DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

21 CFR Part 892

[Docket No. FDA-2025-N-1182]


Medical Devices; Radiology Devices; Classification of the 
Radiological Acquisition and/or Optimization Guidance System

AGENCY: Food and Drug Administration, HHS.

ACTION: Final amendment; final order.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA, the Agency, or we) is 
classifying the radiological acquisition and/or optimization guidance 
system into class II (special controls). The special controls that 
apply to the device type are identified in this order and will be part 
of the codified language for the radiological acquisition and/or 
optimization guidance system's classification. We are taking this 
action because we have determined that classifying the device into 
class II (special controls) will provide a reasonable assurance of 
safety and effectiveness of the device. We believe this action will 
also enhance patients' access to beneficial innovative devices in part 
by reducing regulatory burdens.

DATES: This order is effective June 2, 2025. The classification was 
applicable on February 7, 2020.

FOR FURTHER INFORMATION CONTACT: Shahram Vaezy, Center for Devices and 
Radiological Health, Food and Drug Administration, 10903 New Hampshire 
Ave., Bldg. 66, Rm. 3656, Silver Spring, MD 20993-0002, 301-796-6242, 
[email protected].

SUPPLEMENTARY INFORMATION: 

I. Background

    Upon request, FDA has classified the radiological acquisition and/
or optimization guidance system as class II (special controls), which 
we have determined will provide a reasonable assurance of safety and 
effectiveness. In addition, we believe this action will enhance 
patients' access to beneficial innovation, in part by reducing 
regulatory burdens by placing the device into a lower device class than 
the automatic class III assignment.
    The automatic assignment of class III occurs by operation of law 
and without any action by FDA, regardless of the level of risk posed by 
the new device. Any device that was not in commercial distribution 
before May 28, 1976, is automatically classified as, and remains 
within, class III and requires premarket approval unless and until FDA 
takes an action to classify or reclassify the device (see 21 U.S.C. 
360c(f)(1)). We refer to these devices as ``postamendments devices'' 
because they were not in commercial distribution prior to the date of 
enactment of the Medical Device Amendments of 1976, which amended the 
Federal Food, Drug, and Cosmetic Act (FD&C Act).
    FDA may take a variety of actions in appropriate circumstances to 
classify or reclassify a device into class I or II. We may issue an 
order finding a new device to be substantially equivalent under section 
513(i) of the FD&C Act (see 21 U.S.C. 360c(i)) to a predicate device 
that does not require premarket approval. We determine whether a new 
device is substantially equivalent to a predicate device by means of 
the procedures for premarket notification under section 510(k) of the 
FD&C Act (21 U.S.C. 360(k)) and part 807 (21 CFR part 807).
    FDA may also classify a device through ``De Novo'' classification, 
a common name for the process authorized under section 513(f)(2) of the

[[Page 23284]]

FD&C Act (see also part 860, subpart D (21 CFR part 860, subpart D)). 
Section 207 of the Food and Drug Administration Modernization Act of 
1997 (Pub. L. 105-115) established the first procedure for De Novo 
classification. Section 607 of the Food and Drug Administration Safety 
and Innovation Act (Pub. L. 112-144) modified the De Novo application 
process by adding a second procedure. A device sponsor may utilize 
either procedure for De Novo classification.
    Under the first procedure, the person submits a 510(k) for a device 
that has not previously been classified. After receiving an order from 
FDA classifying the device into class III under section 513(f)(1) of 
the FD&C Act, the person then requests a classification under section 
513(f)(2).
    Under the second procedure, rather than first submitting a 510(k) 
and then a request for classification, if the person determines that 
there is no legally marketed device upon which to base a determination 
of substantial equivalence, that person requests a classification under 
section 513(f)(2) of the FD&C Act.
    Under either procedure for De Novo classification, FDA is required 
to classify the device by written order within 120 days. The 
classification will be according to the criteria under section 
513(a)(1) of the FD&C Act. Although the device was automatically placed 
within class III, the De Novo classification is considered to be the 
initial classification of the device.
    We believe this De Novo classification will enhance patients' 
access to beneficial innovation, in part by reducing regulatory 
burdens. When FDA classifies a device into class I or II via the De 
Novo process, the device can serve as a predicate for future devices of 
that type, including for 510(k)s (see section 513(f)(2)(B)(i) of the 
FD&C Act). As a result, other device sponsors do not have to submit a 
De Novo request or premarket approval application to market a 
substantially equivalent device (see section 513(i) of the FD&C Act, 
defining ``substantial equivalence''). Instead, sponsors can use the 
less-burdensome 510(k) process, when necessary, to market their device.

