NUCLEAR REGULATORY COMMISSION


Report to Congress on Abnormal Occurrences Fiscal Year 2003; 
Dissemination of Information

    Section 208 of the Energy Reorganization Act of 1974 (Public Law 
93-438) defines an abnormal occurrence (AO) as an unscheduled incident 
or event which the U.S. Nuclear Regulatory Commission (NRC) determines 
to be significant from the standpoint of public health or safety. The 
Federal Reports Elimination and Sunset Act of 1995 (Public Law 104-66) 
requires that AOs be reported to Congress annually. During fiscal year 
2003, 14 events that occurred at facilities licensed or otherwise 
regulated by the NRC and/or Agreements States were determined to be 
AOs. The report describes five medical events at facilities licensed by 
the NRC. Three events involved patients undergoing therapeutic 
brachytherapy treatments, one event involved an unintentional 
therapeutic dose of sodium iodide (I-131) to an embryo/fetus, and one 
event involved a diagnostic overexposure of a minor. The report also 
discusses nine AOs at facilities licensed by Agreement States. 
Agreement States are those states which have entered into a formal 
agreement with the NRC pursuant to Section 274 of the Atomic Energy Act 
(AEA) to regulate certain quantities of AEA material at facilities 
located within their borders. Currently, there are 33 Agreement States. 
Seven Agreement State events were medical events (five therapeutic and 
two diagnostic), one Agreement State event involved overexposure to a 
radiographer, and one Agreement State event involved overexposure to 
members of the public from a damaged gauge. As required by Section 208, 
the discussion for each event includes the date and place, the nature 
and probable consequences, the cause or causes, and the action taken to 
prevent recurrence. Each event is also being described in NUREG-0090, 
Vol. 26, ``Report to Congress on Abnormal Occurrences, Fiscal Year 
2003.'' This report will be available electronically at the NRC Web 
site http://www.nrc.gov/reading-rm/doc-collections/nuregs/staff/ staff/.

Nuclear Power Plants

    During this period, no events occurred at U.S. nuclear power plants 
that were significant enough to be reported as AOs.
* * * * *

Fuel Cycle Facilities (Other Than Nuclear Power Plants)

    During this period, no events occurred at U.S. fuel cycle 
facilities that were significant enough to be reported as AOs.
* * * * *

Other NRC Licenses (Industrial Radiographers, Medical Institutions, 
etc.)

    The NRC determined that the following events which occurred at 
facilities, licensed or otherwise regulated by the NRC, during this 
reporting period were significant enough to be reported as AOs:

03-01 Intravascular Brachytherapy (IVB) Medical Event at the Queen's 
Medical Center in Honolulu, Hawaii

    Date and Place--October 9, 2002; the Queen's Medical Center; 
Honolulu, Hawaii.
    Nature and Probable Consequences--A patient undergoing IVB 
treatment for cardiacrestenosis received an underdose to the intended 
treatment site, but a dose above the AO criterion to an unintended 
site. This medical event occurred because the strontium-90 (Sr-90) 
source contained in the device's source train (catheter) did not reach 
the intended treatment site. The patient undergoing IVB was prescribed 
treatment of 18.4 Gray (Gy) (1,840 rads)

[[Page 24689]]

to the left anterior descending (LAD) artery to prevent scar tissue 
blockage. Sixteen Sr-90 seeds with a total activity of 2.224 
gigabecquerel (GBq) (60.11 millicuries [mCi]) were positioned in the 
patient using fluoroscopy. Because the radiation oncologist and 
cardiologist believed that they could see the proximal and distal 
markers of the source train on the fluoroscopy monitor, the physicist 
did not perform a survey to ensure that the source train was in the 
patient's chest.
    After the end of the treatment, the radiation oncologist was unable 
to retrieve all of the Sr-90 radioactive sources. After a second 
attempt to retrieve the sources failed, the oncologist pulled the 
treatment catheter from the patient and placed it in the bailout box. 
The bailout box is an acrylic box approximately 12 inches (in) by 10 in 
by 6 in with a hinged acrylic lid. Acrylic is used because of its 
shielding properties to attenuate the beta radiation from the catheter 
system. While inspecting the catheter, the oncologist discovered a kink 
at the location wherein the distal seed and marker became lodged. The 
kink was attributed to the patient's anatomy (small curves in the blood 
vessel, branching off the aorta where the catheter was inserted). A 
review of the cinematography images revealed that only one Sr-90 seed 
reached the intended treatment site while 5 seeds were positioned in 
the beginning LAD and 10 seeds were outside the cinematography field of 
view. Instead of receiving the intended 18.4 Gy (1,840 rads), the LAD 
received approximately 1.25 Gy (125 rads). The remaining dose was 
delivered to an unintended section of the LAD and aorta. No adverse 
effects due to this medical event are expected.
    Cause or Causes--This medical event was caused by human error as 
the licensee did not perform a survey to verify that the radioactive 
sources were in the proper location. The patient's anatomy was a 
contributing factor in that there were curves in a small blood vessel 
branching off the aorta.

