DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
[Docket No. FDA-2010-D-0226]
Draft Guidance for Industry, Third Parties and Food and Drug
Administration Staff; Medical Device ISO 13485:2003 Voluntary Audit
Report Submission Program; Availability
AGENCY: Food and Drug Administration, HHS.
ACTION: Notice.
-----------------------------------------------------------------------
SUMMARY: The Food and Drug Administration (FDA) is announcing the
availability of the draft guidance entitled ``Medical Device ISO
13485:2003 Voluntary Audit Report Submission Program.'' This draft
guidance is intended to provide information on the implementation of a
section of the Food and Drug Administration Amendments Act of 2007
(FDAAA), which amends a section of the Federal Food, Drug, and Cosmetic
Act (the act). This guidance document describes how FDA's Center for
Devices and Radiological Health (CDRH) and Center for Biologics
Evaluation and Research (CBER) are implementing this provision of the
law. This draft guidance is not final nor is it in effect at this time.
DATES: Although you can comment on any guidance at any time (see 21
CFR 10.115(g)(5)), to ensure that the agency considers your comment on
this draft guidance before it begins work on the final version of the
guidance, submit
[[Page 28258]]
written or electronic comments on the draft guidance by August 18,
2010. Submit written or electronic comments on the collection of
information July 19, 2010.
ADDRESSES: Submit written requests for single copies of the draft
guidance document entitled ``Medical Device ISO 13485:2003 Voluntary
Audit Report Submission Program'' to the Division of Small
Manufacturers, International, and Consumer Assistance (DSMICA), Center
for Devices and Radiological Health, Food and Drug Administration,
10903 New Hampshire Ave., Bldg. 66, Room 4613, Silver Spring, MD 20993
or to the Office of Communications, Outreach and Development (HFM-40),
Center for Biologics Evaluation and Research, Food and Drug
Administration, 1401 Rockville Pike, suite 200N, Rockville, MD 20852-
1448. The draft guidance may also be obtained by mail by calling CBER
at 1-800-835-4709 or 301-827-1800. Send one self-addressed adhesive
label to assist that office in processing your request, or fax your
request to 301-847-8149. See the SUPPLEMENTARY INFORMATION section for
information on electronic access to the guidance.
Submit written comments concerning this draft guidance to the
Division of Dockets Management (HFA-305), Food and Drug Administration,
5630 Fishers Lane, rm. 1061, Rockville, MD 20852. Submit electronic
comments to http://www.regulations.gov. Identify comments with the
docket number found in brackets in the heading of this document.
FOR FURTHER INFORMATION CONTACT: Kimberly A. Trautman, Center for
Devices and Radiological Health, Food and Drug Administration, 10903
New Hampshire Ave., Bldg. 66, Room 3422, Silver Spring, MD 20993, 301-
796-5515, or Stephen Ripley, Center for Biologics Evaluation and
Research (HFM-17), Food and Drug Administration, 1401 Rockville Pike,
suite 200N, Rockville, MD 20852, 301-827-6210.
SUPPLEMENTARY INFORMATION:
I. Background
This draft guidance is intended to provide information on the
implementation of section 228 of FDAAA, which amends section 704(g)(7)
of the act (21 U.S.C. 374(g)(7)). Under this draft guidance, device
manufacturers whose establishment has been audited under one of the
regulatory systems implemented by the Global Harmonization Task Force
(GHTF) founding members\1\ using ISO 13485:2003 ``Medical devices--
Quality management systems--Requirements for regulatory purposes,'' may
voluntarily submit the resulting audit report to FDA. If, based on that
report, FDA determines there is minimal probability--in light of the
relationship between the quality system deficiencies observed and the
particular device and manufacturing processes involved--that the
establishment will produce nonconforming and/or defective finished
devices, then FDA intends to use the audit results as part of its risk
assessment to determine whether that establishment can be removed from
FDA's routine work plan for 1 year. The medical device ISO 13485:2003
Voluntary Audit Report Submission Program outlined in this draft
guidance is another way in which FDA may leverage audits performed by
other GHTF regulators and accredited third parties in order to assist
the agency in setting risk-based inspectional priorities.
---------------------------------------------------------------------------
\1\ The GHTF founding members auditing systems include the
Canadian Medical Devices Conformity Assessment System; the European
Union Notified Body accreditation system; the Therapeutics Goods
Administration of Australia Inspectorate; and the Japanese Medical
Device Ministry of Health, Labour and Welfare system.
