    <SECTION>
        <SECTNO>§ 15.241</SECTNO>
        <SUBJECT>Operation in the band 174-216 MHz.</SUBJECT>
        <P>(a) Operation under the provisions of this section is restricted to biomedical telemetry devices.</P>
        <P>(b) Emissions from the device shall be confined within a 200 kHz band which shall lie wholly within the frequency range of 174-216 MHz.</P>
        <P>(c) The field strength of any emissions radiated within the specified 200 kHz band shall not exceed 1500 microvolts/meter at 3 meters. The field strength of emissions radiated on any frequency outside of the specified 200 kHz band shall not exceed 150 microvolts/meter at 3 meters. The emission limits in this paragraph are based on measurement instrumentation employing an average detector. The provisions in § 15.35 for limiting peak emissions apply.</P>
    </SECTION>