Actions Taken To Prevent Recurrence

    Licensee--Based on the cause and contributing factors of the 
medical event, the licensee modified its procedures to require 
additional documented verification of the position of the markers by 
the radiological technologist and medical physicist in addition to the 
required verification by the radiation oncologist and cardiologist.
    NRC--On November 13, 2002, the NRC issued a Notice of Violation to 
the licensee for the failure to follow the manufacturer's operation 
procedures for the IVB device as specified in its license.
    This event is closed for the purpose of this report.
* * * * *
    03-02 Dose to Fetus at Community Hospital of Anderson in Anderson, 
Indiana

    Date and Place--August 8, 2003; Community Hospital; Anderson, 
Indiana.
    Nature and Probable Consequences--On August 8, 2003, the Community 
Hospital of Anderson reported that a 35-year-old female patient was 
administered 1.1 GBq (29.8 mCi) of sodium iodide-131 (I-131) for the 
treatment of hyperthyroidism. At the time of the therapy, the patient 
was unaware that she was pregnant and, as a result, an unintentional 
dose to her embryo/fetus was delivered. On August 25, 2003, the 
patient's gynecologist informed the hospital and the patient that she 
had been approximately 15 weeks pregnant at the time of the therapy.
    The NRC staff contracted with a medical consultant to review the 
possible deterministic effects of the dose to the embryo/fetus as a 
result of the event. The medical report indicated that the total 
effective dose equivalent (whole body) to the embryo/fetus was 
approximately 0.074 Gy (7.4 rads) and the committed dose equivalent to 
the embryo/fetal thyroid was approximately 278 Gy (27,800 rads). The 
licensee anticipated that the fetal thyroid would be ablated. The NRC 
medical consultant, contracted to review this event, also anticipated 
that the fetal thyroid would be ablated.
    Cause or Causes--The event appeared to be an isolated occurrence. 
The root cause of the event was determined to be human error. Although 
the authorized physician user and the chief technologist asked the 
patient on several occasions, prior to the administration of the I-131 
dosage, if she were pregnant or believed that she could possibly be 
pregnant, the patient denied the possibility of pregnancy. Due to other 
preexisting medical conditions and consultations by other physicians 
informing the patient that she was unable to conceive, the patient 
believed that she could not become pregnant and declined taking a 
pregnancy test prior to the I-131 therapy. Further, the hospital staff, 
knowing that the patient was also a physician on staff at the hospital, 
did not pursue a pregnancy test because they believed that the patient 
was aware of her pregnancy status.

Actions Taken To Prevent Recurrence

    Licensee--The licensee conducted a thorough investigation of the 
event, including identification of the root cause. The root cause of 
the event was identified as human error by the patient. The event 
appeared to be an isolated occurrence. No further actions were deemed 
necessary to prevent recurrence.
    NRC--The NRC conducted an inspection on August 26 and 27, 2003, 
with continued in-office review through September 30, 2003. The 
inspectors determined that the licensee made the required notifications 
to the patient, referring physician, and the NRC. No violations of NRC 
requirements were identified.
    This event is closed for the purpose of this report.
* * * * *
03-03 IVB Medical Event at Washington Hospital Center in Washington, DC

    Date and Place--May 6, 2003; Washington Hospital Center; 
Washington, DC.
    Nature and Probable Consequences--A patient undergoing IVB 
treatment of two areas within the right coronary artery for the 
treatment of restenosis was prescribed a dose of 23 Gy (2,300 rads) to 
each treatment site. Some difficulty was experienced in inserting the 
catheter to the first treatment site, but in the judgment of the 
treatment team, the catheter appeared to be inserted properly. 
Fluoroscopy was used to guide insertion and to position the source 
train. Upon completion of the first treatment, the catheter was moved 
to the second treatment position, as planned. When the source train was 
sent out for the second treatment, resistance was met and this time the 
catheter was replaced. The second treatment was successfully given.
    In documenting the treatment, the licensee reviewed the films taken 
during the treatment and printed a copy of the films for the patient's 
record. During this documentation, the medical physicist noted that the 
source markers were not in the right position and suspected that the 
treatment area was not covered for the first treatment given. The 
radiation oncologist and interventional cardiologist reviewed the films 
and determined that the source train was approximately 40 millimeters 
(mm) (1.6 in) away from the intended treatment site. Therefore, the 23 
Gy (2,300 rads) dose was delivered to an unintended treatment site.
    The NRC contracted a medical consultant to review the medical event 
and assess the probable deterministic

[[Page 24690]]

effects of the treatment to the wrong area of the patient's coronary 
artery. The medical consultant concluded that the dose to the normal 
segment of the right coronary artery reported in this case was well 
below the tolerance dose for coronary arteries and no effect was 
expected other than fibrosis of the right coronary artery vessel wall.
    Cause or Causes--This medical event was caused by human error, in 
that the licensee did not properly visualize the placement of the 
source train due, in part, to a lapse in time in the fluoroscopy 
performed during the treatment and the inherent inability to 
differentiate between the proximal and distal markers of the source 
train. In addition, a kink in the catheter may have prevented the 
source train from traversing to the correct area of the right coronary 
artery.

Actions Taken To Prevent Recurrence

    Licensee--The licensee immediately implemented measures to further 
enhance source positioning verification prior to initiation of future 
treatments. The measures included verification of fluoroscope 
calibration and reinstruction of the treatment team to fully appreciate 
the movement of both ends of the source train at the site prior to 
treatment. Further, the licensee recommended to the device manufacturer 
that they redesign the proximal and distal markers to make them more 
radiographically distinct from each other and from the guiding catheter 
marker.
    NRC--No violations of NRC requirements were identified. The NRC 
issued Information Notice 2003-09 describing medical events resulting 
from source positioning errors and is in the process of reviewing all 
events related to IVB since inception of this technology.
    This event is closed for the purpose of this report.
* * * * *
    03-04 Iodine-125 (I-125) Brachytherapy Seed Medical Event at 
Guthrie Healthcare System in Sayre, Pennsylvania