---------------------------------------------------------------------------
This notice of availability and draft guidance satisfy the public
notice requirement of section 704(g)(7)(F) of the act, which provides
that FDA shall give public notice of the ISO standard(s) under which
FDA accepts voluntary submissions of audit reports.
II. Significance of Guidance
This draft guidance is being issued consistent with FDA's good
guidance practices regulation (21 CFR 10.115). The draft guidance, when
finalized, will represent the agency's current thinking on ``Medical
Device ISO 13485:2003 Voluntary Audit Report Submission Program.'' It
does not create or confer any rights for or on any person and does not
operate to bind FDA or the public. An alternative approach may be used
if such approach satisfies the requirements of the applicable statute
and regulations.
III. Electronic Access
Persons interested in obtaining a copy of the draft guidance may do
so by using the Internet. To receive ``Draft Guidance for Industry,
Third Parties and FDA Staff: Medical Device ISO 13485:2003 Voluntary
Audit Report Submission Program'' you may either send an e-mail request
to [email protected] to receive an electronic copy of the document or
send a fax request to 301-847-8149 to receive a hard copy. Please use
the document number 1705 to identify the guidance you are requesting.
A search capability for all CDRH guidance documents is available at
http://www.fda.gov/cdrh/guidance.html. Guidance documents are also
available at http://www.regulations.gov or at http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/Guidances/default.htm.
IV. Paperwork Reduction Act of 1995
Under the Paperwork Reduction Act of 1995 (PRA) (44 U.S.C. 3501-
3520), Federal agencies must obtain approval from the Office of
Management and Budget (OMB) for each collection of information they
conduct or sponsor. ``Collection of information'' is defined in 44
U.S.C. 3502(3) and 5 CFR 1320.3(c) and includes agency requests or
requirements that members of the public submit reports, keep records,
or provide information to a third party. Section 3506(c)(2)(A) of the
PRA (44 U.S.C. 3506(c)(2)(A)) requires Federal agencies to provide a
60-day notice in the Federal Register concerning each proposed
collection of information before submitting the collection to OMB for
approval. To comply with this requirement, FDA is publishing notice of
the proposed collection of information set forth in this document.
With respect to the following collection of information, FDA
invites comments on these topics: (1) Whether the proposed collection
of information is necessary for the proper performance of FDA's
functions, including whether the information will have practical
utility; (2) the accuracy of FDA's estimate of the burden of the
proposed collection of information, including the validity of the
methodology and assumptions used; (3) ways to enhance the quality,
utility, and clarity of the information to be collected; and (4) ways
to minimize the burden of the collection of information on respondents,
including through the use of automated collection techniques, when
appropriate, and other forms of information technology.
Title: Draft Guidance for Industry, Third Parties and FDA Staff:
Medical Device ISO 13485:2003 Voluntary Audit Report Submission
Program.
Description: Section 228 of the Food and Drug Administration
Amendments Act of 2007 (FDAAA), amended section 704(g)(7) of the Food,
Drug, and Cosmetic Act (the act) (21 U.S.C. 374(g)(7)) to add the
following provision:
``(F) For the purpose of setting risk-based inspectional
priorities, the Secretary shall accept voluntary
[[Page 28259]]
submissions of reports of audits assessing conformance with appropriate
quality system standards set by the International Organization for
Standardization (ISO) and identified by the Secretary in public notice.
If the owner or operator of an establishment elects to submit audit
reports under this subparagraph, the owner or operator shall submit all
such audit reports with respect to the establishment during the
preceding 2-year periods.''
The ``Draft Guidance for Industry, Third Parties and FDA Staff:
Medical Device ISO 13485:2003 Voluntary Audit Report Submission
Program'' will describe how FDA's CDRH and CBER are implementing this
provision of the law and providing public notice as required. The
proposed collections of information are necessary to satisfy the
previously mentioned statutory requirements for implementing this
voluntary submission program.
FDA estimates the burden of this collection of information as
follows:
Table 1.--Estimated Annual Reporting Burden\1\
----------------------------------------------------------------------------------------------------------------
No. of Annual Frequency Total Annual Hours per
Type of Respondent Respondents per Response Responses Response Total Hours
----------------------------------------------------------------------------------------------------------------
Domestic or foreign 1,600 1 1,600 2 3,200
device manufacturer
whose establishment
was audited under
ISO 13485:2003
----------------------------------------------------------------------------------------------------------------
\1\ There are no capital costs or operating and maintenance costs associated with this collection of
information.
