[Federal Register Volume 90, Number 231 (Thursday, December 4, 2025)]
[Rules and Regulations]
[Pages 55978-55996]
From the Federal Register Online via the Government Publishing Office [www.gpo.gov]
[FR Doc No: 2025-21955]



[[Page 55977]]

Vol. 90

Thursday,

No. 231

December 4, 2025

Part II





Department of Health and Human





-----------------------------------------------------------------------





Food and Drug Administration





-----------------------------------------------------------------------





21 CFR Parts 801, 803, 812, et al.





Medical Devices; Quality Management System Regulation Technical 
Amendments; Final Rule

Federal Register / Vol. 90 , No. 231 / Thursday, December 4, 2025 / 
Rules and Regulations

[[Page 55978]]


-----------------------------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Parts 801, 803, 812, 860, 862, 864, 866, 868, 872, 874, 876, 
878, 880, 882, 886, 888, 890, and 892

[Docket No. FDA-2025-N-4635]


Medical Devices; Quality Management System Regulation Technical 
Amendments

AGENCY: Food and Drug Administration, HHS.

ACTION: Final rule; technical amendments.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA, the Agency, or we) is 
amending certain medical device regulations to revise references and 
language in existing Code of Federal Regulations (CFR) provisions to 
conform with the final rule ``Medical Devices; Quality System 
Regulation Amendments'' (QMSR Final Rule). This rule does not impose 
any new requirements on affected parties. This action is editorial in 
nature to correct errors, conform regulatory references, and ensure 
accuracy and clarity in the Agency's regulations.

DATES: This rule is effective February 2, 2026.

FOR FURTHER INFORMATION CONTACT: Daniel Schieffer, Office of Policy, 
Center for Devices and Radiological Health, 10903 New Hampshire Ave., 
Bldg. 66, Rm. 5562, Silver Spring, MD 20993, 301-796-3350, 
[email protected].

SUPPLEMENTARY INFORMATION:

I. Background

    As a part of this technical amendment, FDA is making changes to 21 
CFR parts 801, 803, 812, 860, 862, 864, 866, 868, 872, 874, 876, 878, 
880, 882, 886, 888, 890, and 892 to revise existing regulatory 
provisions for consistency with the QMSR Final Rule \1\ to correct 
errors, conform regulatory references, and ensure accuracy and clarity 
in the Agency's medical device regulations. The changes published in 
this notice are non-substantive and editorial in nature.
---------------------------------------------------------------------------

    \1\ In this technical amendment, FDA uses the terms below in the 
following manner: when referring to 21 CFR part 820 as amended and 
effective on February 2, 2026, FDA uses the terms ``Quality 
Management System Regulation'' or ``QMSR.'' When referring to the 
regulation at 21 CFR part 820 in effect before February 2, 2026, FDA 
uses the terms ``Quality System Regulation'' or ``QSR.''
---------------------------------------------------------------------------

    On February 2, 2024, FDA issued the QMSR Final Rule. This rule 
amended the device current good manufacturing practice (CGMP) 
requirements of the Quality System Regulation (QSR) to harmonize and 
modernize the regulation. The QMSR incorporates by reference an 
international standard for device quality management systems (ISO 
13485:2016, Medical devices--Quality management systems--Requirements 
for regulatory purposes) and establishes additional requirements and 
provisions that clarify certain expectations and concepts used in ISO 
13485. The purpose of this technical amendment is to update references 
from the QSR in existing FDA regulations to references to the QMSR that 
becomes effective on February 2, 2026, and to correct errors and ensure 
consistency and clarity in FDA's regulations.

II. Description of the Technical Amendments

    We are amending 179 sections of Title 21 of the CFR, spread 
throughout 18 parts, in this technical amendment. Each section that is 
being modified includes a change to conform existing regulations for 
consistency with the QMSR. FDA is also clarifying the authority 
citations for three parts. The changes made in this technical amendment 
can be summarized as follows:
     One hundred and sixty-two of these amendments change the 
wording in the classification regulations for certain class I device 
types.\2\ Specifically, these classification regulations currently 
exempt their respective devices, with various qualifications depending 
on the device, from the majority of the QSR. The QSR provisions that 
are not exempted in all of these sections (except one \3\) are 
Sec. Sec.  820.180 and 820.198 (21 CFR 820.180 and 820.198). This 
amendment updates the references to the non-exempted activities for 
these devices from Sec. Sec.  820.180 (Records--General requirements) 
and 820.198 (Records--Complaint files) in the QSR to the equivalent 
requirement in the QMSR (Sec.  820.35 (Control of Records) (21 CFR 
820.35)).
---------------------------------------------------------------------------

    \2\ Class I devices are those devices for which the general 
controls within the Federal Food, Drug, and Cosmetic Act (FD&C Act) 
(controls authorized by or under section 501, 502, 510, 516, 518, 
519, or 520 (21 U.S.C. 351, 352, 360, 360f, 360h, 360i, or 360j) or 
any combination of such sections) are sufficient to provide 
reasonable assurance of safety and effectiveness; or those devices 
for which insufficient information exists to determine that general 
controls are sufficient to provide reasonable assurance of safety 
and effectiveness or to establish special controls to provide such 
assurance, but because the devices are not purported or represented 
to be for a use in supporting or sustaining human life or for a use 
which is of substantial importance in preventing impairment of human 
health, and do not present a potential unreasonable risk of illness 
or injury, are to be regulated by general controls (section 
513(a)(1)(A) of the FD&C Act (21 U.S.C. 360c(a)(1)(A))).
    \3\ Section 864.3260(b) (21 CFR 864.3260(b)) currently exempts 
over-the-counter (OTC) test sample collection systems for drugs of 
abuse testing, if the device is not labeled or otherwise represented 
as sterile, from all of the QSR except for Sec.  820.198.
---------------------------------------------------------------------------

     Section 801.30(a)(2) (21 CFR 801.30(a)(2)), which is a 
specific exemption to the unique device identification requirement, is 
being modified in the same way as the classification regulations in the 
preceding bullet by changing references from Sec. Sec.  820.180 and 
820.198 in the QSR to the equivalent requirement in the QMSR (Sec.  
820.35).
     In the classification regulation for the Keratoscope 
(Sec.  886.1350(b) (21 CFR 886.1350(b))), in addition to the updated 
language described in the first bullet, we are also adding the words 
``Classification. Class I (general controls).'' This is because the 
statement that the device was classified as class I \4\ was 
inadvertently deleted from the CFR when the section was last edited in 
2000.\5\
---------------------------------------------------------------------------

    \4\ 55 FR 48436 at 48438.
    \5\ See 65 FR 2296 at 2320.
---------------------------------------------------------------------------

     Nine class II classification regulations \6\ are being 
modified to change references to Sec.  820.30 (Design Controls) (21 CFR 
820.30) in the QSR to the equivalent provision in the QMSR (Sec.  
820.10(c) (21 CFR 820.10(c))).
---------------------------------------------------------------------------

