this equipment because it had
not received compliance information from 398 manufacturers (26. 7
percent). 7 According to VHA, as of March 16, 1999, the number of
nonresponsive manufacturers had been reduced to 126 (8. 5
percent). 8 As shown in table 2, about 19 percent of the
manufacturers in VHA's database of suppliers had at least one
biomedical equipment item that was either noncompliant or
conditionally compliant.
7 Year 2000 Computing Crisis: Leadership Needed to Collect and
Disseminate Critical Biomedical Equipment Information (GAO/T-AIMD-
98-310, September 24, 1998). 8 According to VHA, 101 of the 126
letters requesting compliance information sent to manufacturers
were marked return to sender.
Page 10 GAO/T-AIMD-99-152
Table 2: Status of Manufacturer Responses to VHA as of March 16,
1999
a For inclusion in this category, 100 percent of the
manufacturer's products had to be considered compliant. b For
inclusion in this category, only one of the manufacturer's
products had to be considered noncompliant. c For inclusion in
this category, the manufacturer had to have no noncompliant
equipment, no equipment pending, and at least one conditional-
compliant item. d For inclusion in this category, the manufacturer
had to have no noncompliant equipment and at least one equipment
item pending. e For inclusion in this category, VHA had to have no
compliance information from the manufacturer.
Source: VHA. We did not independently verify these data.
To identify specific biomedical equipment in the inventories of
VHA's medical facilities that still require Y2K compliance status
information from manufacturers, VHA's Chief Network Officer sent a
letter to the directors of VHA's 22 Veterans Integrated Service
Networks (VISN). This letter requested that they (1) review VHA's
list of manufacturers that have yet to respond to VHA's request
for compliance information and compare it with a
list of manufacturers from which their medical facilities still
require compliance information and (2) indicate the equipment item
that the facility owns for each manufacturer. According to VHA's
Y2K project
director, as of mid- March with 135 of 147 medical reporting sites
47 biomedical equipment items involving 35 manufacturers were
identified as still requiring compliance status information. The
project director told us that VHA medical facilities have been
instructed to replace or eliminate
equipment in their inventories for which they do not know the
compliance status by June 30. According to VHA's February 1999
status report on medical devices, medical facilities estimated
that the total cost of renovations will be about $41 million.
Category Number of manufacturers Percentage of manufacturers
Compliant manufacturers a 816 55. 2 Noncompliant manufacturers b
126 8. 5 Conditional- compliant manufacturers c 156 10. 5 Pending
manufacturers d 29 2.0 Manufacturers merged or bought out 226 15.
3 Nonresponsive manufacturers e 126 8. 5
Total 1,479 100. 0
Page 11 GAO/T-AIMD-99-152
We have previously reported that most manufacturers citing
noncompliant products listed incorrect display of date and/ or
time as the Y2K problem. 9 According to VA, these cases do not
present a risk to patient safety because health care providers,
such as physicians and nurses, can work around the problem. Of
more serious concern are situations in which devices depend on
date calculations the results of which can be incorrect. One
manufacturer cited the example of a product used for planning
delivery of radiation treatment using a radioactive isotope as the
source. An error in
calculating the strength of the radiation source on the day of
treatment could result in a dose that is too high or too low,
which could have an adverse effect on the patient. Other examples
of equipment presenting risk to patient safety identified by
manufacturers to FDA include hemodialysis delivery systems;
therapeutic apheresis systems; 10 alpha- fetoprotein kits for
neural tube defects; 11 various types of medical imaging
equipment; and
systems that store, track, and recall images in chronological
order. To track the compliance status of its biomedical equipment,
VHA uses a monthly status report on medical devices based on
information provided by the VISNs. According to the February 1999
report, approximately 426,000 of 531,000 medical devices in VHA
medical facilities are Y2K
compliant. Of the remaining devices, 86, 452 were identified as
conditionalcompliant or were not assessed for Y2K compliance
because the manufacturers certified that the equipment contained
no software or
embedded chips, and 19, 073 were reported as being noncompliant.