Based on FDA's experience with the founding regulatory members of
GHTF, FDA expects that the vast majority of manufacturers who will
participate in the Voluntary Audit Report Submission Program will be
manufacturers who are certified by Health Canada under ISO
13485:2003.\2\ In 2008, approximately 2,650 manufacturers or
manufacturing sites had been certified by Health Canada.
---------------------------------------------------------------------------
\2\ The majority of these manufacturers are also certified under
ISO 13485:2003 by the European Union Notified Body accreditation
system.
---------------------------------------------------------------------------
In addition, FDA only expects firms who do not have major
deficiencies or observations in their ISO 13485:2003 audits to be
willing to submit their audit reports to FDA under the Voluntary Audit
Report Submission Program. FDA analyzed its inspection data from Fiscal
Year (FY) 2008 (October 1, 2007 to October 1, 2008) and determined that
the total number of inspections finalized in FY2008 for medical devices
was 1,965. The breakdown for the 1,965 compliance decisions is as
follows:
Table 2.--Compliance Decision Breakdown
----------------------------------------------------------------------------------------------------------------
Compliance Decision\1\ Number Approximate Percentage
----------------------------------------------------------------------------------------------------------------
Official Action Indicated 148 8%
----------------------------------------------------------------------------------------------------------------
Voluntary Action Indicated 775 40%
----------------------------------------------------------------------------------------------------------------
No Action Indicated 1,025 52%
----------------------------------------------------------------------------------------------------------------
Pending Final Decision 17 1%
----------------------------------------------------------------------------------------------------------------
\1\ June 15, 2006, Compliance Program 7382.845 Inspection of Medical Device Manufacturers Part V http://www.fda.gov/cdrh/comp/guidance/7382.845.html#p5p5.pdf.
Because FDA only expects firms who do not have major deficiencies
or observations to be willing to submit their audit reports to FDA
under the Voluntary Audit Report Submission Program, FDA only expects
to receive audit reports that would have been classified by FDA as No
Action Indicated (NAI).
Assuming that the percentage breakdown of compliance decisions for
all inspections conducted in FY2008 can be extrapolated and applied to
audits of manufacturers certified under ISO 13485:2003 by Health
Canada, FDA can estimate the number of Canadian establishments that
would have had an inspection classified as an NAI. Since 52 percent of
all compliance decisions resulted in a NAI decision, FDA estimates that
1,378 of the facilities certified under ISO 13485:2003 by Health Canada
(52 percent of the total 2,650 facilities) would have had an inspection
classified as an NAI. Since FDA only expects to receive audit reports
that would have been classified by FDA as NAI, FDA expects 1,378, or
approximately 1,400, audit reports to be submitted.
Since FDA expects that the vast majority of manufacturers who will
participate in the Voluntary Audit Report Submission Program will be
manufacturers certified by Health Canada under ISO 13485:2003, FDA
expects the number of reports to be submitted from manufacturers
certified by regulatory systems established by other founding GHTF
members to be minimal. For purposes of calculating the reporting
burden, FDA estimates that approximately 10 percent of total audit
reports submitted under this program will be from these other
manufacturers. Since 90 percent of the audit reports are expected to be
submitted by manufacturers certified by Health Canada (approximately
1,400 audit reports as calculated previously in this document), then
the total number of audit reports FDA expects to receive in a year is
1,556, or approximately 1,600, audit reports.
FDA further estimates that the gathering, scanning, and submission
of the audit reports, certificates, and related correspondence would
take approximately 2 hours.
This draft guidance also refers to previously approved collections
of information found in FDA regulations. These collections of
information are subject to review by OMB under the PRA. The collections
of information in 21 CFR part 820 have been approved
[[Page 28260]]
under OMB control number 0910-0073 and the collections of information
for the Inspection by Accredited Persons Program have been approved
under OMB control number 0910-0569.
V. Comments
Interested persons may submit to the Division of Dockets Management
(see ADDRESSES), written or electronic comments regarding this
document. Submit a single copy of electronic comments or two paper
copies of any mailed comments, except that individuals may submit one
paper copy. Comments are to be identified with the docket number found
in brackets in the heading of this document. Received comments may be
seen in the Division of Dockets Management between 9 a.m. and 4 p.m.,
Monday through Friday.
Dated: May 17, 2010.
Leslie Kux,
Acting Assistant Commissioner for Policy.
[FR Doc. 2010-12098 Filed 5-19-10; 8:45 am]