    \6\ Class II devices are those devices for which general 
controls by themselves are insufficient to provide reasonable 
assurance of safety and effectiveness, and for which there is 
sufficient information to establish special controls to provide such 
assurance, including the issuance of performance standards, 
postmarket surveillance, patient registries, development and 
dissemination of guidelines, recommendations, and other appropriate 
actions the Agency deems necessary to provide such assurance 
(section 513(a)(1)(B) of the FD&C Act).
---------------------------------------------------------------------------

     In part 812 (Investigational Device Exemptions) an 
additional two sections (Sec. Sec.  812.1(a) and 812.35(a)(3)(iii)(A) 
(21 CFR 812.1(a) and 812.35(a)(3)(iii)(A))) are being modified to also 
change references from Sec.  820.30 in the QSR to the equivalent 
provision in the QMSR (Sec.  820.10(c)). A second revision to Sec.  
812.35 revises Sec.  812.35(a)(3)(iv)(A) to insert a clarifying 
reference to Sec.  820.10(c) of the QMSR.
     One change in Sec.  801.45(e) (21 CFR 801.45(e)) changes a 
reference from the design history file in the QSR to the design and 
development files of the QMSR.
     Four sections are modified in this technical amendment by 
updating the phrase ``quality system regulation'' to

[[Page 55979]]

the phrase ``quality management system regulation.''
     FDA is clarifying the authority citations in parts 862, 
876, and 878 to italicize the `l' in 21 U.S.C. 360l to more clearly 
distinguish it from a numeral 1.
    In some of the edited sections we have also made minor wording 
changes to standardize language across sections and/or correct non-
substantive typographic errors. The changes to all 179 sections are 
non-substantive and are intended to conform existing regulations to the 
requirements in the QMSR that mirror the requirements in the QSR. This 
rule does not impose any new regulatory requirements on affected 
parties. The amendments are editorial in nature and should not be 
construed as modifying any substantive standards or requirements.

III. Notice and Public Comment

    Publication of this document constitutes final action under the 
Administrative Procedure Act (APA) (5 U.S.C. 551-559). Section 553 of 
the APA generally exempts ``rules of agency organization, procedure, or 
practice'' from proposed rulemaking (i.e., notice and comment 
rulemaking) (5 U.S.C. 553(b)(4)(A)). Rules are also exempt when an 
agency finds ``good cause'' that notice and comment rulemaking 
procedures would be ``impracticable, unnecessary, or contrary to the 
public interest'' (5 U.S.C. 553(b)(4)(B)).
    FDA has determined that this rulemaking meets the APA's notice and 
comment exemption requirements under 5 U.S.C. 553(b)(4)(B). All the 
revisions in this rule are technical or non-substantive changes. These 
revisions update the language in certain regulations to be consistent 
with other regulations and the FD&C Act without changing the actions 
required under the regulations. Such technical, non-substantive changes 
are ``a routine determination, insignificant in nature and impact, and 
inconsequential to the industry and to the public.'' Mack Trucks, Inc. 
v. EPA, 682 F.3d 87, 94 (D.C. Cir. 2012) (quotation marks and citation 
omitted). FDA accordingly for good cause finds that notice and public 
procedure thereon are unnecessary for these amendments.

IV. Effective Date

    This amendment is effective on February 2, 2026. This is the same 
date that the QMSR final rule becomes effective.

List of Subjects

21 CFR Part 801

    Labeling, Medical devices, Reporting and recordkeeping 
requirements.

21 CFR Part 803

    Imports, Medical devices, Reporting and recordkeeping requirements.

21 CFR Part 812

    Health records, Medical devices, Medical research, Reporting and 
recordkeeping requirements.

21 CFR Part 860

    Administrative practice and procedure, Medical devices.

21 CFR Part 862

    Medical devices.

21 CFR Part 864

    Blood, Medical devices, Packaging and containers.

21 CFR Part 866

    Biologics, Laboratories, Medical devices.

21 CFR Parts 868, 872, 874, 876, 878, 880, and 882

    Medical devices.

21 CFR Part 886

    Medical devices, Ophthalmic goods and services.

21 CFR Parts 888 and 890

    Medical devices.

21 CFR Part 892

    Medical devices, Radiation protection, X-rays.

    Therefore, under the Federal Food, Drug, and Cosmetic Act, and 
under the authority delegated to the Commissioner of Food and Drugs, 21 
CFR parts 801, 803, 812, 860, 862, 864, 866, 868, 872, 874, 876, 878, 
880, 882, 886, 888, 890, and 892 are amended as follows:

PART 801--LABELING

0
1. The authority citation for part 801 continues to read as follows:

    Authority:  21 U.S.C. 321, 331-334, 351, 352, 360d, 360i, 360j, 
371, 374.


0
2. In Sec.  801.30, revise paragraph (a)(2) to read as follows:


Sec.  801.30  General exceptions from the requirement for the label of 
a device to bear a unique device identifier.

    (a) * * *
    (2) A class I device that FDA has by regulation exempted from the 
good manufacturing practice requirements of part 820 of this chapter, 
exclusive of any continuing requirement for recordkeeping under Sec.  
820.35 of this chapter.
* * * * *

0
3. In Sec.  801.45, revise paragraph (e) to read as follows:


Sec.  801.45  Devices that must be directly marked with a unique device 
identifier.

* * * * *
    (e) Exception to be noted in design and development files. A 
labeler that decides to make use of an exception under paragraph (d) of 
this section must document the basis of that decision in the design and 
development files required by Sec.  820.10(c) of this chapter.

PART 803--MEDICAL DEVICE REPORTING

0
4. The authority citation for part 803 continues to read as follows:

    Authority:  21 U.S.C. 352, 360, 360i, 360j, 371, 374.


0
5. In Sec.  803.18, revise paragraph (e) to read as follows:


Sec.  803.18  What are the requirements for establishing and 
maintaining MDR files or records that apply to me?

* * * * *
    (e) If you are a manufacturer, you may maintain MDR event files as 
part of your complaint file, under part 820 of this chapter, if you 
prominently identify these records as MDR reportable events. We will 
not consider your submitted MDR report to comply with this part unless 
you evaluate an event in accordance with the quality management system 
requirements described in part 820 of this chapter. You must document 
and maintain in your MDR event files an explanation of why you did not 
submit or could not obtain any information required by this part, as 
well as the results of your evaluation of each event.

PART 812--INVESTIGATIONAL DEVICE EXEMPTIONS

0
6. The authority citation for part 812 continues to read as follows:

    Authority: 21 U.S.C. 331, 351, 352, 353, 355, 360, 360c-360f, 
360h-360j, 360hh-360pp, 360rr-360ss, 360bbb-8b, 371, 372, 374, 379e, 
381, 382; 42 U.S.C. 216, 241, 262.