Of the noncompliant devices identified, 15, 621 are to be
repaired, 1,582 are to be replaced, 757 are to be used as is, 255
are to be retired, and 858 are still awaiting a decision on the
remedy. According to VHA's Chief Biomedical Engineer, most of the
noncompliant devices identified incorrectly displayed date/ time.
As we reported last September, FDA was also trying to determine
the Y2K compliance status of biomedical equipment. 12 Its goal is
to provide a
9 Year 2000 Computing Crisis: Compliance Status of Many Biomedical
Equipment Items Still Unknown (GAO/AIMD-98-240, September 18,
1998). 10 Such equipment allows therapeutic apheresis, which
refers to the exchange or purification of blood plasma.
Therapeutic apheresis is recognized as a successful treatment for
more than 40 autoimmune diseases.
11 Devices that use computer calculations of gestational status to
help assess the risk of neural tube defects in the fetuses of
pregnant women. 12 GAO/AIMD-98-240, September 18, 1998.
Page 12 GAO/T-AIMD-99-152
comprehensive, centralized source of information on the Y2K
compliance status of biomedical equipment used in the United
States and to make this information publicly available on a web
site. At the time, however, FDA had
a disappointing response rate from manufacturers to its letter
requesting compliance information. And, while FDA made this
information available to the public, it was not detailed enough to
be useful. Specifically, FDA's list
of compliant equipment lacked information on particular make and
model. To provide more detailed information on the compliance
status of biomedical equipment, as well as to integrate more
detailed compliance information gathered by VHA, we recommended
that VA and the Department of Health and Human Services (HHS)
jointly develop a single data clearinghouse that provides such
information to all users. We said development of the clearinghouse
should involve representatives from the
health- care industry, such as the Department of Defense and the
Health Industry Manufacturers Association. We recommended that the
clearinghouse contain such information as (1) the compliance
status of all biomedical equipment by make and model and (2) the
identity of manufacturers that are no longer in business. We also
recommended that
VHA and FDA determine what actions should be taken regarding
biomedical equipment manufacturers that have not provided
compliance information. In response to our recommendation, FDA in
conjunction with VHA has established the Federal Year 2000
Biomedical Equipment Clearinghouse.
With the assistance of VHA, the Department of Defense, and the
Health Industry Manufacturers Association, FDA has made progress
in obtaining compliance- status information from manufacturers.
For example, according to FDA, 4,251 biomedical equipment
manufacturers had
submitted data to the clearinghouse as of April 5, 1999. As shown
in figure 1, about 54 percent of the manufacturers reported having
products that do not employ a date, while about 16 percent
reported having date- related problems such as incorrect display
of date/ time. FDA is still awaiting responses from 399
manufacturers.
Page 13 GAO/T-AIMD-99-152
Figure 1: Biomedical Equipment Compliance- Status Information
Reported to FDA by Manufacturers as of April 5, 1999
Note: Total number of manufacturers = 4,251. Source: FDA.
FDA has also expanded the information in the clearinghouse. For
example, users can now find information on manufacturers that have
merged with or have been bought out by other firms. In
collaboration with the National Patient Safety Partnership, 13 FDA
is in the process of obtaining more detailed information from
manufacturers on noncompliant products, such
as make and model and descriptions of the impact of the Y2K
problem on products left uncorrected. For example, FDA sent a
letter dated March 29, 1999, to medical device manufacturers
requesting that they submit to the
clearinghouse complete lists of individual product models that are
Y2K compliant.
13 The National Patient Safety Partnership is a coalition of
public and private health care providers, including VA, the
American Medical Association, the American Hospital Association,
the American Nurses Association, and the Joint Commission on
Accreditation of Healthcare Organizations.