    Date and Place--May 24, 2001 (identified on June 12, 2003); Robert 
Packer Hospital (part of Guthrie Healthcare System), Sayre, 
Pennsylvania.
    Nature and Probable Consequences--In 2001, a patient received a 
permanent brachytherapy implant using I-125 seeds as treatment for 
prostate carcinoma. The authorized user prescribed a dose of 144 Gy 
(14,400 rads) to the prostate. The implant was performed under 
ultrasound guidance using 18 needles and 50 radioactive sources, as 
prescribed in the written directive. In June 2003, the patient returned 
for consultation regarding additional treatment after a diagnostic test 
indicated that the prostate cancer may have returned. A computerized 
tomography (CT) scan taken May 27, 2003, revealed that many of the 
seeds were not in the prostate but in adjacent tissue where they would 
have been ineffective in the treatment. The CT scan showed the array of 
seeds approximately 3 centimeters from the prostate. A review was then 
conducted of the May 2001 CT scan performed shortly after the initial 
implant procedure. This CT scan showed the array of I-125 seeds in the 
same location as in the May 2003 CT scan. The seed configuration 
resulted in a negligible dose to the prostate and a dose of 60 to 80 Gy 
(6,000 to 8,000 rads) to an adjacent structure, the penile bulb. The 
probable deterministic effects to the patient are being determined by 
NRC medical consultants. The patient and the patient's referring 
physician were notified of the event.
    Cause or Causes--The cause of this event is under investigation by 
the licensee.

Actions Taken To Prevent Recurrence

    Licensee--This event occurred in 2001 and involved an entirely 
different radiation oncology team than is currently employed by the 
licensee. The current radiation oncology team uses a different prostate 
implant protocol than was used in 2001. Reviews of the licensee's 
current prostate implant program by both the NRC and an independent 
physics consultant indicate that treatments performed since October 
2002 have been accurate.
    NRC--The NRC staff conducted a special safety inspection on June 
19, 2003. Subsequent to this inspection, the licensee (Guthrie 
Healthcare System) began to audit other prostate implants performed in 
2001 and identified additional cases of possible treatment errors. On 
July 28, 2003, the NRC issued a Confirmatory Action Letter (CAL) 
specifying actions the licensee agreed to perform, including evaluation 
of the root cause of the events and performance of an audit of past and 
current prostate implants. The NRC conducted a second special 
inspection on August 14, 2003. As of the date of this report, the 
licensee has reported a total of 21 possible medical events and is 
continuing the actions required by the CAL. It appears that the 
treatment errors may have been less extreme for the additional 20 cases 
reported by the licensee. An NRC medical consultant is currently 
evaluating these cases. NRC staff will consider enforcement options 
upon the completion of the licensee's and NRC's investigations.
    This event is considered open for the purpose of this report.
* * * * *
    03-05 Diagnostic Medical Event at Deaconess Hospital, Evansville, 
Indiana

    Date and Place--March 28, 2003; Deaconess Hospital; Evansville, 
Indiana.
    Nature and Probable Consequences--A nine-year-old patient, who had 
been prescribed a dosage of 0.148 MBq (4 [mu]Ci) in an I-131 capsule 
for a thyroid uptake study, instead received 15.6 MBq (421 [mu]Ci) of 
I-131 in liquid form. Because the patient was unable to swallow the 
capsule, the technologist placed a telephone request to a local 
commercial radiopharmacy for liquid I-131; however, the technologist 
erroneously ordered 15.6 MBq (421 [mu]Ci) of I-131 for the patient. The 
licensee identified the error while reviewing related paperwork on 
April 2, 2003. The referring physician, the patient, and the patient's 
family were informed of this event on April 3, 2003. The intended 
thyroid dose was approximately 0.13 Gy (13 rads), but the NRC's 
contracted medical consultant estimated that the patient received a 
thyroid dose of 13.7 Gy (1,370 rads) and an effective dose equivalent 
of 0.42 Gy (42 rads). According to the medical consultant, no acute 
radiation effects were anticipated to any organ, since no organ (except 
the thyroid) received more than 0.01 Gy (1.0 rad). The 13.7 Gy (1,370 
rads) dose will not cause radiation thyroiditis. The medical consultant 
also stated that there was insufficient data on juveniles to be 
reassured that a radiation dose in excess of 13.7 Gy (1,370 rads) to 
the thyroid would have no long-term consequences, given the increased 
radiosensitivity of the thyroid glands of children.
    Cause or Causes--This medical event was caused by human error in 
ordering the correct dosage.

Actions Taken To Prevent Recurrence

    Licensee--Corrective actions include (1) develop and use a 
standardized order form for liquid I-131 that will be faxed to the 
local nuclear pharmacy as written confirmation of the dosage ordered; 
(2) modify the computerized unit dose manager system to prevent an 
inappropriate dosage of I-131 from being entered into the computer 
system; (3) provide the local nuclear pharmacy with typical dosage 
ranges used by the licensee, which will be put into the nuclear 
pharmacy's computer and used as a secondary check to verify that the 
dosage ordered is appropriate for the

[[Page 24691]]

study or treatment to be performed; and (4) provide in-service training 
to the nuclear medicine technicians regarding the medical event.
    NRC--On August 29, 2003, a Notice of Violation was issued for a 
violation that included the failure to order the correct quantity of I-
131 as directed by the authorized user, to have a written directive 
dated and signed by an authorized user prior to the administration of 
the 15.6 MBq (421 [mu]Ci) I-131 dosage, and to administer a dosage 
within 20% of the prescribed dosage range for a thyroid uptake study 
using I-131.
    This event is considered closed for the purpose of this report.
* * * * *