0
7. In Sec.  812.1, revise paragraph (a) to read as follows:


Sec.  812.1  Scope.

    (a) The purpose of this part is to encourage, to the extent 
consistent with the protection of public health and safety and with 
ethical standards, the discovery and development of useful devices 
intended for human use, and to that end to maintain optimum freedom for 
scientific investigators in their pursuit of this purpose. This part 
provides procedures for the conduct of clinical investigations of 
devices. An approved investigational device exemption (IDE) permits a 
device that

[[Page 55980]]

otherwise would be required to comply with a performance standard or to 
have premarket approval to be shipped lawfully for the purpose of 
conducting investigations of that device. An IDE approved under Sec.  
812.30 or considered approved under Sec.  812.2(b) exempts a device 
from the requirements of the following sections of the Federal Food, 
Drug, and Cosmetic Act (the act) and the regulations in this chapter 
issued thereunder: Misbranding under section 502 of the act, 
registration, listing, and premarket notification under section 510, 
performance standards under section 514, premarket approval under 
section 515, a banned device regulation under section 516, records and 
reports under section 519, restricted device requirements under section 
520(e), good manufacturing practice requirements under section 520(f) 
except for the requirements found in Sec.  820.10(c), if applicable 
(unless the sponsor states an intention to comply with these 
requirements under Sec.  812.20(b)(3) or Sec.  812.140(b)(4)(v)) and 
color additive requirements under section 721.
* * * * *

0
8. In Sec.  812.35, revise paragraphs (a)(3)(iii)(A) and (a)(3)(iv)(A) 
to read as follows:


Sec.  812.35  Supplemental applications.

    (a) * * *
    (3) * * *
    (iii) * * *
    (A) Credible information to support developmental changes in the 
device (including manufacturing changes) includes data generated under 
the design and development activities of Sec.  820.10(c) of this 
chapter, preclinical/animal testing, peer reviewed published 
literature, or other reliable information such as clinical information 
gathered during a trial or marketing.
* * * * *
    (iv) * * *
    (A) For a developmental or manufacturing change to the device, the 
notice shall include a summary of the relevant information gathered 
during the course of the investigation upon which the change was based; 
a description of the change to the device or manufacturing process 
(cross-referenced to the appropriate sections of the original device 
description or manufacturing process); and, if the design and 
development activities of Sec.  820.10(c) of this chapter were used to 
assess the change, a statement that no new risks were identified by 
appropriate risk analysis and that the verification and validation 
testing, as appropriate, demonstrated that the design outputs met the 
design input requirements. If another method of assessment was used, 
the notice shall include a summary of the information which served as 
the credible information supporting the change.
* * * * *

PART 860--MEDICAL DEVICE CLASSIFICATION PROCEDURES

0
9. The authority citation for part 860 continues to read as follows:

    Authority: 21 U.S.C. 321(h), 353(g), 360c, 360d, 360e, 360i, 
360j, 371, 374.


0
10. In Sec.  860.15, revise paragraph (a) to read as follows:


Sec.  860.15  Exemptions from sections 510, 519, and 520(f) of the 
Federal Food, Drug, and Cosmetic Act.

    (a) A panel recommendation to the Commissioner that a device be 
classified or reclassified into class I will include a recommendation 
as to whether the device should be exempted from some or all of the 
requirements of one or more of the following sections of the Federal 
Food, Drug, and Cosmetic Act: Section 510 (registration, product 
listing, and premarket notification), section 519 (records and reports) 
and section 520(f) (good manufacturing practice requirements of the 
quality management system regulation), and, in the case of a 
recommendation for classification into class II, whether the device 
should be exempted from the premarket notification requirement under 
section 510.
* * * * *

0
11. In Sec.  860.84, revise paragraph (d)(4) to read as follows:


Sec.  860.84  Classification procedures for ``preamendments devices.''

* * * * *
    (d) * * *
    (4) In the case of a recommendation for classification into class 
I, a recommendation as to whether the device should be exempted from 
the requirements of one or more of the following sections of the 
Federal Food, Drug, and Cosmetic Act: Section 510 (registration, 
product listing, and premarket notification), section 519 (records and 
reports), and section 520(f) (good manufacturing practice requirements 
of the quality management system regulation) and, in the case of a 
recommendation for classification into class II, whether the device 
should be exempted from the premarket notification requirement under 
section 510, in accordance with Sec.  860.15;
* * * * *

0
12. In Sec.  860.240, revise paragraph (c)(2) to read as follows:


Sec.  860.240  Procedures for review of a De Novo request.

* * * * *
    (c) * * *
    (2) That implementation of Quality Management System Regulation 
(part 820 of this chapter) requirements, in addition to other general 
controls and any specified special controls, provide adequate assurance 
that critical and/or novel manufacturing processes produce devices that 
meet specifications necessary to ensure reasonable assurance of safety 
and effectiveness.

PART 862--CLINICAL CHEMISTRY AND CLINICAL TOXICOLOGY DEVICES

0
13. The authority citation for part 862 is revised to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
14. In Sec.  862.1220, revise paragraph (b)(2) to read as follows:


Sec.  862.1220  Acute kidney injury test system.

* * * * *
    (b) * * *
    (2) As part of the risk management activities performed as part of 
your 21 CFR 820.10(c) design and development activities, you must 
document the appropriate end user device training program provided in 
your premarket notification submission to satisfy the special control 
in paragraph (b)(1) of this section that will be offered while 
marketing the device as part of your efforts to mitigate the risk of 
incorrect interpretation of test results.
* * * * *

0
15. In Sec.  862.2050, revise paragraph (b) to read as follows:


Sec.  862.2050  General purpose laboratory equipment labeled or 
promoted for a specific medical use.

* * * * *
    (b) Classification. Class I (general controls). The device is 
identified in paragraph (a) of this section and is exempt from the 
premarket notification procedures in subpart E of part 807 of this 
chapter subject to the limitations in Sec.  862.9. The device is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
16. In Sec.  862.2270, revise paragraph (b) to read as follows:


Sec.  862.2270  Thin-layer chromatography system for clinical use.

* * * * *

[[Page 55981]]

    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  862.9. Particular components of 
TLC systems, i.e., the thin-layer chromatography apparatus, TLC 
atomizer, TLC developing tanks, and TLC ultraviolet light, are exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

PART 864--HEMATOLOGY AND PATHOLOGY DEVICES

0
17. The authority citation for part 864 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
18. In Sec.  864.1850, revise paragraph (b) to read as follows:


Sec.  864.1850  Dye and chemical solution stains.

* * * * *
    (b) Classification. Class I (general controls). These devices are 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  864.9. These 
devices are also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
19. In Sec.  864.2240, revise paragraph (b) to read as follows:


Sec.  864.2240  Cell and tissue culture supplies and equipment.