2,299 880
669 403
0 500
1,000 1,500
2,000 2,500
Products do not employ a date
All products employing date are compliant
Products with date related problems
Product status reported in manufacturer's web site
Page 14 GAO/T-AIMD-99-152
Review of Biomedical Equipment Test Results Lacking
We reported last September that VHA and FDA relied on
manufacturers to validate, test, and certify that equipment is Y2K
compliant. 14 We also reported that there was no assurance that
the manufacturers adequately addressed the Y2K problem for
noncompliant equipment because FDA did not require medical device
manufacturers to submit test results to it certifying compliance.
Accordingly, we recommended that VA and HHS take prudent steps to
jointly review manufacturers' compliance test results
for critical care/ life support biomedical equipment. We were
especially concerned that VA and FDA review test results for
equipment previously determined to be noncompliant but now deemed
by manufacturers to be compliant, or equipment for which concerns
about compliance remain. We also recommended that VA and HHS
determine what legislative, regulatory, or other changes were
necessary to obtain assurances that the manufacturers' equipment
was compliant, including performing independent verification and
validation of the manufacturers' certifications.
At the time, VA stated that it had no legislative or regulatory
authority to implement the recommendation to review test results
from manufacturers. In its response, HHS stated that it did not
concur with our recommendation to review test results supporting
medical device manufacturers' certifications that their equipment
is compliant. It believed that the submission of appropriate
certifications of compliance was sufficient to ensure that the
certifying manufacturers are in compliance. HHS also stated that
it did not have the resources to undertake such a review, yet we
are not aware of HHS' requesting resources from the Congress for
this purpose. More recently, VHA's Chief Biomedical Engineer told
us that VHA medical facilities are not requesting test results for
critical care/ life support biomedical equipment; they also are
not currently reviewing the test results available on
manufacturers' web sites. He said that VHA's priority is
determining the compliance status of its biomedical equipment
inventory and replacing noncompliant equipment. The director of
FDA's Division of Electronics and Computer Science likewise said
FDA sees no need to question manufacturers' certifications.
14 GAO/AIMD-98-240, September 18, 1998.
Page 15 GAO/T-AIMD-99-152
In contrast to VHA's and FDA's positions, some hospitals in the
private sector believe that testing biomedical equipment is
necessary to prove that they have exercised due diligence in the
protection of patient health and safety. Officials at three
hospitals told us that their biomedical engineers established
their own test programs for biomedical equipment and, in many
cases, contacted the manufacturers for their test protocols.
Several of these engineers informed us that their testing
identified some
noncompliant equipment that the manufacturers had earlier
certified as compliant. According to these engineers, to date, the
equipment found to be noncompliant all had display problems and
was not critical care/ life
support equipment. We were told that equipment found to be
incorrectly certified as compliant included a cardiac
catheterization unit, a pulse oxymeter, medical imaging equipment,
and ultrasound equipment. VHA, FDA, and the Emergency Care
Research Institute 15 continue to believe that manufacturers are
best qualified to analyze embedded systems or software to
determine Y2K compliance. They further believe that manufacturers
are the ones with full access to all design and operating
parameters contained in the internal software or embedded chips in
the equipment. VHA believes that such testing can potentially
cause irreparable damage to expensive health care equipment,
causing it to lock up or otherwise cease functioning. Further, a
number of manufacturers also have recommended that users not
conduct verification and validation testing.
We continue to believe that rather than relying solely on
manufacturers' certifications, organizations such as VHA or FDA
can provide users of medical devices with a greater level of
confidence that the devices are Y2K compliant through independent
reviews of manufacturers' compliance test results. The question of
whether to independently verify and validate biomedical equipment
that manufacturers have certified as compliant is one that must be
addressed jointly by medical facilities' clinical staff,
biomedical engineers, and corporate management. The overriding
criterion
should be ensuring patient health and safety. 15 An international,
nonprofit health services research agency. This organization
believes that superficial testing of biomedical equipment by users
may provide false assurances, as well as create legal liability
exposure for health care institutions.