Agreement State Licensees

    The NRC determined that the following events, which occurred at 
Agreement State licensed facilities during this reporting period, were 
significant enough for reporting as AOs:

AS 03-01 IVB Medical Event at Union Memorial Hospital in Baltimore, 
Maryland

    Date and Place--May 22, 2003, Union Memorial Hospital; Baltimore, 
Maryland.
    Nature and Probable Consequences--During a cardiac brachytherapy 
procedure conducted at the licensee's facility, a malfunction of the 
drive mechanism occurred with an IVB device containing a phosphorous-32 
source with an activity of 3.48 GBq (94 mCi). The malfunction occurred 
during the treatment of the third of three patients. The first two 
treatments were completed without incident. The treatment of the third 
patient was initiated with the dummy source successfully reaching the 
proper dwell position (confirmed visually via fluoroscopy) and 
returning to the cartridge. The active source was then advanced into 
the catheter, but when the source movement light continued to blink 
well after the anticipated transit time, the licensee initiated a 
fluoroscopic view of the treatment site. The source was not observed in 
the fluoroscopic field of view, so the licensee assumed a machine 
malfunction had occurred and initiated emergency procedures. Radiation 
surveys were performed, which confirmed that the source had stopped 
inside the patient. The indicator light on the console continued to 
indicate that the source was in transit even after the licensee 
confirmed the source was in the patient and not at the treatment site. 
The licensee was unable to retract the source to its shielded position 
using the machine interrupt, the system stop button, or the handwheel. 
At that point, the attending physician removed the catheter and source 
from the patient and accidently dropped them on the operating room 
floor. After the power cord was removed from the wall receptacle, the 
source retracted into its shielded position. The licensee stated that 
it took approximately 45 to 60 seconds to remove the source from the 
patient. The manufacturer's representative present during the treatment 
indicated that this period was 60 to 90 seconds. The licensee estimated 
a worst case dose to the wall of the patient's artery as approximately 
10.38 Gy (1,038 rads) based on a 60-second exposure time. The source 
delivery unit was taken to the licensee's ``hot'' laboratory after the 
event and the daily quality assurance (QA) checks were performed in the 
physics and clinical modes. The unit passed both QA checks. The 
manufacturer's representative present during the procedure immediately 
notified the manufacture's technical center. The device was returned to 
the manufacturer for evaluation and a new device was provided to the 
licensee.
    Cause or Causes--This medical event was caused by equipment 
malfunction. The manufacturer was able to simulate a similar type of 
failure on two occasions and is focusing on a timer chip as the 
possible cause of the malfunction. The manufacturer believes that a 
hardware problem and not the device's software caused the failure. The 
State of Maryland ruled out human error as the cause of the drive 
mechanism malfunction.

Actions Taken To Prevent Recurrence

    Licensee--Corrective actions included the implementation of revised 
procedures regarding dosimetry, emergency response, and notification of 
incidents. Training for the revised procedures was completed on 
November 12, 2003. The licensee also revised its annual Radiation 
Safety Training Program to ensure compliance with pertinent State 
regulations and revised procedures.
    State Agency--The State of Maryland conducted an investigation, and 
the State concurs with the licensee corrective actions that included 
implementation of revised procedures and an annual emergency exercise.
    This event is considered closed for the purpose of this report.
* * * * *
AS 03-02 Industrial Radiography Occupational Overexposure at a 
Temporary Jobsite in Ghent, Kentucky

    Date and Place--On November 12, 2002, the Kentucky Radiation Health 
& Toxic Agents (KRHTA) Branch was notified, by the licensee, that for 
the month of October 2002, a radiographer's total annual occupational 
dose was exceeded while working at a temporary jobsite near Ghent, 
Kentucky.
    Nature and Probable Consequences--The licensee reported an 
overexposure to a radiographer of 314 mSv (31.4 rem). A 3.81 
terabecquerel (TBq) (103 Ci) Ir-192 source was being retracted after an 
exposure. The radiographer who had entered the area was in the area for 
approximately 3 minutes before realizing the source was not fully 
retracted. Upon realizing that the source was not fully retracted, the 
radiographer immediately left the area, extended the source, and then 
retracted it to the housed position. The radiographer's dosimetry was 
sent for processing and results indicated a whole body exposure of only 
48.6 mSv (4.86 rem). However, the licensee, with assistance from the 
source manufacturer's Radiation Safety Officer (RSO), completed a 
reconstruction of the whole body exposure to the radiographer. The 
final result indicated an exposure of 300 mSv (30 rem) whole body from 
the event. This exposure was added to the radiographer's year-to-date 
exposure of 14 mSv (1.4 rem), for a total yearly whole body exposure of 
314 mSv (31.4 rem). Discussions with the KRHTA Branch, along with 
independent calculations, confirmed the 300 mSv (30 rem) event 
exposure. The licensee stated that the thermoluminescent dosimeter 
(TLD) and operating ratemeter were in the radiographer's pocket, an 
area that did not reflect true whole body exposure, and the alarm 
ratemeter was never heard in an alarming condition.
    Cause or Causes--This event was caused by inadequate operating 
procedures for the exposure device, improper placement of the TLD in 
the radiographer's pocket (rather than on his body), improper storage 
of the alarm ratemeter in his pocket (rather than on his body), and 
failure to survey the exposure device upon completion of the 
radiograph.