* * * * *
    (b) Classification. Class I (general controls). These devices are 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  864.9. If the 
devices are not labeled or otherwise represented as sterile, they are 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
20. In Sec.  864.3010, revise paragraph (b) to read as follows:


Sec.  864.3010  Tissue processing equipment.

* * * * *
    (b) Classification. Class I (general controls). These devices are 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  864.9. The 
devices are also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
21. In Sec.  864.3250, revise paragraph (b) to read as follows:


Sec.  864.3250  Specimen transport and storage container.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  864.9. If the device is not 
labeled or otherwise represented as sterile, it is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
22. In Sec.  864.3260, revise paragraph (b) to read as follows:


Sec.  864.3260  OTC test sample collection systems for drugs of abuse 
testing.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification requirements in part 807, 
subpart E of this chapter subject to the limitations in Sec.  864.9 if 
it is sold, distributed, and used in accordance with the restrictions 
set forth in Sec.  809.40 of this chapter. If the device is not labeled 
or otherwise represented as sterile, it is exempt from the current good 
manufacturing practice requirements of the quality management system 
regulation in part 820 of this chapter, except for requirements 
concerning complaint files under Sec.  820.35(a) of this chapter.

0
23. In Sec.  864.3600, revise paragraph (b) to read as follows:


Sec.  864.3600  Microscopes and accessories.

* * * * *
    (b) Classification. Class I (general controls). These devices are 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  864.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
24. In Sec.  864.4010, revise paragraph (b) to read as follows:


Sec.  864.4010  General purpose reagent.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  864.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
25. In Sec.  864.7010, revise paragraph (b)(3) to read as follows:


Sec.  864.7010   Flow cytometric test system for hematopoietic 
neoplasms.

* * * * *
    (b) * * *
    (3) As part of the risk management activities performed under 21 
CFR 820.10(c) design and development, product labeling and instruction 
manuals must include clear examples of all expected phenotypic patterns 
and gating strategies using well-defined clinical samples 
representative of both abnormal and normal cellular populations. These 
samples must be selected based upon the indications described in 
paragraph (b)(1)(i) of this section.

PART 866--IMMUNOLOGY AND MICROBIOLOGY DEVICES

0
26. The authority citation for part 866 continues to read as follows:

    Authority: 21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
27. In Sec.  866.2120, revise paragraph (b) to read as follows:


Sec.  866.2120  Anaerobic chamber.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  866.9. The 
device is also exempt from the good manufacturing practice requirements 
of the quality management system regulation in part 820 of this 
chapter, except for requirements concerning records and complaint files 
under Sec.  820.35 of this chapter.

0
28. In Sec.  866.2180, revise paragraph (b) to read as follows:

[[Page 55982]]

Sec.  866.2180  Manual colony counter.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  866.9. The 
device is also exempt from the good manufacturing practice requirements 
of the quality management system regulation in part 820 of this 
chapter, except for requirements concerning records and complaint files 
under Sec.  820.35 of this chapter.

0
29. In Sec.  866.2190, revise paragraph (b)(7) to read as follows:


Sec.  866.2190  Automated image assessment system for microbial 
colonies on solid culture media.

* * * * *
    (b) * * *
    (7) Under 21 CFR 820.10(c) design and development, device 
manufacturers must, as appropriate:
    (i) Conduct human factors/usability validation testing with the 
final version of the labeling and related materials to adequately 
mitigate the risk of failure to operate the instrument correctly.
    (ii) Document a device training program that will be offered to the 
end user to adequately mitigate the risk of failure to operate the 
instrument correctly.

0
30. In Sec.  866.2440, revise paragraph (b) to read as follows:


Sec.  866.2440  Automated medium dispensing and stacking device.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  866.9. The 
device is also exempt from the good manufacturing practice requirements 
of the quality management system regulation in part 820 of this 
chapter, except for requirements concerning records and complaint files 
under Sec.  820.35 of this chapter.

0
31. In Sec.  866.2540, revise paragraph (b) to read as follows:


Sec.  866.2540  Microbiological incubator.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  866.9. The 
device is also exempt from the good manufacturing practice requirements 
of the quality management system regulation in part 820 of this 
chapter, except for requirements concerning records and complaint files 
under Sec.  820.35 of this chapter.

0
32. In Sec.  866.2600, revise paragraph (b) to read as follows:


Sec.  866.2600  Wood's fluorescent lamp.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  866.9. The 
device is also exempt from the good manufacturing practice requirements 
of the quality management system regulation in part 820 of this 
chapter, except for requirements concerning records and complaint files 
under Sec.  820.35 of this chapter.

0
33. In Sec.  866.3215, revise paragraph (b)(6) to read as follows:


Sec.  866.3215  Device to detect and measure non-microbial analyte(s) 
in human clinical specimens to aid in assessment of patients with 
suspected sepsis.

* * * * *
    (b) * * *
    (6) As part of the risk management activities performed under 21 
CFR 820.10(c) design and development, you must document an appropriate 
end user device training program that will be offered as part of your 
efforts to mitigate the risk of failure to correctly operate the 
instrument.
* * * * *

0
34. In Sec.  866.3309, revise paragraph (b)(7) to read as follows:


Sec.  866.3309  Herpes virus nucleic acid-based cutaneous and 
mucocutaneous lesion panel.

* * * * *
    (b) * * *
    (7) The risk management activities performed as part of the 
manufacturer's 21 CFR 820.10(c) design and development activities must 
document an appropriate end user device training program that will be 
offered as part of efforts to mitigate the risk of failure to correctly 
operate the instrument.

0
35. In Sec.  866.3361, revise paragraph (b)(4) to read as follows:


Sec.  866.3361  Mass spectrometer system for clinical use for the 
identification of microorganisms.

* * * * *
    (b) * * *
    (4) As part of the risk management activities performed under 21 
CFR 820.10(c) design and development, you must document an appropriate 
end user device training program that will be offered as part of your 
efforts to mitigate the risk of failure to correctly operate the 
instrument.
* * * * *

0
36. In Sec.  866.3960, revise paragraph (b)(8) to read as follows:


Sec.  866.3960  Nucleic acid-based device for the amplification, 
detection, and identification of microbial pathogens directly from 
whole blood specimens.

* * * * *
    (b) * * *
    (8) As part of the risk management activities performed under 21 
CFR 820.10(c) design and development, you must document an appropriate 
end user device training program that will be offered as part of your 
efforts to mitigate the risk of failure to correctly operate the 
instrument.

0
37. In Sec.  866.3970, revise paragraph (b)(10) to read as follows:


Sec.  866.3970  Device to detect and identify microbial pathogen 
nucleic acids in cerebrospinal fluid.

* * * * *
    (b) * * *
    (10) As part of the risk management activities performed under 21 
CFR 820.10(c) design and development, you must document an appropriate 
end user device training program that will be offered as part of your 
efforts to mitigate the risk of failure to correctly operate the 
instrument.