Page 16 GAO/T-AIMD-99-152
VHA Pharmaceutical Operations Also Face Y2K Risks
Another critical component to VA's ability to deliver health care
at the turn of the century is ensuring that the automated systems
supporting VHA's medical facility pharmacies and its consolidated
mail outpatient pharmacies (CMOP) are Y2K compliant. VHA reported
that in 1998 it filled about 72 million prescriptions for 3.4
million veterans, at an estimated cost of about $2 billion. About
half of the prescriptions were filled by the over 200 pharmacies
located in VA's medical centers, clinics, and nursing homes. These
pharmacies rely on the pharmaceutical applications in the Veterans
Health Information Systems and Technology Architecture (VISTA) for
(1) drug distribution and inventory management, (2) dispensing of
drugs to inpatients and outpatients, (3) patient medication
information, and (4) an electronic connection between the
pharmacies and the CMOPs. Y2K failures in these applications could
impair the pharmacies' ability to fill prescriptions.
The remaining 50 percent of VHA's prescriptions are filled by
seven CMOPs, geographically located throughout the United States.
These facilities are supported by automated systems provided by
one of two contractors SI/ Baker, Inc. and Siemens ElectroCom. 16
For example, the CMOP electronically receives a prescription for a
veteran through the medical center. The prescription is downloaded
to highly automated dispensing
equipment to be filled. The filled prescription is then validated
by a pharmacist who compares the medication against a computerized
image of the prescribed medication. Afterward, the prescription is
packaged and an automatically generated mailing label is applied
for delivery to the veteran.
Finally, the medical center is electronically notified that the
prescription has been filled. Because of the reliance on
automation, the CMOPs' ability to fill prescriptions could be
delayed or interrupted if a Y2K failure occurred.
VHA has determined that the automated systems supporting its CMOPs
are not Y2K compliant. Specifically, neither of the systems
provided by their contractors is Y2K compliant. According to the
Y2K coordinator for the SI/ Baker facilities, failure to make the
SI/ Baker systems Y2K compliant may delay the filling of
outpatient prescriptions. The SI/ Baker systems are
used by three of VHA's CMOPs Hines, Illinois; Charleston, South
Carolina; and Murfreesboro, Tennessee; they handle about 58
percent of all prescriptions filled by CMOPs. In contrast to the
SI/ Baker systems, according to a contractor hired by the CMOPs
that use these systems, 16 These include operating systems,
databases, and pharmacy fulfillment application software.
Page 17 GAO/T-AIMD-99-152
failure to make the Siemens ElectroCom systems Y2K compliant may
result in delays in processing management reports for
prescriptions filled, but not in the actual filling of
prescriptions. Although the CMOPs plan to replace their
noncompliant systems with compliant ones, these systems are not
scheduled to be implemented until mid- to late- 1999. As shown in
table 3, the earliest estimated completion date for implementing a
compliant system is June 30, 1999, while the latest is December 1,
1999. 17 This leaves little time to address any unexpected
implementation problems. Table 3: Schedule of Estimated
Implementation Completion Dates and Current Daily Workload by
Consolidated Mail Outpatient Pharmacies
a Siemens ElectroCom automation. b SI/ Baker, Inc. automation.
Source: VA.
Given the late schedule for implementing compliant systems, it is
crucial that the CMOPs develop business continuity and contingency
plans to ensure that veterans will continue to receive their
medications if these
systems are not implemented in time or fail to operate properly.