II. De Novo Classification

    On August 27, 2019, FDA received Caption Health, Inc.'s request for 
De Novo classification of the Caption Guidance. FDA reviewed the 
request in order to classify the device under the criteria for 
classification set forth in section 513(a)(1) of the FD&C Act.
    We classify devices into class II if general controls by themselves 
are insufficient to provide reasonable assurance of safety and 
effectiveness, but there is sufficient information to establish special 
controls that, in combination with the general controls, provide 
reasonable assurance of the safety and effectiveness of the device for 
its intended use (see 21 U.S.C. 360c(a)(1)(B)). After review of the 
information submitted in the request, we determined that the device can 
be classified into class II with the establishment of special controls. 
FDA has determined that these special controls, in addition to the 
general controls, will provide reasonable assurance of the safety and 
effectiveness of the device.
    Therefore, on February 7, 2020, FDA issued an order to the 
requester classifying the device into class II. In this final order, 
FDA is codifying the classification of the device by adding 21 CFR 
892.2100.\1\ We have named the generic type of device radiological 
acquisition and/or optimization guidance system, and it is identified 
as a device that is intended to aid in the acquisition and/or 
optimization of images and/or diagnostic signals. The device interfaces 
with the acquisition system, analyzes its output, and provides guidance 
and/or feedback to the operator for improving image and/or signal 
quality.
---------------------------------------------------------------------------

    \1\ FDA notes that the ACTION caption for this final order is 
styled as ``Final amendment; final order,'' rather than ``Final 
order.'' Beginning in December 2019, this editorial change was made 
to indicate that the document ``amends'' the Code of Federal 
Regulations. The change was made in accordance with the Office of 
Federal Register's (OFR) interpretations of the Federal Register Act 
(44 U.S.C. chapter 15), its implementing regulations (1 CFR 5.9 and 
parts 21 and 22), and the Document Drafting Handbook.
---------------------------------------------------------------------------

    FDA has identified the following risks to health associated 
specifically with this type of device and the measures required to 
mitigate these risks in table 1.

       Table 1--Radiological Acquisition and/or Optimization Guidance System Risks and Mitigation Measures
----------------------------------------------------------------------------------------------------------------
                          Identified risks to health                                  Mitigation measures
----------------------------------------------------------------------------------------------------------------
Device Error--Failure to provide guidance on acquiring diagnostic-quality      Design verification and
 images or signals, leading to delay, prolonged examination, or additional      validation, and Labeling.
 unnecessary procedures, due to:
     Algorithm failure.
     Hardware or software failure.
User Error--Operator failure to follow the guidance provided by the device to  Design verification and
 acquire diagnostic-quality images or signals, leading to delay, prolonged      validation, and Labeling.
 examination, or additional unnecessary procedures, due to human error.
----------------------------------------------------------------------------------------------------------------

    FDA has determined that special controls, in combination with the 
general controls, address these risks to health and provide reasonable 
assurance of safety and effectiveness. For a device to fall within this 
classification, and thus avoid automatic classification in class III, 
it would have to comply with the special controls named in this final 
order. The necessary special controls appear in the regulation codified 
by this order. This device is subject to premarket notification 
requirements under section 510(k) of the FD&C Act.

III. Analysis of Environmental Impact

    The Agency has determined under 21 CFR 25.34(b) that this action is 
of a type that does not individually or cumulatively have a significant 
effect on the human environment. Therefore, neither an environmental 
assessment nor an environmental impact statement is required.