PART 868--ANESTHESIOLOGY DEVICES

0
38. The authority citation for part 868 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
39. In Sec.  868.1030, revise paragraph (b) to read as follows:


Sec.  868.1030  Manual algesimeter.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
40. In Sec.  868.1965, revise paragraph (b) to read as follows:


Sec.  868.1965  Switching valve (ploss).

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. The 
device is also exempt from the current

[[Page 55983]]

good manufacturing practice requirements of the quality management 
system regulation in part 820 of this chapter, except for requirements 
concerning records and complaint files under Sec.  820.35 of this 
chapter.

0
41. In Sec.  868.5220, revise paragraph (b) to read as follows:


Sec.  868.5220  Blow bottle.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
42. In Sec.  868.5420, revise paragraph (b) to read as follows:


Sec.  868.5420  Ether hook.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
43. In Sec.  868.5760, revise paragraph (b) to read as follows:


Sec.  868.5760  Cuff spreader.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
44. In Sec.  868.5795, revise paragraph (b) to read as follows:


Sec.  868.5795  Tracheal tube cleaning brush.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
45. In Sec.  868.6175, revise paragraph (b) to read as follows:


Sec.  868.6175  Cardiopulmonary emergency cart.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
46. In Sec.  868.6225, revise paragraph (b) to read as follows:


Sec.  868.6225  Nose clip.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  868.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

PART 872--DENTAL DEVICES

0
47. The authority citation for part 872 continues to read as follows:

    Authority: 21 U.S.C. 351, 360 360c, 360e, 360j, 360l, 371.


0
48. In Sec.  872.1905, revise paragraph (b) to read as follows:


Sec.  872.1905  Dental x-ray film holder.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
49. In Sec.  872.3140, revise paragraph (b) to read as follows:


Sec.  872.3140   Resin applicator.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, the device 
is exempt from the current good manufacturing practice requirements of 
the quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
50. In Sec.  872.3150, revise paragraph (b) to read as follows:


Sec.  872.3150  Articulator.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, the device 
is exempt from the current good manufacturing practice requirements of 
the quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
51. In Sec.  872.3220, revise paragraph (b) to read as follows:


Sec.  872.3220  Facebow.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, the device 
is exempt from the current good manufacturing practice requirements of 
the quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
52. In Sec.  872.3670, revise paragraph (b) to read as follows:


Sec.  872.3670  Resin impression tray material.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If

[[Page 55984]]

the device is not labeled or otherwise represented as sterile, it is 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
53. In Sec.  872.3730, revise paragraph (b) to read as follows:


Sec.  872.3730  Pantograph.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
54. In Sec.  872.6010, revise paragraph (b) to read as follows:


Sec.  872.6010  Abrasive device and accessories.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
55. In Sec.  872.6050, revise paragraph (b) to read as follows:


Sec.  872.6050  Saliva absorber.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
56. In Sec.  872.6140, revise paragraph (b) to read as follows:


Sec.  872.6140  Articulation paper.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
57. In Sec.  872.6200, revise paragraph (b) to read as follows:


Sec.  872.6200  Base plate shellac.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
58. In Sec.  872.6290, revise paragraph (b) to read as follows:


Sec.  872.6290  Prophylaxis cup.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
59. In Sec.  872.6300, revise paragraph (b) to read as follows:


Sec.  872.6300  Rubber dam and accessories.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  872.9. If the device is not 
labeled or otherwise represented as sterile, it is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
60. In Sec.  872.6570, revise paragraph (b) to read as follows:


Sec.  872.6570  Impression tube.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
61. In Sec.  872.6650, revise paragraph (b) to read as follows:


Sec.  872.6650  Massaging pick or tip for oral hygiene.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
62. In Sec.  872.6670, revise paragraph (b) to read as follows:


Sec.  872.6670  Silicate protector.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
63. In Sec.  872.6855, revise paragraph (b) to read as follows:


Sec.  872.6855  Manual toothbrush.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements

[[Page 55985]]

concerning records and complaint files under Sec.  820.35 of this 
chapter.

0
64. In Sec.  872.6870, revise paragraph (b) to read as follows:


Sec.  872.6870  Disposable fluoride tray.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
65. In Sec.  872.6880, revise paragraph (b) to read as follows:


Sec.  872.6880  Preformed impression tray.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
66. In Sec.  872.6890, revise paragraph (b) to read as follows:


Sec.  872.6890  Intraoral dental wax.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  872.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

PART 874--EAR, NOSE, AND THROAT DEVICES

0
67. The authority citation for part 874 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
68. In Sec.  874.1500, revise paragraph (b) to read as follows:


Sec.  874.1500  Gustometer.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  874.9. If the device is not 
labeled or otherwise represented as sterile, it is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
69. In Sec.  874.3540, revise paragraph (b) to read as follows:


Sec.  874.3540  Prosthesis modification instrument for ossicular 
replacement surgery.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  874.9. If the device is not 
labeled or otherwise represented as sterile, it is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
70. In Sec.  874.5220, revise paragraph (b) to read as follows:


Sec.  874.5220  Ear, nose, and throat drug administration device.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  874.9. If the 
device is not labeled or otherwise represented as sterile, it is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

PART 876--GASTROENTEROLOGY-UROLOGY DEVICES

0
71. The authority citation for part 876 is revised to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
72. In Sec.  876.5210, revise paragraph (b) to read as follows:


Sec.  876.5210  Enema kit.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  876.9. The device is exempt from 
the current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
73. In Sec.  876.5250, revise paragraph (b)(2) to read as follows:


Sec.  876.5250  Urine collector and accessories.

* * * * *
    (b) * * *
    (2) Class I (general controls). For a urine collector and 
accessories not intended to be connected to an indwelling catheter, the 
device is exempt from the premarket notification procedures in subpart 
E of part 807 of this chapter subject to the limitations in Sec.  
876.9. If the device is not labeled or otherwise represented as 
sterile, it is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
74. In Sec.  876.5920, revise paragraph (b) to read as follows:


Sec.  876.5920  Protective garment for incontinence.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  876.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
75. In Sec.  876.5970, revise paragraph (b) to read as follows:


Sec.  876.5970  Hernia support.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  876.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

[[Page 55986]]

PART 878--GENERAL AND PLASTIC SURGERY DEVICES

0
76. The authority citation for part 878 is revised to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
77. In Sec.  878.3800, revise paragraph (b) to read as follows:


Sec.  878.3800  External aesthetic restoration prosthesis.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  878.9. If the 
device is intended for use without an external prosthesis adhesive to 
fasten it to the body, the device is exempt from the current good 
manufacturing practice requirements of the quality management system 
regulation in part 820 of this chapter, except for requirements 
concerning records and complaint files under Sec.  820.35 of this 
chapter.