Actions Taken To Prevent Recurrence

    Licensee--The licensee's corrective actions included revision of 
the operating procedure for retracting the source into the exposure 
device, personnel training on the revised procedure and proper wearing 
of dosimetry devices, and annual refresher training on proper operation 
and responses of survey instrumentation. Additionally, the radiographer 
involved will receive an additional 40 hours of

[[Page 24692]]

radiation safety training prior to returning to work in radiography, 
and will be evaluated at least once a month for the next year.
    State Agency--The KRHTA Branch conducted an onsite investigation 
and concurred with the licensee's dose assessment and identification of 
the causes of the event. The licensee was issued a Notice of Violation 
and has provided corrective actions to the Commonwealth of Kentucky.
    This event is closed for the purposes of this report.
* * * * *
AS 03-03 Diagnostic Medical Event at Rush Copley Medical Center in 
Aurora, Illinois

    Date and Place--July 28, 2003; Rush Copley Medical Center; Aurora, 
Illinois.
    Nature and Probable Consequences--The Illinois Emergency Management 
Agency received a call on July 29, 2003, from a nuclear medicine 
technician at Rush Copley Medical Center in Aurora, Illinois. The 
technician reported that a patient who was to receive 148 MBq (4 mCi) 
of thallium-201 (Tl-201) for a heart test instead received 148 MBq (4 
mCi) of I-131 on July 28, 2003. The patient had been admitted the day 
before the event with an order for a treadmill heart stress test to be 
performed. The patient remained hospitalized at the facility until 
discharged after July 30, 2003.
    The circumstances of the event, as reported by the technician, are 
that both the exterior lead container and the syringe were labeled as 
containing a diagnostic unit dose of Tl-201. Although the injection 
occurred the previous day, it was not determined that I-131 was 
involved until the morning of July 29, 2003. Service engineers were 
called to the site on both days to inspect the gamma cameras used after 
attempts to image the patient failed. The reason became evident when a 
gamma camera flood source that had been made from what was thought to 
be the remaining Tl-201 material in the syringe from July 29 showed 
peaks consistent with I-131, rather than the expected Tl-201. The 
syringe had been assayed by the medical center before injection. The 
assayed amount showed the dose to be within the prescribed range for a 
typical 148 MBq (4 mCi) Tl-201 diagnostic administration.
    On Friday, July 25, 2003, the nuclear pharmacy received an order 
for five unit dose syringes of I-131 for the Veterinary Service Center 
(VSC) and two unit dose syringes of Tl-201 for Rush Copley Medical 
Center. When the computer generated orders and associated labels were 
segregated, one of the prescriptions for the Tl-201 was mistakenly 
substituted for I-131. The pharmacist did not realize the error and the 
I-131 dose (syringe) and its container were labeled with one of the Tl-
201 labels generated for the original order. On Monday, July 28, 2003, 
the pharmacy facility manager noted that only four I-131 prescriptions 
had been filled for VSC. Assuming the I-131 dose had not been filled 
with the others the previous Friday, July 25, 2003, he filled an 
additional syringe with I-131 to complete the order for VSC.
    The medical center estimates that a small amount of residual 
activity remained adhered to the walls of the syringe. Therefore, it 
estimates the amount of injected I-131 to be 148 MBq (4 mCi). Based on 
the package insert information for this material and assuming that an 
injected sodium iodide solution of I-131 results in a radiation 
absorbed dose similar to oral administration and that the patient had 
normal thyroid function (25% uptake), the dose to the patient's thyroid 
is approximately 51.95 Gy (5,195 rads).
    The medical center technician indicated that the patient involved 
had been contacted by the referring physician, onsite oncologists, and 
the medical center's administrator and lawyer and was informed as to 
what had happened at the initial time of discovery of the event. Later, 
a copy of the medical center's report to the agency was also provided 
to the patient. The medical center offered to perform routine blood 
analysis throughout the year to monitor any changes in thyroid 
activity. The patient had been advised as to the potential health 
effects of the medical event during that time and the need for routine 
followup testing. The patient has not returned to the medical center 
for any additional testing, diagnosis, or consultation.
    The medical center's oncologist indicated that it is very unlikely 
that any medical changes will be noted in the patient because the dose 
administered is only slightly larger than that typically ordered for 
whole body scans using I-131. Blood tests were taken immediately 
following the discovery of the event. Those tests suggest that the 
patient was hypothyroid as a preexisting condition to admittance.
    Cause or Causes--The medical event was caused by the mislabeling of 
the I-131 unit dose syringe. Other factors that led to the medical 
event include improper segregation of the prescriptions at the pharmacy 
and lack of a second means of verifying proper completion of the order.

Actions Taken To Prevent Recurrence

    Licensee--The pharmacy ceased dispensing therapeutic quantities of 
I-131 in unit dose syringes. Therapeutic doses of I-131 will only be 
dispensed in capsule form. This will preclude the possibility of a unit 
dose of diagnostic material being mistakenly filled with a quantity of 
therapeutic material. Additional corrective actions included (1) 
retraining of pharmacists, (2) implementation of a dual verification 
system for all prescriptions received, (3) implementation of a triple 
check system for dispensing compounds, and (4) testing a new bar code 
system for tracking all prescriptions.
    State Agency--On July 30, 2003, the State agency sent an 
investigator to the medical center and the nuclear pharmacy to observe 
licensed activities and to review the circumstances of the event. 
During those onsite visits, preliminary information reported by the 
medical center and pharmacy was confirmed. The pharmacy was cited for 
failure to properly fill the prescription as ordered by the physician. 
The State agency is holding this action item open pending enforcement 
action and will include a review of the corrective actions taken during 
the next routine inspection. The agency does not expect any additional 
significant information to be received or other notable action to be 
taken outside of the enforcement process.
    This event is considered closed for the purpose of this report.
* * * * *
AS 03-04 High Dose-Rate Afterloader (HDR) Medical Event at Saint 
Joseph's Hospital in Houston, Texas