As of March 31, VA had not completed a business continuity and
contingency plan for the CMOPs. The Y2K coordinator for the
Siemens ElectroCom
17 In April 15, 1999, testimony before the Subcommittee on
Oversight and Investigations, Committee on Veterans' Affairs,
House of Representatives, VHA's Y2K project director said that
estimated implementation dates would be revised so that the latest
date is August 30, 1999. Location Estimated completion
date Current daily workload (prescriptions filled)
West Los Angeles, California a June 30, 1999
15, 000 Bedford, Massachusetts a June 30, 1999 15, 000 Dallas,
Texas a June 30, 1999 14, 000 Leavenworth, Kansas a July 31, 1999
16, 000 Charleston, South Carolina b September 1, 1999
23, 000 Murfreesboro, Tennessee b September 30, 1999 38, 000
Hines, Illinois b December 1, 1999 21, 000
Page 18 GAO/T-AIMD-99-152
system has been tasked with developing this plan, which is to be
completed by the end of this month.
Further, VA did not include the CMOP systems in its quarterly
reports of mission- critical systems to OMB. According to VHA's
Y2K project director, VHA considered the CMOP systems to be COTS
products and, therefore, did not report them as mission- critical
systems. Given the criticality of these systems to VHA's ability
to fill prescriptions at the turn of the century, we believe VA
should reassess this decision. Reporting CMOPs as missioncritical
to VA top management and OMB would help ensure that necessary
attention is paid and action is taken.
VA Taking Action to Determine Y2K Readiness of Pharmaceutical and
Medical- Surgical Manufacturers
VA, like other users of pharmaceutical and medical- surgical
products, needs to know whether it will have a sufficient supply
of these items for its customers. Therefore, it has taken a
leadership role in the federal
government in determining whether manufacturers supplying these
products to VHA are Y2K- ready. This information is essential to
VHA's medical facilities and CMOPs because of their just- in- time
18 inventory
policy. Accordingly, they must know whether their manufacturers'
processes, which are highly automated, 19 are at risk, as well as
whether the rest of the supply chain will function properly.
To determine the Y2K readiness of their suppliers, on January 8,
1999, VA's National Acquisition Center (NAC) 20 sent a survey to
384 pharmaceutical firms and 459 medical- surgical firms with
which it does business. The survey contained questions on the
firms' overall Y2K status and inquired about actions taken to
assess, inventory, and plan for any perceived impact that the
century turnover would have on their ability to operate at normal
levels. In addition, the firms were asked to provide status
information on progress made to become Y2K compliant and a
reliable estimated date when compliance will be achieved for
business processes such as (1) ordering and receipt of raw
materials, (2) mixing and processing
product, (3) completing final product processing, (4) packaging
and 18 This term refers to maintaining a limited inventory on
hand. 19 Pharmaceutical manufacturers rely on automated systems
for production, packaging, and distribution of their products, as
well as for ordering of raw materials and supplies.
20 This organization is responsible for supporting VHA's health
care delivery system by providing an acquisition program for items
such as medical, dental, and surgical supplies and equipment;
pharmaceuticals; and chemicals. NAC is part of VA's Office of
Acquisition and Materiel Management.
Page 19 GAO/T-AIMD-99-152
labeling product, and (5) distributing finished product to
distributors/ wholesalers and end customers.
According to NAC officials, of the 455 firms that responded to the
survey as of March 31, 1999, about 55 percent completed all or
part of the survey. The remainder provided general information on
their Y2K readiness status or literature 21 on their efforts. As
shown in table 4, more than half (52
percent) of the pharmaceutical firms surveyed responded, with just
less than one- third (32 percent) of those respondents reporting
that they are compliant. Among the pharmaceutical firms that had
not responded as of March 31, however, were two of VA's five
largest suppliers. 22 The three large pharmaceutical suppliers
that did respond provided general information on their Y2K
readiness status, rather than answering the survey, and estimated
that they will be compliant by June 30, 1999.
Table 4: Status of Companies Surveyed by VHA as of March 31, 1999
a Estimated compliance status ranged from March 31, 1999 through
January 1, 2000; about 71 percent of pharmaceutical firms and 80
percent of medical- surgical firms estimated they will be
compliant by July 31, 1999. One firm responded that it will be
compliant by January 1, 2000.