0
78. In Sec.  878.3910, revise paragraph (b) to read as follows:


Sec.  878.3910  Noninflatable extremity splint.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  878.9. If the device is not 
labeled or otherwise represented as sterile, it is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

PART 880--GENERAL HOSPITAL AND PERSONAL USE DEVICES

0
79. The authority citation for part 880 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
80. In Sec.  880.2700, revise paragraph (b) to read as follows:


Sec.  880.2700  Stand-on patient scale.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
81. In Sec.  880.2740, revise paragraph (b) to read as follows:


Sec.  880.2740  Surgical sponge scale.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
82. In Sec.  880.2930, revise paragraph (b) to read as follows:


Sec.  880.2930  Apgar timer.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
83. In Sec.  880.5075, revise paragraph (b) to read as follows:


Sec.  880.5075  Elastic bandage.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
84. In Sec.  880.5120, revise paragraph (b) to read as follows:


Sec.  880.5120  Manual adjustable hospital bed.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
85. In Sec.  880.5150, revise paragraph (b) to read as follows:


Sec.  880.5150  Nonpowered flotation therapy mattress.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
86. In Sec.  880.5160, revise paragraph (b) to read as follows:


Sec.  880.5160  Therapeutic medical binder.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
87. In Sec.  880.5270, revise paragraph (b) to read as follows:


Sec.  880.5270  Neonatal eye pad.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  880.9. If the device is not 
labeled or otherwise represented as sterile, it is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
88. In Sec.  880.5300, revise paragraph (b) to read as follows:


Sec.  880.5300  Medical absorbent fiber.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt

[[Page 55987]]

from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
89. In Sec.  880.5640, revise paragraph (b) to read as follows:


Sec.  880.5640  Lamb feeding nipple.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
90. In Sec.  880.5680, revise paragraph (b) to read as follows:


Sec.  880.5680  Pediatric position holder.

* * * * *
    (b) Classification. Class I (general controls). Except when the 
device is an infant positioner for prescription use in highly monitored 
settings or an infant sleep position holder, it is exempt from the 
premarket notification procedures in subpart E of part 807 of this 
chapter subject to the limitations in Sec.  880.9. The device is exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
91. In Sec.  880.5780, revise paragraph (b)(2) to read as follows:


Sec.  880.5780  Medical support stocking.

* * * * *
    (b) * * *
    (2) Classification. Class I. The device is exempt from the 
premarket notification procedures in subpart E of part 807 of this 
chapter, subject to the limitations in Sec.  880.9. The device is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
92. In Sec.  880.5820, revise paragraph (b) to read as follows:


Sec.  880.5820  Therapeutic scrotal support.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
93. In Sec.  880.6025, revise paragraph (b) to read as follows:


Sec.  880.6025  Absorbent tipped applicator.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
94. In Sec.  880.6050, revise paragraph (b) to read as follows:


Sec.  880.6050  Ice bag.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
95. In Sec.  880.6060, revise paragraph (b) to read as follows:


Sec.  880.6060  Medical disposable bedding.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
96. In Sec.  880.6070, revise paragraph (b) to read as follows:


Sec.  880.6070  Bed board.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
97. In Sec.  880.6080, revise paragraph (b) to read as follows:


Sec.  880.6080  Cardiopulmonary resuscitation board.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
98. In Sec.  880.6085, revise paragraph (b) to read as follows:


Sec.  880.6085  Hot/cold water bottle.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
99. In Sec.  880.6140, revise paragraph (b) to read as follows:


Sec.  880.6140  Medical chair and table.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

[[Page 55988]]


0
100. In Sec.  880.6185, revise paragraph (b) to read as follows:


Sec.  880.6185  Cast cover.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
101. In Sec.  880.6190, revise paragraph (b) to read as follows:


Sec.  880.6190  Mattress cover for medical purposes.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
102. In Sec.  880.6200, revise paragraph (b) to read as follows:


Sec.  880.6200  Ring cutter.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
103. In Sec.  880.6230, revise paragraph (b) to read as follows:


Sec.  880.6230  Tongue depressor.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
104. In Sec.  880.6265, revise paragraph (b) to read as follows:


Sec.  880.6265  Examination gown.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
105. In Sec.  880.6350, revise paragraph (b) to read as follows:


Sec.  880.6350  Battery-powered medical examination light.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
106. In Sec.  880.6430, revise paragraph (b) to read as follows:


Sec.  880.6430  Liquid medication dispenser.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
107. In Sec.  880.6450, revise paragraph (b) to read as follows:


Sec.  880.6450  Skin pressure protectors.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
108. In Sec.  880.6730, revise paragraph (b) to read as follows:


Sec.  880.6730  Body waste receptacle.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
109. In Sec.  880.6785, revise paragraph (b) to read as follows:


Sec.  880.6785  Manual patient transfer device.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
110. In Sec.  880.6800, revise paragraph (b) to read as follows:


Sec.  880.6800  Washers for body waste receptacles.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. The 
device also is exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
111. In Sec.  880.6900, revise paragraph (b) to read as follows:


Sec.  880.6900  Hand-carried stretcher.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter,

[[Page 55989]]

subject to the limitations in Sec.  880.9. The device is also exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
112. In Sec.  880.6960, revise paragraph (b) to read as follows:


Sec.  880.6960  Irrigating syringe.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
113. In Sec.  880.6980, revise paragraph (b) to read as follows:


Sec.  880.6980  Vein stabilizer.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  880.9. If the 
device is not labeled or otherwise represented as sterile, it is also 
exempt from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

PART 882--NEUROLOGICAL DEVICES

0
114. The authority citation for part 882 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
115. In Sec.  882.1200, revise paragraph (b) to read as follows:


Sec.  882.1200  Two-point discriminator.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  882.9. The device is also exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
116. In Sec.  882.1500, revise paragraph (b) to read as follows:


Sec.  882.1500  Esthesiometer.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  882.9. The device is also exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
117. In Sec.  882.1525, revise paragraph (b) to read as follows:


Sec.  882.1525  Tuning fork.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  882.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
118. In Sec.  882.1700, revise paragraph (b) to read as follows:


Sec.  882.1700  Percussor.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to the limitations in Sec.  882.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

PART 886--OPHTHALMIC DEVICES

0
119. The authority citation for part 886 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
120. In Sec.  886.1140, revise paragraph (b) to read as follows:


Sec.  886.1140  Ophthalmic chair.

* * * * *
    (b) Classification. Class I. The AC-powered device and the manual 
device are exempt from the premarket notification procedures in subpart 
E of part 807 of this chapter, subject to the limitations in Sec.  
886.9. The manual device is also exempt from the current good 
manufacturing practice requirements of the quality management system 
regulation in part 820 of this chapter, except for requirements 
concerning records and complaint files under Sec.  820.35 of this 
chapter.