    Date and Place--June 9, 10, and 11, 2003; Saint Joseph's Hospital; 
Houston, Texas.
    Nature and Probable Consequences--A cancer patient undergoing 
therapeutic radiation treatment for breast cancer received a 
superficial skin dose of 70 Gy (7,000 rads) to a circular area 
approximately 10 mm (0.4 in) in diameter. This error occurred using an 
HDR device. Deeper absorbed doses of 34 Gy (3,400 rads), 15 Gy (1,500 
rads), and 10 Gy (1,000 rads) have been estimated at depths of 10 mm 
(0.4 in), 20 mm (0.8 in), and 30 mm (1.2 in), respectively. These 
deeper doses were absorbed by the subcutaneous fat and muscle of the 
lower left chest wall. The patient had a slight erythema of the skin 
which measured 5 to 10 mm (0.2 to 0.4 in) in diameter approximately 2 
weeks after the radiation therapy injury.
    The incorrect placement of the source in the catheter was detected 
on June 11, 2003, between treatment fractions 5 and

[[Page 24693]]

6. The patient and referring physician were notified of the treatment 
error and the facts involved with this treatment. The patient elected 
to continue treatment with a modified treatment plan after the source 
location was corrected. A new plan was generated representing a 
composite of the unintended dose to the skin of the lower left chest 
wall and the intentional dose prescribed in the original treatment 
plan.
    The attending physician, who was present during treatment, followed 
the patient's progress for any needed medical intervention due to 
exposure to the HDR source. The patient's erythema of the skin failed 
to heal and developed into an ulceration. The ulceration was surgically 
excised by the referring physician. After excision, the area fully 
healed within a period of approximately two months. The patient 
continues to be monitored by the referring physician.
    Causes or Causes--During the setup of the HDR unit with the 
approved treatment plan, the source was instructed to stop at the 20th 
position from the catheter tip. The 20th stop resulted in the source 
stopping at 20 cm (7.9 in) from the catheter tip instead of the planned 
20 mm (0.8 in) from the catheter tip. This was due to failure to 
correct the default value step size from 10 mm to 1 mm (0.4 in. to .04 
in) as specified in the treatment plan. This failure was a human error 
in the copying of the treatment plan into the device's control console 
after the initial QA test. After the QA test the physician requested 
that the plan instruction be copied into a new plan, after the initial 
QA films had been approved. This procedure is required as the device 
manufacturer does not have a separate QA mode that allows QA without 
recording the QA tests as a fractional treatment.

Actions Taken To Prevent Recurrence

    Licensee--The facility instituted a policy of comparing the console 
instructions to the approved QA record prior to each treatment 
fraction. In addition the medical physicist has made two suggestions 
for product improvement (1) the addition of a physics QA mode to allow 
the physicist to test a treatment plan without having it recorded as a 
treatment fraction to the patient; and (2) the placement of a display 
on the operator's console that graphically displays the actual position 
of the source within the catheter. Presently, the source position must 
be deduced by multiplying the current dwell stop by the step size.
    State Agency--The licensee's comments and suggested product 
improvements were forwarded to the manufacturer's regulatory affairs 
office. The licensee was cited for failure to verify that the specific 
details of the administration were in accordance with the treatment 
plan and the written directive. Escalated enforcement actions were 
taken against the licensee.
    This event is closed for the purposes of this report.
* * * * *
AS 03-05 Overexposure at Monsanto Chemical Plant in Luling, Louisiana
    Date and Place--June 28, 2003, to July 10, 2003; Monsanto Chemical 
Plant; Luling, Louisiana.
    Nature and Probable Consequences--The licensee notified the 
Louisiana Office of Environmental Services on July 10, 2003, that a 
radiation overexposure had occurred to members of the public due to a 
loss of control of a 37 GBq (1 Ci) cesium-137 (Cs-137) source that 
became dislodged from a damaged fixed gauge. The licensee stated that 
on June 29, 2003, a Monsanto maintenance technician noticed that the 
gauge's handle mechanism had broken off and fallen to the floor. The 
technician picked up the broken pieces and placed them on the Monsanto 
Planner's desk. The Planner was not present. The Planner returned to 
work on July 1, 2003, but did not discover the pieces until July 10, 
2003. The Planner thought the parts were the gauge's locking mechanism 
and went to the area where the fixed gauge had been mounted and 
realized that the gauge's source was missing. After realizing that the 
parts contained the unshielded Cs-137 source, the licensee evacuated 
the building and secured the area. On July 11, 2003, a representative 
from a consulting company arrived on-site to perform an area survey, 
retrieve the source from the Planner's desk, and place the source in a 
secure storage area. The licensee requested that the manufacturer 
evaluate the failed gauge and conduct an assessment of the remaining 
gauges. On July 19, 2003, a representative from the device manufacturer 
removed the source from the Monsanto plant.
    It was determined that the Planner occupied the desk for 
approximately 50 to 60 hours and received a whole body dose of 
approximately 400 mSv (40 rem). This determination was based on an 
analysis of the Planner's schedule and work habits together with the 
radiation dose rate of the source. The technician who carried the 
source to the Planner's desk received an extremity dose of 
approximately 18,000 mSv (1,800 rem) to the hand. Reenactments were 
performed to estimate the exposures to 100 individuals employed by the 
plant. The estimates were determined by the time spent and proximity to 
the source. The highest exposure was estimated to be 740 mSv (74 rem) 
and the next highest exposure 180 mSv (18 rem). Altogether, 42 
nonradiation workers exceeded the 1 mSv (0.1 rem) exposure limit to 
members of the general public. The workers are considered to be members 
of the public, and not radiation workers, because they are not exposed 
to radiation from licensed radioactive material as a normal part of 
their work. Others may have also been exposed at lower levels. Blood 
tests were performed for seven individuals, but revealed no cell 
changes. No one has shown signs of sickness or erythema.
    The licensee contacted the Radiological Emergency Assistance 
Center/Training Site (REAC/TS) in Oak Ridge, Tennessee, and requested 
its assistance in having a cytogenetic blood study performed for the 
Planner. The licensee reported that it appears that vibration of the 
gauge caused the source holder and the attached source to fall. Surveys 
of the relevant areas and wipe tests on the source did not reveal any 
source leakage.
    Cause or Causes--Monsanto believes the cause of the incident was 
corrosion of the epoxy that holds the source in place. However, the end 
plate was held in place by one tack weld and the vibration of the gauge 
could have contributed to the gauge becoming dislodged.