Source: VA. We did not independently verify these data.
Table 4 also shows that 54 percent of the medical- surgical firms
surveyed responded, with about two- thirds (166) of them reporting
that they are Y2K compliant. All five of VA's largest medical-
surgical suppliers have responded. Specifically, two reported
being compliant, two reported they would be compliant by June 30,
1999, and the remaining supplier did not
report an expected compliance date. 21 This includes annual and
quarterly financial reports required by the Securities and
Exchange Commission for companies listed on the New York Stock
Exchange. 22 On April 14, 1999, a NAC official told us that of the
two suppliers that had not responded as of March 31, one responded
on April 12, and the other responded on April 14.
Responses Pharmaceutical Medical- surgical
Y2K compliant 65 166 Will be compliant by 1/ 1/ 2000 or earlier a
90 70 Provided no compliant date 50 14
Total number of responses 205 250
Non- responses 179 209
Total number of firms surveyed 384 459
Page 20 GAO/T-AIMD-99-152
On March 17, 1999, NAC sent a second letter to its pharmaceutical
and medical- surgical suppliers, informing them of VA's plans to
make Y2K readiness information previously provided to VA available
to the public through a web site ( www. va. gov/ oa& mm/ nac/
y2k). VA made the survey results available on its web site on
April 13, 1999. The letter also requested that manufacturers that
had not previously responded provide information on their
readiness. NAC's Executive Director said that he would personally
contact any major VA supplier that does not respond. On a broader
level, VHA has taken a leadership role in obtaining and sharing
information on the Y2K readiness of the pharmaceutical industry.
Specifically, VHA chairs the Year 2000 Pharmaceuticals
Acquisitions and Distributions Subcommittee, which reports to the
Chair of the President's Council on
Year 2000 Conversion. The purpose of this subcommittee is to bring
together federal and pharmaceutical representatives to address
issues concerning supply and distribution as they relate to the
year 2000. The
subcommittee consists of FDA, federal health care providers, and
industry trade associations such as the Pharmaceutical Research
and Manufacturers of America (PhRMA), Generic Pharmaceutical
Industry Association, the National Association of Chain Drug
Stores, National Wholesale Druggists' Association, and consumer
advocates. Several of these trade associations
have surveyed their members on their Y2K readiness and made the
results available to the public. However, the information is not
manufacturerspecific or as detailed as VHA's survey results.
FDA's Y2K Efforts for Pharmaceutical and Biological Products
Industries Were Initially Focused on Awareness
FDA's oversight and regulatory responsibility for pharmaceutical
and biological products 23 is to ensure that they are safe and
effective for public use. Because of its concern about the Y2K
impact on manufacturers of these products, FDA has taken several
actions to raise the Y2K awareness
of the pharmaceutical and biological products industries. In
addition, it is thinking about conducting a survey to determine
the industry's Y2K readiness. One of FDA's actions to raise
industry awareness was the January 1998 issuance of industry
guidance by the Center for Biologics Evaluation and Research
(CBER) on the Y2K impact of computer systems and software
applications used in the manufacture of blood products. In
addition, as shown in table 5, FDA has issued several letters to
pharmaceutical and biological trade associations and sole- source
drug manufacturers. 23 Biological products include vaccines,
blood, and blood products.
Page 21 GAO/T-AIMD-99-152
Table 5: FDA Letters to Manufacturers Regarding Y2K
Source: FDA.
Further, on February 11, 1999, FDA's director of emergency and
investigation operations sent a memorandum on FDA's interim
inspection policy for the Y2K problem to the directors of FDA's
investigations branch. The policy emphasizes FDA's Y2K awareness
efforts for manufacturers. It states that FDA inspectors are to
(1) inform the firm of FDA's Y2K web page
(URL http:// www. fda. gov/ cdrh/ yr2000/ year2000.html), (2)
provide the firm with copies of the appropriate FDA Y2K awareness
letter, (3) explain that Y2K problems could potentially affect
aspects of the firm's operations,
including some areas not regulated by FDA, and that FDA
anticipates that firms will take prudent steps to ensure that they
are not adversely affected by Y2K, and (4) provide firms with a
copy of FDA's compliance policy guide Year 2000 (Y2K) Computer
Problems.