0
121. In Sec.  886.1150, revise paragraph (b) to read as follows:


Sec.  886.1150  Visual acuity chart.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
122. In Sec.  886.1170, revise paragraph (b) to read as follows:


Sec.  886.1170  Color vision tester.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
123. In Sec.  886.1190, revise paragraph (b) to read as follows:


Sec.  886.1190  Distometer.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
124. In Sec.  886.1200, revise paragraph (b) to read as follows:


Sec.  886.1200  Optokinetic drum.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter,

[[Page 55990]]

subject to the limitations in Sec.  886.9. The device is also exempt 
from the current good manufacturing practice requirements of the 
quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
125. In Sec.  886.1320, revise paragraph (b) to read as follows:


Sec.  886.1320  Fornixscope.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.
0
126. In Sec.  886.1330, revise paragraph (b) to read as follows:


Sec.  886.1330  Amsler grid.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
127. In Sec.  886.1350, revise paragraph (b) to read as follows:


Sec.  886.1350  Keratoscope.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter subject to Sec.  886.9. The battery-powered device 
is exempt from the current good manufacturing practice requirements of 
the quality management system regulation in part 820 of this chapter, 
except for requirements concerning records and complaint files under 
Sec.  820.35 of this chapter.

0
128. In Sec.  886.1375, revise paragraph (b) to read as follows:


Sec.  886.1375  Bagolini lens.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
129. In Sec.  886.1380, revise paragraph (b) to read as follows:


Sec.  886.1380  Diagnostic condensing lens.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
130. In Sec.  886.1390, revise paragraph (b) to read as follows:


Sec.  886.1390  Flexible diagnostic Fresnel lens.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
131. In Sec.  886.1395, revise paragraph (b) to read as follows:


Sec.  886.1395  Diagnostic Hruby fundus lens.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
132. In Sec.  886.1400, revise paragraph (b) to read as follows:


Sec.  886.1400  Maddox lens.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
133. In Sec.  886.1415, revise paragraph (b) to read as follows:


Sec.  886.1415  Ophthalmic trial lens frame.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
134. In Sec.  886.1460, revise paragraph (b) to read as follows:


Sec.  886.1460  Stereopsis measuring instrument.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
135. In Sec.  886.1500, revise paragraph (b) to read as follows:


Sec.  886.1500  Headband mirror.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
136. In Sec.  886.1605, revise paragraph (b) to read as follows:


Sec.  886.1605  Perimeter.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The

[[Page 55991]]

device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
137. In Sec.  886.1650, revise paragraph (b) to read as follows:


Sec.  886.1650  Ophthalmic bar prism.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
138. In Sec.  886.1655, revise paragraph (b) to read as follows:


Sec.  886.1655  Ophthalmic Fresnel prism.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
139. In Sec.  886.1665, revise paragraph (b) to read as follows:


Sec.  886.1665  Ophthalmic rotary prism.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
140. In Sec.  886.1700, revise paragraph (b) to read as follows:


Sec.  886.1700  Pupillometer.

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the manual device are exempt from the premarket notification 
procedures in subpart E of part 807 of this chapter, subject to the 
limitations in Sec.  886.9. The manual device is also exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
141. In Sec.  886.1770, revise paragraph (b) to read as follows:


Sec.  886.1770  Manual refractor.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
142. In Sec.  886.1780, revise paragraph (b)(2) to read as follows:


Sec.  886.1780  Retinoscope.

* * * * *
    (b) * * *
    (2) Class I (general controls) for the battery-powered device. The 
class I battery-powered device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter 
subject to Sec.  886.9. The battery-powered device is exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
143. In Sec.  886.1790, revise paragraph (b) to read as follows:


Sec.  886.1790  Nearpoint ruler.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
144. In Sec.  886.1810, revise paragraph (b) to read as follows:


Sec.  886.1810  Tangent screen (campimeter).

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the battery-powered device are exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter, 
subject to the limitations in Sec.  886.9. The battery-powered device 
is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
145. In Sec.  886.1840, revise paragraph (b) to read as follows:


Sec.  886.1840  Simulatan (including crossed cylinder).

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
146. In Sec.  886.1860, revise paragraph (b) to read as follows:


Sec.  886.1860  Ophthalmic instrument stand.

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the battery-powered device are exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter, 
subject to the limitations in Sec.  886.9. The battery-powered device 
is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
147. In Sec.  886.1870, revise paragraph (b) to read as follows:


Sec.  886.1870  Stereoscope.

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the battery-powered device are exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter, 
subject to the limitations in Sec.  886.9. The battery-powered device 
is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
148. In Sec.  886.1880, revise paragraph (b) to read as follows:

[[Page 55992]]

Sec.  886.1880  Fusion and stereoscopic target.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
149. In Sec.  886.1905, revise paragraph (b) to read as follows:


Sec.  886.1905  Nystagmus tape.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
150. In Sec.  886.1910, revise paragraph (b) to read as follows:


Sec.  886.1910  Spectacle dissociation test system.

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the battery-powered device are exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter, 
subject to the limitations in Sec.  886.9. The battery-powered device 
is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
151. In Sec.  886.4230, revise paragraph (b) to read as follows:


Sec.  886.4230  Ophthalmic knife test drum.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
152. In Sec.  886.4445, revise paragraph (b) to read as follows:


Sec.  886.4445  Permanent magnet.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
153. In Sec.  886.4750, revise paragraph (b) to read as follows:


Sec.  886.4750  Ophthalmic eye shield.

* * * * *
    (b) Classification. Class I (general controls). When made only of 
plastic or aluminum, the device is exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter 
subject to Sec.  886.9. When made only of plastic or aluminum, the 
devices are exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
154. In Sec.  886.4770, revise paragraph (b) to read as follows:


Sec.  886.4770  Ophthalmic operating spectacles (loupes).

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
155. In Sec.  886.4855, revise paragraph (b) to read as follows:


Sec.  886.4855  Ophthalmic instrument table.

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the manual device are exempt from the premarket notification 
procedures in subpart E of part 807 of this chapter, subject to the 
limitations in Sec.  886.9. The manual device is also exempt from the 
current good manufacturing practice requirements of the quality 
management system regulation in part 820 of this chapter, except for 
requirements concerning records and complaint files under Sec.  820.35 
of this chapter.

0
156. In Sec.  886.5120, revise paragraph (b) to read as follows:


Sec.  886.5120  Low-power binocular loupe.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
157. In Sec.  886.5540, revise paragraph (b) to read as follows:


Sec.  886.5540  Low-vision magnifier.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
158. In Sec.  886.5600, revise paragraph (b) to read as follows:


Sec.  886.5600  Ptosis crutch.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
159. In Sec.  886.5800, revise paragraph (b) to read as follows:


Sec.  886.5800  Ophthalmic bar reader.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements

[[Page 55993]]

concerning records and complaint files under Sec.  820.35 of this 
chapter.