IV. Paperwork Reduction Act of 1995

    This final order establishes special controls that refer to 
previously approved collections of information found in other FDA 
regulations and guidance. These collections of information are subject 
to review by the Office of Management and Budget (OMB) under the 
Paperwork Reduction Act of 1995 (44 U.S.C. 3501-3521). The collections 
of information in part 860, subpart D, regarding De Novo classification 
have been approved under OMB control number 0910-0844; the collections 
of information in 21 CFR part 814, subparts A through E, regarding 
premarket approval, have been approved under OMB control

[[Page 23285]]

number 0910-0231; the collections of information in part 807, subpart 
E, regarding premarket notification submissions, have been approved 
under OMB control number 0910-0120; the collections of information in 
21 CFR part 820, regarding quality system regulation, have been 
approved under OMB control number 0910-0073; and the collections of 
information in 21 CFR parts 801 and 809, regarding labeling, have been 
approved under OMB control number 0910-0485.

List of Subjects in 21 CFR Part 892

    Medical devices, Radiation protection, X-rays.

    Therefore, under the Federal Food, Drug, and Cosmetic Act and under 
authority delegated to the Commissioner of Food and Drugs, 21 CFR part 
892 is amended as follows:

PART 892--RADIOLOGY DEVICES

0
1. The authority citation for part 892 continues to read as follows:

    Authority:  21 U.S.C. 351, 360, 360c, 360e, 360j, 360l, 371.


0
2. Add Sec.  892.2100 to subpart B to read as follows:


Sec.  892.2100  Radiological acquisition and/or optimization guidance 
system.

    (a) Identification. A radiological acquisition and/or optimization 
guidance system is a device that is intended to aid in the acquisition 
and/or optimization of images and/or diagnostic signals. The device 
interfaces with the acquisition system, analyzes its output, and 
provides guidance and/or feedback to the operator for improving image 
and/or signal quality.
    (b) Classification. Class II (special controls). The special 
controls for this device are:
    (1) Design verification and validation must include:
    (i) A detailed, technical device description, including a detailed 
description of the impact of any software and hardware on the device's 
functions, the associated capabilities and limitations of each part, 
and the associated inputs and outputs.
    (ii) A detailed, technical report on the non-clinical performance 
testing of the subject device in the intended use environments, using 
relevant consensus standards when applicable.
    (iii) A detailed report on the clinical performance testing, 
obtained from either clinical testing, accepted virtual/physical 
systems designed to capture clinical variability, comparison to a 
closely-related device with established clinical performance, or other 
sources that are justified appropriately. The choice of the method must 
be justified given the risk of the device and the general acceptance of 
the test methods. The report must include the following:
    (A) A thorough description of the testing protocol(s).
    (B) A thorough, quantitative evaluation of the diagnostic utility 
and quality of images/data acquired, or optimized, using the device.
    (C) A thorough, quantitative evaluation of the performance in a 
representative user population and patient population, under 
anticipated conditions and environments of use.
    (D) A thorough discussion on the generalizability of the clinical 
performance testing results.
    (E) A thorough discussion on use-related risk analysis/human 
factors data.
    (iv) A detailed protocol that describes, in the event of a future 
change, the level of change in the device technical specifications or 
indications for use at which the change or changes could significantly 
affect the safety or effectiveness of the device and the risks posed by 
these changes. The assessment metrics, acceptance criteria, and 
analytical methods used for the performance testing of changes that are 
within the scope of the protocol must be included.
    (v) Documentation of an appropriate training program, including 
instructions on how to acquire and process quality images and video 
clips, and a report on usability testing demonstrating the 
effectiveness of that training program on user performance, including 
acquiring and processing quality images.
    (2) The labeling required under Sec.  801.109(c) of this chapter 
must include:
    (i) A detailed description of the device, including information on 
all required and/or compatible parts.
    (ii) A detailed description of the patient population for which the 
device is indicated for use.
    (iii) A detailed description of the intended user population, and 
the recommended user training.
    (iv) Detailed instructions for use, including the information 
provided in the training program used to meet the requirements of 
paragraph (b)(1)(iv) of this section.
    (v) A warning that the images and data acquired using the device 
are to be interpreted only by qualified medical professionals.
    (vi) A detailed summary of the reports required under paragraphs 
(b)(1)(ii) and (iii) of this section.
    (vii) A statement on upholding the As Low As Reasonably Achievable 
(ALARA) principle with a discussion on the associated device controls/
options.