In addition, FDA and PhRMA jointly held a government/ industry
forum on the Y2K preparedness of the pharmaceutical and biotech
industries on February 22, 1999. The objectives of this forum were
to (1) share information on Y2K programs conducted by health care
providers, pharmaceutical companies, FDA, and other federal
agencies, (2) provide a vehicle for networking, and (3) raise
awareness.
On March 29, 1999, FDA revised its February 11, 1999, interim
inspection policy. The revision states that field inspectors are
now to inquire about Date FDA source Recipient Purpose
October 1998 Center for
Drug Evaluation and Research Pharmaceutical
manufacturer trade associations To relay to members FDA's
expectation that the
pharmaceutical industry would (1) make resolution of Y2K a high
priority, (2) ensure that production systems were fixed and tested
prior to January 1, 2000, and (3) urge manufacturers to develop
Y2K contingency plans.
October 1998 Center for Biologics
Evaluation and Research Biologics manufacturer trade associations
Same as above. January 1999 Center for
Drug Evaluation and Research Sole- source drug manufacturers Same
as above. Also (1) noted that the impact of Y2K
on pharmaceutical safety, efficacy, and availability merits
special attention for firms who are the sole manufacturers of drug
components, bulk ingredients, and finished products and (2) stated
that pharmaceutical industry suppliers must have Y2Kcompliant
systems to protect against disruption in the flow of product
components, packaging materials, and
equipment to pharmaceutical manufacturers.
Page 22 GAO/T-AIMD-99-152
manufacturers' efforts to ensure that their computer- controlled
or datesensitive manufacturing processes and distribution systems
are Y2K compliant. Inspectors are to include this information in
their reports, along with a determination of activities that firms
have completed or started to ensure that they will be Y2K
compliant.
Further, FDA inspectors may review documentation in cases in which
firms have made changes to their computerized production or
manufacturing control systems to address Y2K problems. The purpose
of this review is to ensure that the changes were made in
accordance with the firms' procedures and applicable regulations.
If inspectors determine that a firm
has not taken steps to ensure Y2K compliance, they are to notify
their district managers and the responsible FDA center.
FDA's interim policy describes steps inspectors are to take in
reviewing manufacturers' Y2K compliance. However, FDA stated that
the primary focus of its inspections for the remainder of 1999
will be to ensure that products sold in the United States are safe
and effective for public use and comply with federal statutes and
regulations, including good manufacturing practice (GMP). 24 FDA
officials explained that the agency does not have sufficient
resources to perform both regulatory oversight of the
manufacturers and in- depth evaluations of firms' Y2K compliance
activities.
Nevertheless, according to the March 29, 1999, memorandum, field
inspectors are to note any concerns they may have with a firm's
Y2K readiness in the administrative remarks section of their
inspection reports. These reports are to be reviewed by FDA
district managers. If the Y2K concern appears to present a serious
problem to a firm's ability to produce safe, effective medication,
the district manager can discuss this issue with FDA's Office of
Regulatory Affairs and determine a course of action.
However, FDA officials have stressed that the agency cannot take
any regulatory action toward the firm until a Y2K- related problem
affects a pharmaceutical or biological product.
Like VHA, FDA is interested in the impact of Y2K readiness of
pharmaceutical and biological products on the availability of
products for health care facilities and individual patients. FDA's
Acting Deputy 24 GMP requirements include federal standards for
ensuring that products are high in quality and produced under
sanitary conditions (21 CFR parts 210, 211).