0
160. In Sec.  886.5810, revise paragraph (b) to read as follows:


Sec.  886.5810  Ophthalmic prism reader.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
161. In Sec.  886.5870, revise paragraph (b) to read as follows:


Sec.  886.5870  Low-vision telescope.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
162. In Sec.  886.5910, revise paragraph (b) to read as follows:


Sec.  886.5910   Image intensification vision aid.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  886.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
163. In Sec.  886.5915, revise paragraph (b) to read as follows:


Sec.  886.5915  Optical vision aid.

* * * * *
    (b) Classification. Class I (general controls). The AC-powered 
device and the battery-powered device are exempt from the premarket 
notification procedures in subpart E of part 807 of this chapter, 
subject to the limitations in Sec.  886.9. The battery-powered device 
is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

PART 888--ORTHOPEDIC DEVICES

0
164. The authority citation for part 888 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
165. In Sec.  888.5850, revise paragraph (b) to read as follows:


Sec.  888.5850  Non-powered orthopedic traction apparatus and 
accessories.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  888.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
166. In Sec.  888.5890, revise paragraph (b) to read as follows:


Sec.  888.5890  Non-invasive traction component.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  888.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
167. In Sec.  888.5940, revise paragraph (b) to read as follows:


Sec.  888.5940  Cast component.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  888.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
168. In Sec.  888.5980, revise paragraph (b) to read as follows:


Sec.  888.5980  Manual cast application and removal instrument.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  888.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

PART 890--PHYSICAL MEDICINE DEVICES

0
169. The authority citation for part 890 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
170. In Sec.  890.3025, revise paragraph (b) to read as follows:


Sec.  890.3025  Prosthetic and orthotic accessory.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
171. In Sec.  890.3075, revise paragraph (b) to read as follows:


Sec.  890.3075  Cane.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
172. In Sec.  890.3150, revise paragraph (b) to read as follows:


Sec.  890.3150  Crutch.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this

[[Page 55994]]

chapter, except for requirements concerning records and complaint files 
under Sec.  820.35 of this chapter.

0
173. In Sec.  890.3410, revise paragraph (b) to read as follows:


Sec.  890.3410  External limb orthotic component.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
174. In Sec.  890.3420, revise paragraph (b) to read as follows:


Sec.  890.3420  External limb prosthetic component.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
175. In Sec.  890.3475, revise paragraph (b) to read as follows:


Sec.  890.3475  Limb orthosis.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
176. In Sec.  890.3490, revise paragraph (b) to read as follows:


Sec.  890.3490  Truncal orthosis.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
177. In Sec.  890.3520, revise paragraph (b) to read as follows:


Sec.  890.3520  Plinth.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
178. In Sec.  890.3640, revise paragraph (b) to read as follows:


Sec.  890.3640  Arm sling.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
179. In Sec.  890.3665, revise paragraph (b) to read as follows:


Sec.  890.3665  Congenital hip dislocation abduction splint.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
180. In Sec.  890.3675, revise paragraph (b) to read as follows:


Sec.  890.3675  Denis Brown splint.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
181. In Sec.  890.3700, revise paragraph (b) to read as follows:


Sec.  890.3700  Nonpowered communication system.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
182. In Sec.  890.3790, revise paragraph (b) to read as follows:


Sec.  890.3790  Cane, crutch, and walker tips and pads.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
183. In Sec.  890.3825, revise paragraph (b) to read as follows:


Sec.  890.3825  Mechanical walker.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
184. In Sec.  890.3910, revise paragraph (b) to read as follows:


Sec.  890.3910  Wheelchair accessory.

* * * * *
    (b) Classification. Class I (general controls). If the device is 
not intended for use as a protective restraint as defined in Sec.  
880.6760 of this chapter, it is exempt from the premarket notification 
procedures in subpart E of part 807 of this chapter, subject to the

[[Page 55995]]

limitations in Sec.  890.9. The device is also exempt from the current 
good manufacturing practice requirements of the quality management 
system regulation in part 820 of this chapter, except for requirements 
concerning records and complaint files under Sec.  820.35 of this 
chapter.

0
185. In Sec.  890.3940, revise paragraph (b) to read as follows:


Sec.  890.3940  Wheelchair platform scale.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
186. In Sec.  890.5050, revise paragraph (b) to read as follows:


Sec.  890.5050  Daily activity assist device.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. If the 
device is not labeled or otherwise represented as sterile, the device 
is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
187. In Sec.  890.5125, revise paragraph (b) to read as follows:


Sec.  890.5125  Nonpowered sitz bath.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
188. In Sec.  890.5350, revise paragraph (b) to read as follows:


Sec.  890.5350  Exercise component.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
189. In Sec.  890.5370, revise paragraph (b) to read as follows:


Sec.  890.5370  Nonmeasuring exercise equipment.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
190. In Sec.  890.5700, revise paragraph (b) to read as follows:


Sec.  890.5700  Cold pack.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807. The device also is exempt from the current good manufacturing 
practice requirements of the quality management system regulation in 
part 820, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
191. In Sec.  890.5730, revise paragraph (b) to read as follows:


Sec.  890.5730  Moist heat pack.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
192. In Sec.  890.5925, revise paragraph (b) to read as follows:


Sec.  890.5925  Traction accessory.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  890.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

PART 892--RADIOLOGY DEVICES

0
193. The authority citation for part 892 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
194. In Sec.  892.1920, revise paragraph (b) to read as follows:


Sec.  892.1920  Radiographic head holder.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  892.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
195. In Sec.  892.1940, revise paragraph (b) to read as follows:


Sec.  892.1940  Radiologic quality assurance instrument.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  892.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

0
196. In Sec.  892.1950, revise paragraph (b) to read as follows:


Sec.  892.1950  Radiographic anthropomorphic phantom.

* * * * *
    (b) Classification. Class I (general controls). The device is 
exempt from the premarket notification procedures in subpart E of part 
807 of this chapter, subject to the limitations in Sec.  892.9. The 
device is also exempt from the current good manufacturing practice 
requirements of the quality management system regulation in part 820 of 
this chapter, except for requirements concerning records and complaint 
files under Sec.  820.35 of this chapter.

[[Page 55996]]


0
197. In Sec.  892.5725, revise paragraph (b)(2) to read as follows:


Sec.  892.5725  Absorbable perirectal spacer.

* * * * *
    (b) * * *
    (2) The risk management activities performed as part of the 
manufacturer's Sec.  820.10(c) of this chapter design and development 
activities must document an appropriate end user initial training 
program which will be offered as part of efforts to mitigate the risk 
of failure to correctly operate the device, including, but not limited 
to, documentation of an appropriate end user initial training program 
on the proper spacer deployment technique.
* * * * *

Lowell M. Zeta,
Acting Deputy Commissioner for Policy, Legislation, and International 
Affairs.
[FR Doc. 2025-21955 Filed 12-3-25; 8:45 am]
BILLING CODE 4164-01-P