Actions Taken To Prevent Recurrence

    Licensee--The decision has been made to take this type of device 
out of service and replace it with a newer model. Until the devices are 
removed from service, weekly visual inspections on the devices will be 
performed. The Planner and Monsanto engineers/technicians were trained 
only to recognize the radiation posting on the device. Now the safety 
training includes pictures of the device, its components, and the 
radioactive capsule.
    State Agency--The licensee was cited for two violations. One 
violation was for the exposure of a nonradiation worker in excess of 1 
mSv (0.1 rem) in a year, and the other was for creating a radiation 
area in an unrestricted area that exceeded 0.02 mSv (0.002 rem) in any 
one hour. The event was referred to the State of Louisiana's 
Enforcement Section.
    This event is closed for the purpose of this report.
* * * * *

[[Page 24694]]

AS 03-06 Brachytherapy Medical Event at University Hospitals of 
Cleveland in Cleveland, Ohio

    Date and Place--May 13, 2003; University Hospitals of Cleveland; 
Cleveland, Ohio.
    Nature and Probable Consequences--On May 22, 2003, the Ohio 
Department of Health notified the NRC Operations Center of an apparent 
brachytherapy medical event at University Hospitals of Cleveland. The 
licensee reported a radiation treatment to the wrong target area during 
a brachytherapy prostate procedure using 59 I-125 seeds, each 
containing 13 MBq (0.351 mCi) for a total activity of 765 MBq (20.71 
mCi). The treatment resulted in a distribution of seeds in areas other 
than prescribed.
    An unintended area of the prostate gland received approximately 1.4 
Gy (140 rads) due to seeds implanted outside of the intended cancer 
cell site. The licensee determined that 31% of the bladder received 72 
Gy (7,200 rads) and 3% of the rectum received 72 Gy (7,200 rads).
    Cause or Causes--Unusual anatomical aspects of the seminal/prostate 
vesicle under ultrasound hampered the physician's ability to correctly 
place the seeds fully within the intended preplan margins. In addition, 
seed visualization on fluoroscopy was suboptimal.

Actions Taken To Prevent Recurrence

    Licensee--Faculty and staff will increase efforts to identify 
unusual prostate anatomical features during the preplanning process; 
specifically, they will continue to cross-check and verify seed 
position in relation to underlying anatomy. Corrective actions taken by 
the licensee include (1) the introduction of stabilization needles to 
assist in keeping the prostate fixed relative to the base plate, the 
ultrasound probe, and surrounding tissues during the localization and 
the seed deposition process and (2) the use of a more radio-opaque seed 
to facilitate positive location during procedures viewed under 
fluoroscopy. The patient and referring physician were notified of the 
medical event.
    State Agency--The Ohio Department of Health performed an 
investigation of the event.
    This event is closed for the purpose of this report.
* * * * *
AS 03-07 Diagnostic Medical Event at Christus Santa Rosa; San Antonio, 
Texas

    Date and Place--June 11, 2003; Christus Santa Rosa; San Antonio, 
Texas.
    Nature and Probable Consequences--A patient received 85.1 MBq (2.3 
mCi) of I-131 instead of the prescribed dosage of 11.1 MBq (0.3 mCi) of 
I-131. The licensee discovered the error when the patient returned 
after 48 hours for a scan. The physician's written order requesting a 
thyroid scan for thyroiditis was misunderstood by the technologist as a 
request for a ``whole body image'' instead of a ``thyroid up-take and 
scan''. As a result, the technologist ordered the wrong dose for the 
prescribed procedure. Both the referring physician and the patient have 
been informed of the error.
    Cause or Causes--The medical event was caused by human error. The 
wrong dosage was administered to the patient because the written order 
for the I-131 procedure was misread by the administering technologist.