Page 23 GAO/T-AIMD-99-152
Commissioner for Policy informed us on March 24, 1999, that the
agency is thinking about surveying pharmaceutical and biological
products manufacturers, distributors, product repackagers, and
others in the drug dispensing chain, on their Y2K readiness and
contingency planning. In anticipation of a possible survey, the
agency has published a notice in the March 22, 1999, Federal
Register regarding this matter. The Acting Deputy Commissioner
said that potential survey questions on contingency planning would
include steps the manufacturers are taking to ensure an adequate
supply of bulk manufacturing materials from overseas suppliers.
This is a key issue because, as we reported in March 1998, 25
according to FDA, as much as 80 percent of the bulk pharmaceutical
chemicals used by
U. S. manufacturers to produce prescription drugs is imported. In
summary, VBA and VHA continue to make progress in preparing their
mission- critical systems for the year 2000. However, key actions
remain to be taken in the areas of mission- critical systems
testing, VHA facility systems compliance, and CMOP systems
compliance. We also reiterate the
need for VHA and FDA to take prudent steps to ensure that the test
results of critical care/ life support biomedical equipment are
obtained and reviewed. Finally, VHA needs information on the Y2K
readiness of specific pharmaceutical and medical- surgical
manufacturers. Until this information is obtained and publicized,
VHA medical facilities and veterans will remain in doubt as to
whether an adequate supply of pharmaceutical and biological
products will be available. FDA and the pharmaceutical and
biological trade associations can play key roles in helping VHA
obtain this
information and publicize the results in a single data
clearinghouse. In carrying out this assignment, we reviewed and
analyzed VA's Y2K documents and plans, comparing them against our
guidance on Y2K activities. We also reviewed and analyzed FDA
documentation relating to its Y2K efforts on biomedical devices
and pharmaceutical manufacturers. In addition, we visited selected
VHA medical centers, VA data centers, and
VHA consolidated mail outpatient pharmacies to discuss their Y2K
activities, and interviewed VA and FDA officials on those
activities. We also interviewed officials of the Emergency Care
Research Institute regarding their statements on biomedical
equipment testing. Finally, we interviewed selected private
hospital officials about their Y2K actions and
25 Food and Drug Administration: Improvements Needed in the
Foreign Drug Inspection Program (GAO/HEHS-98-21, March 17, 1998).
Page 24 GAO/T-AIMD-99-152
pharmaceutical trade associations on their Y2K readiness surveys
of pharmaceutical manufacturers. Mr. Chairman, this concludes my
statement. I would be pleased to respond to any questions that you
or other members of the Committee may have at this time.
(511748) Lett er
Ordering Information The first copy of each GAO report and
testimony is free. Additional copies are $2 each. Orders should be
sent to the following address, accompanied by a check or money
order made out to the Superintendent of Documents, when necessary,
VISA and
MasterCard credit cards are accepted, also. Orders for 100 or more
copies to be mailed to a single address are discounted 25 percent.
Orders by mail: U. S. General Accounting Office P. O. Box 37050
Washington, DC 20013
or visit: Room 1100 700 4th St. NW (corner of 4th and G Sts. NW)
U. S. General Accounting Office Washington, DC
Orders may also be placed by calling (202) 512- 6000 or by using
fax number (202) 512- 6061, or TDD (202) 512- 2537.
Each day, GAO issues a list of newly available reports and
testimony. To receive facsimile copies of the daily list or any
list from the past 30 days, please call (202) 512- 6000 using a
touchtone phone. A recorded menu will provide information on how
to obtain these lists.
For information on how to access GAO reports on the INTERNET, send
an e- mail message with info in the body to:
info@ www. gao. gov or visit GAO's World Wide Web Home Page at:
http:// www. gao. gov
United States General Accounting Office Washington, D. C. 20548-
0001
Official Business Penalty for Private Use $300
Address Correction Requested Bulk Mail
Postage & Fees Paid GAO Permit No. GI00
*** End of document. ***