Actions Taken To Prevent Recurrence

    Licensee--The licensee implemented revised procedures mandating 
that a physician review all prescriptions requiring the use of I-131 
and concur on the correct dosage.
    State Agency--The State accepted the licensee's report and 
corrective actions as appropriate.
    This event is closed for the purpose of this report.
* * * * *
AS 03-08 Therapy Medical Event at Marian Medical Center in Santa Maria, 
California

    Date and Place--April 25, 2002; Marian Medical Center; Santa Maria, 
California. This event was not determined to be an AO until the 
preparation of the FY 2003 report.
    Nature and Probable Consequences--A patient was prescribed a 
therapeutic dose to the thyroid of I-131 with an activity of 296 MBq (8 
mCi) but was erroneously administered 3,700 MBq (100 mCi) of I-131 
instead. The error was discovered immediately and was reported to the 
RSO and the referring physician. After consultation, the RSO and 
referring physician prescribed suppressive and hydration therapy to the 
patient immediately in order to minimize the patient's absorbed dose. 
The suppressive therapy blocked the thyroid from absorbing the total 
dose and the hydration therapy was given to accelerate the excretion of 
the radioactivity from the body.
    The dose to the patient was calculated to be 0.03 Gy (3 rads) to 
the whole body and 38.7 Gy (3,870 rads) to the thyroid. No adverse 
health effects are expected.
    Cause or Causes--The State found that the medical event occurred 
due to human error. Two I-131 capsules had been delivered that day for 
two patients who were to receive iodine therapy. The capsule containing 
3.7 GBq (100 mCi) was given to the first patient. The error was 
recognized before the second patient was treated; therefore, the second 
I-131 capsule was never administered. The technologist failed to check 
the labeling and did not verify the dose using a dose calibrator.

Actions Taken To Prevent Recurrence

    Licensee--Corrective actions included (1) counseling the 
technologist to review the labels on the vial and to check the dose in 
the dose calibrator before administration, (2) providing in-service 
training to technologists on proper procedures, (3) implementing new 
procedures requiring the doctor to check the label to ensure the 
patient will be administered the correct dose, and (4) administering I-
131 to no more that one patient daily.
    State Agency--The State has reviewed and accepted the licensee's 
corrective actions.
    This event is closed for the purposes of this report.
* * * * *
AS 03-09 Gamma Stereotactic Radiosurgery Device Medical Event at 
Bayfront Medical Center, Inc., in St. Petersburg, Florida

    Date and Place--Between August and October 2002; Bayfront Medical 
Center; St. Petersburg, Florida.
    Nature and Probable Consequences--On October 31, 2002, the Florida 
Bureau of Radiation Control was notified that 10 patients undergoing 
Gamma Stereotactic Radiosurgery (gamma knife) had received a dose or 
doses at least 50% greater than prescribed. The prescribed treatments 
ranged from 12.2 to 24 Gy (1,220 to 2,400 rads) at the 50% isodose 
curve; however, the delivered doses to the patients ranged between 19.2 
and 38.4 Gy (1,920 and 3,840 rads) at the 50% isodose curve, which is 
60% greater than the treatment prescribed. The patients were diagnosed 
with a variety of brain disorders (vascular diseases, tumors, and 
functional targets such as selected nerves). A treatment plan was 
developed and reviewed by the physicist, and the doses were 
administered using a gamma knife device. On October 30, 2002, while 
performing a routine QA, the RSO discovered that the physics parameters 
in the treatment planning file had an incorrect calibration factor. 
Further investigation identified that the system had an older 
calibration date which resulted in the incorrect information that the 
sources had 60% less activity. The medical events were discovered 
during a review of all patient files.

[[Page 24695]]

    The medical events were reported to two authorized users and three 
referring physicians. Notification of the medical event was provided to 
nine of the patients or patients' responsible guardians and they were 
subsequently provided a copy of the report pertinent to that patient. 
The authorized user does not anticipate any change in the patient's 
condition from the additional exposure. The licensee's authorized users 
noted that these doses are still within the published literature. 
During the notifications it was discovered that one of the patients had 
died as a result of the patient's disease. The licensee's authorized 
users stated that this patient was given palliative treatment for four 
metastatic lesions that were not close to any critical structure. The 
patient died approximately 2 months after the treatment, which was the 
typical period of life expectancy for a patient with this type and 
stage of disease.
    Cause or Causes--The State was not able to identify how the 
calibration date was changed in the treatment planning software physics 
protocol file. However, it is the licensee's responsibility, through an 
effective quality management program, to ensure that the treatment is 
administered with high confidence as directed by the authorized user.

Actions Taken To Prevent Recurrence

    Licensee--The licensee has revised its quality management program 
to include additional daily checks to verify that the expected dose 
rate agrees with the dose rate shown on the treatment planning software 
physics protocol output to within 1%. The gamma knife manufacturer 
issued a notice dated November 4, 2002, to all customers utilizing the 
treatment planning system specific to the gamma knife used to treat 
these patients. The notice requested customers to check the physics 
protocol and to run tests to verify dose calibration factors after any 
treatment planning system service or software reinstallation.
    State Agency--The State conducted an onsite investigation that 
included interviews with licensee personnel involved and a 
representative from the device's manufacturer on November 12-13, 2002. 
In the licensee's medical event report, the licensee indicated the 
device manufacturer installed a peripheral printer on August 26, 2002. 
The licensee's report also indicated that on this date the source 
calibration information was changed. During the investigation the 
manufacturer stated that it was unable to recreate the occurrence. 
Telephone interviews were conducted with service personnel from the 
device manufacturer. The State also consulted with an independently 
contracted physicist with experience specific to the gamma knife and 
its treatment planning system to determine the state of the equipment. 
It was determined that the licensee's quality management program did 
not routinely verify calibration information as compared to treatment 
planning dose rates. State actions for this case are still pending.
    This event is closed for the purpose of this report.
* * * * *

    Dated at Rockville, Maryland this 28th day of April 2004.

    For the Nuclear Regulatory Commission
Annette L. Vietti-Cook,
Secretary of the Commission.